Verification of Gated Radiation Therapy: Dosimetric Impact of Residual Motion  

Yeo, Inhwan (Loma Linda University Medical Center)
Jung, Jae Won (East Carolina University)
Publication Information
Progress in Medical Physics / v.25, no.3, 2014 , pp. 128-138 More about this Journal
Abstract
In gated radiation therapy (gRT), due to residual motion, beam delivery is intended to irradiate not only the true extent of disease, but also neighboring normal tissues. It is desired that the delivery covers the true extent (i.e. clinical target volume or CTV) as a minimum, although target moves under dose delivery. The objectives of our study are to validate if the intended dose is surely delivered to the true target in gRT and to quantitatively understand the trend of dose delivery on it and neighboring normal tissues when gating window (GW), motion amplitude (MA), and CTV size changes. To fulfill the objectives, experimental and computational studies have been designed and performed. A custom-made phantom with rectangle- and pyramid-shaped targets (CTVs) on a moving platform was scanned for four-dimensional imaging. Various GWs were selected and image integration was performed to generate targets (internal target volume or ITV) for planning that included the CTVs and internal margins (IM). The planning was done conventionally for the rectangle target and IMRT optimization was done for the pyramid target. Dose evaluation was then performed on a diode array aligned perpendicularly to the gated beams through measurements and computational modeling of dose delivery under motion. This study has quantitatively demonstrated and analytically interpreted the impact of residual motion including penumbral broadening for both targets, perturbed but secured dose coverage on the CTV, and significant doses delivered in the neighboring normal tissues. Dose volume histogram analyses also demonstrated and interpreted the trend of dose coverage: for ITV, it increased as GW or MA decreased or CTV size increased; for IM, it increased as GW or MA decreased; for the neighboring normal tissue, opposite trend to that of IM was observed. This study has provided a clear understanding on the impact of the residual motion and proved that if breathing is reproducible gRT is secure despite discontinuous delivery and target motion. The procedures and computational model can be used for commissioning, routine quality assurance, and patient-specific validation of gRT. More work needs to be done for patient-specific dose reconstruction on CT images.
Keywords
gated radiation therapy; residual motion; dosimetric evaluation;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Keall PJ, Mageras GS, Balter JM, et al: The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Med Phys 3874-900 (2006)
2 Berbeco RI, Nishioka S, Shirato H, Chen GTY, Jiang SB: Residual motion of lung tumours in gated radiotherapy with external respiratory surrogates. Phys Med Biol 50(16):3655-67 (2005)   DOI   ScienceOn
3 Hugo GD, Agazaryan N, Solberg TD: An evaluation of gating window size, delivery method, and composite field dosimetry of respiratory-gated IMRT. Med Phys 29(11):2517 (2002)   DOI   ScienceOn
4 ICRU-62: Prescribing, recording and reporting photon beam therapy (Supplement to ICRU Report 50) (1999)
5 Kubo HD, Wang L: Compatibility of Varian 2100C gated operations with enhanced dynamic wedge and IMRT dose delivery. Med Phys 27(8):1732 (2000)   DOI
6 Hugo GD, Agazaryan N, Solberg TD: The effects of tumor motion on planning and delivery of respiratory-gated IMRT. Med Phys 30(6):1052 (2003)   DOI   ScienceOn
7 Li XA, Stepaniak C, Gore E: Technical and dosimetric aspects of respiratory gating using a pressure-sensor motion monitoring system. Med Phys 33(1):145 (2006)   DOI   ScienceOn
8 Dietrich L, Tücking T, Nill S, Oelfke U: Compensation for respiratory motion by gated radiotherapy: an experimental study. Phys Med Biol 50(10):2405-14 (2005)   DOI
9 Nelms BE, Ehler E, Bragg H, Tome W a: Quality assurance device for four-dimensional IMRT or SBRT and respiratory gating using patient-specific intrafraction motion kernels. J Appl Clin Med Phys 8(4):2683 (2007)
10 Duan J, Shen S, Fiveash JB, et al: Dosimetric effect of respiration-gated beam on IMRT delivery. Med Phys 30(8):2241 (2003)   DOI
11 Létourneau D, Gulam M, Yan D, Oldham M, Wong JW: Evaluation of a 2D diode array for IMRT quality assurance. Radiother Oncol 70(2):199-206 (2004)   DOI   ScienceOn
12 George R, Chung TD, Vedam SS, et al: Audio-visual biofeedback for respiratory-gated radiotherapy: impact of audio instruction and audio-visual biofeedback on respiratory-gated radiotherapy. Int J Radiat Oncol Biol Phys 65(3):924-33 (2006)   DOI   ScienceOn
13 Nelms BE, Simon JA: A survey on planar IMRT QA analysis. J Appl Clin Med Phys 8(3):2448 (2007)