A Comparison Study of Volumetric Modulated Arc Therapy Quality Assurances Using Portal Dosimetry and MapCHECK 2  

Jin, Hosang (Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center)
Jesseph, Fredrick B. (Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center)
Ahmad, Salahuddin (Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center)
Publication Information
Progress in Medical Physics / v.25, no.2, 2014 , pp. 65-71 More about this Journal
Abstract
A Varian Portal Dosimetry system was compared to an isocentrically mounted MapCHECK 2 diode array for volumetric modulated arc therapy (VMAT) QA. A Varian TrueBeam STx with an aS-1000 digital imaging panel was used to acquire VMAT QA images for 13 plans using four photon energies (6, 8, 10 and 15 MV). The EPID-based QA images were compared to the Portal Dose Image Prediction calculated in the Varian Eclipse treatment planning system (TPS). An isocentrically mounted Sun Nuclear MapCHECK 2 diode array with 5 cm water-equivalent buildup was also used for the VMAT QAs and the measurements were compared to a composite dose plane from the Eclipse TPS. A ${\gamma}$ test was implemented in the Sun Nuclear Patient software with 10% threshold and absolute comparison at 1%/1 mm (dose difference/distance-to-agreement), 2%/2 mm, and 3%/3 mm criteria for both QA methods. The two-tailed paired Student's t-test was employed to analyze the statistical significance at 95% confidence level. The average ${\gamma}$ passing rates were greater than 95% at 3%/3 mm using both methods for all four energies. The differences in the average passing rates between the two methods were within 1.7% and 1.6% of each other when analyzed at 2%/2 mm and 3%/3 mm, respectively. The EPID passing rates were somewhat better than the MapCHECK 2 when analyzed at 1%/1 mm; the difference was lower for 8 MV and 10 MV. However, the differences were not statistically significant for all criteria and energies (p-values >0.05). The EPID-based QA showed large off-axis over-response and dependence of ${\gamma}$ passing rate on energy, while the MapCHECK 2 was susceptible to the MLC tongue-and-groove effect. The two fluence-based QA techniques can be an alternative tool of VMAT QA to each other, if the limitations of each QA method (mechanical sag, detector response, and detector alignment) are carefully considered.
Keywords
Pretreatment QA; VMAT; EPID; Portal dosimetry; MapCHECK 2;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Jursinic PA, Sharma R, Reuter J: MapCHECK used for rotational IMRT measurements: step-and-shoot, TomoTherapy, RapidArc. Med Phys 37:2837-2846 (2010)   DOI
2 Rinaldin G, Perna L, Agnello G, et al: Quality assurance of RapidArc treatments: Performances and pre-clinical verifications of a planar detector (MapCHECK2). Phys Medica 30:184-190 (2014)   DOI
3 Shim SJ, Shim JB, Lee SH, et al: Quality assurance of volumetric modulated arc therapy for elekta synergy. Korean J Med Phys 23:33-41 (2012)
4 Varian Medical Systems: TrueBeamTM TrueBeam STxTM Technical reference guide (Version 1.6 Volume 2: Imaging; Docu ID: B501671R01, Revision D). Varian Medical Systems, Inc., Switzerland, (2011), pp.205-207
5 Varian Medical Systems: Eclipse algorithms reference guide (P/N B503486R01B, document version 1.0). Varian Medical Systems, Inc., Finland, (2011), pp.393-423
6 Low DA, Harms WB, Mutic S, Purdy JA: A technique for the quantitative evaluation of dose distributions. Med Phys 25:656-661 (1998)   DOI   ScienceOn
7 Stock M, Kroupa B, Georg D: Interpretation and evaluation of the gamma index and the gamma index angle for the verification of IMRT hybrid plans. Phys Med Biol 50:399-411 (2005)   DOI
8 Sweet J, Nicolaou N: Arc splitting for VMAT patient QA [abstract]. Med Phys 39:3746 (2012)
9 Jin H, Keeling VP, Johnson DA, Ahmad S: Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance. J Appl Clin Med Phys 15:1-13 (2014)
10 Greer PB: Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry. Med Phys 32:3558-3568 (2005)   DOI   ScienceOn
11 Greer PB: Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device. Med Phys 34: 3815-3824 (2007)   DOI   ScienceOn
12 Deng J, Pawlicki T, Chen Y, Li J, Jiang SB, Ma C-M: The MLC tongue-and-groove effect on IMRT dose distributions. Phys Med Biol 46:1039-1060 (2001)   DOI
13 Olch AJ: Evaluation of the accuracy of 3DVH software estimates of dose to virtual ion chamber and film in composite IMRT QA. Med Phys 39:81-86 (2012)   DOI
14 Bailey DW, Kumaraswamy L, Bakhtiari M, Malhotra HK, Podgorsak MB: EPID dosimetry for pretreatment quality assurance with two commercial systems. J Appl Clin Med Phys 13:82-99 (2012)
15 Bakhtiari M, Kumaraswamy L, Bailey DW, de Boer S, Malhotra HK, Podgorsak MB: Using an EPID for patient- specific VMAT quality assurance. Med Phys 38:1366-1373 (2011)   DOI
16 Sharma DS, Mhatre V, Heigrujam M, Talapatra K, Mallik S: Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array. J Appl Clin Med Phys 11:238-248 (2010)
17 Clemente S, Caivano R, Cozzolino M, et al: To evaluate the accuracy of dynamic versus static IMRT delivery using portal dosimetry. Clin Transl Oncol 16:208-212 (2014)   DOI
18 Bailey DW, Kumaraswamy L, Bakhtiari M, Podgorsak MB: A two-dimensional matrix correction for off-axis portal dose prediction errors. Med Phys 40:051704 (2013)   DOI
19 Bailey DW, Kumaraswamy L, Podgorsak MB: An effective correction algorithm for off-axis portal dosimetry errors. Med Phys 36:4089-4094 (2009)   DOI   ScienceOn
20 Huang Y-C, Yeh C-Y, Yeh J-H, et al: Clinical practice and evaluation of electronic portal imaging device for VMAT quality assurance. Med Dosim 38:35-41 (2013)   DOI
21 Iftimia I, Cirino ET, Xiong L, Mower HW: Quality assurance methodology for Varian RapidArc treatment plans. J Appl Clin Med Phys 11:130-143 (2010)   DOI
22 Nicolini G, Vanetti E, Clivio A, et al: The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol 3:24 (2008)   DOI
23 Rowshanfarzad P, Sabet M, O'Connor DJ, McCowan PM, McCurdy BM, Greer PB: Detection and correction for EPID and gantry sag during arc delivery using cine EPID imaging. Med Phys 39:623-635 (2012)   DOI
24 Gloi AM, Buchanan RE, Zuge CL, Goettler AM: RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys 12:39-47 (2011)   DOI