Browse > Article
http://dx.doi.org/10.9718/JBER.2008.29.2.132

Development of an ACL Anchor: Effects of the Design Parameters on the Performance of a New Anterior Cruciate Ligament Fixation Device  

Kim, Jong-Dae (Department of Bionano-System Engineering, Graduate School, Chonbuk National University)
Oh, Chae-Youn (Division of Mechanical and Aerospace System Engineering, Chonbuk National University)
Kim, Cheol-Sang (Department of Bionano-System Engineering, Graduate School, Chonbuk National University)
Publication Information
Journal of Biomedical Engineering Research / v.29, no.2, 2008 , pp. 132-138 More about this Journal
Abstract
We investigated the biomechanical properties of a newly designed self-expansion type anterior cruciate ligament (ACL) anchor. The ACL anchor consists of the ring section giving the elastic force, the wedge for maintaining in contact with the femur tunnel wall and the link suspending hamstring graft or artificial ligament. The main design parameters that determine the performance of this device were the expansion angle (${\theta}$) and the thickness ($t_R$). The Ti6Al4V anchors were heated after inserting in a jig for 1 hour at $800^{\circ}C$ in a protective argon gas atmosphere and allowed to cool to room temperature in the furnace. In order to investigate the influence of the expansion angle and the thickness of the ring on the biomechanical properties of the anchor, the maximum pull-out load, stiffness and slippage of the ACL anchor were measured using the pull-out tester, and statistical analyses were also executed. The present results showed that the design parameters gave a significant effect on the performance of the self- expansion type of anchor. The pull-out load of the ACL anchors significantly increased as the thickness of the ring section was increased, having a similar trend for both expansion angles. The ACL anchor showed about 2.5 times higher values of the pull-out load than that of the minimum load (500N)required for the "accelerated rehabilitation". The optimum ${\theta}$ and $t_R$ values of this ACL anchor were suggested to have sufficient resistance against the pull-out force, high stiffness and relatively low slippage after ACL reconstruction.
Keywords
anterior cruciate ligament; ligament fixator; accelerated rehabilitation; self-expansion; interference screw;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Brown, C. H., Sklar, J. H., 1998, 'Endoscopic anterior cruciate ligament reconstruction using quadrupled hamstring tendons and endobutton femoral fixation,' Techniques in orthopedics, vol. 13, no. 3, pp. 281-298   DOI
2 Fu, F. H., Bennett, C. H., Ma, C. B.,Menetrey, J., Lattermann, C., 1999, 'Current Trends in Anterior Cruciate Ligament Reconstruction. Part I. Biology and Biomechanics of Reconstruction,' The American journal of sports medicine, vol. 27, no. 6, pp. 821-830   DOI   PUBMED
3 Weiler, A., Hoffmann, R. F. G., Siepe, C. J., Kolbeck, S. F., Sudkamp, N. P., 2000, 'The influence of screw geometry on hamstring tendon interference fit fixation,' The American journal of sports medicine, vol. 28, no. 3, pp. 356-359   DOI   PUBMED
4 Weiler, A., Richter, M., Schmidmaier, G.,Kandziora, F., Sudkamp, N. P., 2001, 'The EndoPearlDevice Increases Fixation Strength and Eliminates Construct Slippage of Hamstring Tendon Grafts With Interference Screw Fixation,' Arthroscopy : the journal of arthroscopic & related surgery, vol. 17, no. 4, pp. 353-359   DOI   ScienceOn
5 Yukio, Y., Shinichi, Y., Masahiro, K., Tetsuji, Y., Ryosuke, K., Hirotsugu, M., 2005, 'The load of an implanted graft during and after fixation in anterior cruciate ligament reconstruction,' Knee surgery, sports traumatology, arthroscopy, vol. 13, no. 2, pp. 101-106   DOI
6 C. S. Kim, K. Y. Shin, K. E. Hwang, M. Y. Jung, D. H. Lee, 2004, 'CPC Precipitation Behavior of Titanium Phosphide Surfaces Obtained from Heat Treated Ti6Al4V buried in Hydroxyapatite Paste,' Annals of transplantation, vol. 9, no. IA (Suppl.), pp. 43-47
7 Aune, A. K.,Ekeland, A., Cawley, Patrick, W., 1998, 'Interference screw fixation of hamstring vs. patellar tendon grafts for anterior cruciate ligament reconstruction,' Knee surgery, sports traumatology, arthroscopy, vol. 6, no. 2, pp. 99-102   DOI
8 Brand, J.,Weiler, A., Caborn, D. N. M., Brown, C. H., Johnson, D. L., 2000, 'Current Concepts. Graft Fixation in Cruciate Ligament Reconstruction,' The American journal of sports medicine, vol. 28, no. 5, pp. 761-774   DOI   PUBMED
9 Aglietti, P., Buzzi, R., Zaccherotti,G., De Biase, P., 1994, 'Patellar tendon versus doubled semitendinosus and gracilis tendons for anterior cruciate ligament reconstruction,' The American journal of sports medicine, vol. 22, no. 2, pp. 211-218   DOI   ScienceOn
