Browse > Article
http://dx.doi.org/10.9718/JBER.2013.34.3.105

Estimation of Respiration Using Photoplethysmograph During Sleep  

Park, Jong-Uk (Department of Biomedical Engineering, Yonsei University)
Lee, Jeon (Department of Biomedical Engineering, Yonsei University)
Lee, Hyo-Ki (Department of Biomedical Engineering, Yonsei University)
Kim, Hojoong (Division of Pulmonary and Critical Care Medicine, Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine)
Lee, Kyoung-Joung (Department of Biomedical Engineering, Yonsei University)
Publication Information
Journal of Biomedical Engineering Research / v.34, no.3, 2013 , pp. 105-110 More about this Journal
Abstract
Respiratory signal is one of the important physiological information indicating the status and function of the body. Recent studies have provided the possibility of being able to estimate the respiratory signal by using a change of PWV(pulse width variability), PRV(pulse rate variability) and PAV(pulse amplitude variability) in the PPG (photoplethysmography) signal during daily life. But, it is not clear whether the respiratory monitoring is possible even during sleep. Therefore, in this paper, we estimated the respiration from PWV, PRV and PAV of PPG signals during sleep. In addition, respiration rates of the estimated respiration signal were calculated through a time-frequency analysis, and errors between respiration rates calculated from each parameter and from reference signal were evaluated in terms of 1 sec, 10 sec and 1 min. As a result, it showed the errors in PWV(1s: $36.38{\pm}37.69$ mHz, 10s: $36.53{\pm}38.16$ mHz, 60s: $30.35{\pm}38.72$ mHz), in PRV(1s: $1.45{\pm}1.38$ mHz, 10s: $1.44{\pm}1.37$ mHz, 60s: $0.45{\pm}0.56$ mHz), and in PAV(1s: $1.05{\pm}0.81$ mHz, 10s: $1.05{\pm}0.79$ mHz, 60s: $0.56{\pm}0.93$ mHz). The errors in PRV and PAV are lower than that of PWV. Finally, we concluded that PRV and PAV are more effective than PWV in monitoring the respiration in daily life as well as during sleep.
Keywords
Respiration rate; Photoplethysmography; Pulse width variability; Pulse rate variability; Pulse amplitude variability;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 C.A. Kushida, "Practice Parameters for the Indications for Polysomnography and Related Procedures:An Update for 2005", SLEEP, vol. 28, no. 4, pp. 499-601, 2005.   DOI
2 P. Langely, E.J. Bowers and A. Murray, "Principal Component Analysis as a Tool for Analyzing Beat-to-Beat Changes in ECG Features:Application to ECG-Derived Respiration", IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, vol. 57, no. 4, pp. 821-829, 2010.   DOI   ScienceOn
3 S. Chen, M. Bolic, V.Z. Groza, H.R. Dajani and S. Rajan, "Extraction of Breathing Signal and Suppression of Its Effects in Oscillometric Blood Pressure Measurement", IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, vol. 60, no. 5, pp. 1741-1750, 2011.   DOI   ScienceOn
4 J. Lazaro, E. Gil, R. Bailon, A. Minchole and P. Laguna, "Deriving respiration from photoplethysmographic pulse width", Med Biol Eng Comput, vol. 51, no. 1-2, pp. 233-242, 2013.   DOI   ScienceOn
5 S.H. Baek, "Clinical Vascular Medicine", Korea Vascular Research Workng Group, 2009, pp. 1-295.
6 S.M. Lee, S.J. Lee, J.M. Ahn and J.K. Kim, "A Study on an Optimal Respiration Rate for the ANS Assessment based on RSA Analysis", J. Biomed. Eng. Res, vol. 28, no. 4, pp. 503- 511, 2007   과학기술학회마을
7 C.K Lee, H.S Shin, S.D Min, Y.H Yun and M.H Lee, "A Study on Comparison PPG Variability with Heart Rate Variability in the Sitting Position During Paced Respiration", IFMBE Proceedings, Sep 2009, pp. 1703-1705.
8 H.S Shin, C.K Lee and M.H Lee, "Adaptive threshold method for the peak detection of photoplethysmographic waveform", Computers in Biology and Medicine, vol. 39, no. 12, pp. 1145-1152, 2012.