10 Shelbourne, K. D., Nitz, P., 1990, 'Acceleratedrehabilitation after anterior cruciate ligament reconstruction,' The American journal of sports medicine, vol. 18, no. 3, pp. 292-299   DOI   ScienceOn
11 Cooper, D. E., Deng, X. H., Burstein, A. L., Warren, R. F., 1993, 'The strength of the central third patellar tendon graft: A biomechanical study,' The American journal of sports medicine, vol. 21, no. 6, pp. 818-824   DOI   ScienceOn
12 Lee, Myung Chul, Jo, Hyunchul, Bae, Tae-Soo, Jang, Jin Dae, Seong, Sang Cheol, 2003, 'Analysis of initial fixation strength of press-fit fixation technique in anterior cruciate ligament reconstruction : A comparative study with titanium and bioabsorbable interference screw using porcine lower limb,' Knee surgery, sports traumatology, arthroscopy, vol. 11, no. 2, pp. 91-98   DOI
13 Hwang, K. E., Kim, C. S., 2003, 'Interface characteristics changed by heat treatment of Ti materials with hydroxyapatite, Materials science & engineering. C, Biomimetic and supramolecular systems, vol. 23, no. 3, pp. 401-405   DOI   ScienceOn
14 Robbe, R., Paletta, G. A., 2004, 'Soft-Tissue Graft in Anterior Cruciate Ligament Reconstruction,' Operative techniques in sports medicine, vol. 12, no. 3, pp. 188-194   DOI   ScienceOn
15 Weiler, A.,Hoffmann, R. F. G., Bail, H. J., Rehm, O., Sudkamp, N. P., 2002, 'Tendon healing in a bone tunnel: Part II: Histological analysis after biodegradable interference fit fixation in a model of anterior cruciate ligament reconstruction in sheep,' Arthroscopy : the journal of arthroscopic & related surgery, vol. 18, no. 2, pp. 124-135   DOI   ScienceOn
16 Weimann, A., Rodieck, M., Zantop, T., Hassenpflug, J., Petersen, W., 2005, 'Primary Stability of Hamstring Graft Fixation With Biodegradable Suspension Versus Interference Screws,' Arthroscopy : the journal of arthroscopic & related surgery, vol. 21, no. 3, pp. 266-277   DOI   ScienceOn
17 ASTM F-1839, 2001, 'Standard specification for rigid polyurethane foam for use as a standard material for testing orthopedic devices and instruments,' pp. 1-6
18 Fabbriciani, C. ; Mulas, P.D. ; Ziranu, F. ; Deriu, L. ; Zarelli, D. ; Milano, G., 2005, 'Mechanical analysis of fixation methods for anterior cruciate ligament reconstruction with hamstring tendon graft. An experimental study in sheep knees,' The Knee, vol. 12, no. 2, pp. 135-138   DOI   ScienceOn
19 Zantop, T., Weimann, A., Rummler, M., Hssenpflug, J., Petersen, W., 2004, 'Initial Fixation Strength of Two Bioabsorbable Pins for the Fixation of Hamstring Grafts Compared to Interference Screw Fixation : Single Cycle and Cyclic Loading,' The American journal of sports medicine, vol. 32, no. 3, pp. 641-649   DOI
20 Kudo, T., Tohyama, H., Minami, A.,Yasuda, K., 2005, 'The effect of cyclic loading on the biomechanical characteristics of the femurgrafttibia complex after anterior cruciate ligament reconstruction using Bone Mulch$^{TM}$ screw/WasherLoc$^{TM}$ fixation,' Clinical biomechanics, vol. 20, no. 4, pp. 414-420   DOI   ScienceOn
21 Magen, H. E., Howell, S. M., Hull, M. L., 1999, 'Structural properties of six tibial fixation methods for anterior cruciate ligament soft tissue grafts,' The American journal of sports medicine, vol. 27, no. 1, pp. 35-43   DOI   PUBMED
22 Mann, C. J., Costi, J. J., Stanley, R.M., Dobson, P. J., 2005, 'The effect of screw taper on interference fit during load to failure at the soft tissue/bone interface,' The Knee, vol. 12, no. 5, pp. 370-376   DOI   ScienceOn
23 Morris, M. W. J., Williams, J. L.,Thake, A. J., Lang, Y., Brown, J. N., 2004, 'Optimal screw diameter for interference fixation in a bone tunnel: a porcine model,' Knee surgery, sports traumatology, arthroscopy, vol. 12, no. 3, pp. 486-489
24 Kousa, P., Järvinen, T. L. N., Vihavainen, M., Kannus, P., Järvinen, M., 2003, 'The Fixation Strength of Six Hamstring Tendon Graft Fixation Devices in Anterior Cruciate Ligament Reconstruction Part I: Femoral Site,' The American journal of sports medicine, vol. 31, no. 2, pp. 174-181   DOI   PUBMED
25 ASTM F-1691, 2001, 'Standard test method for determining axial pull-out strength of medical bone screws', pp. 1-3
26 To, J. T.,Howell, S. M., Hull, M. L., 1999, 'Contributions of Femoral Fixation Methods to the Stiffness of Anterior Cruciate Ligament Replacements at Implantation,' Arthroscopy : the journal of arthroscopic & related surgery, vol. 15, no. 4, pp. 379-387   DOI   ScienceOn
27 Piltz, S., Strunk, P., Meyer, L.,Plitz, W., Lob, G., 2004, 'Fixation strength of a novel bioabsorbable expansion bolt for patellar tendon bone graft fixation: an experimental study in calf tibial bone,' Knee surgery, sports traumatology, arthroscopy, vol. 12, no. 5, pp. 376-383