The Output Factor of Small Field in Multileaf Collimator of 6 MV Photon Beams  

Lee, Ho Joon (Department of Radiation Oncology, Daegu Catholic University Medical Center)
Choi, Tae-Jin (Department of Radiation Oncology, Daegu Catholic University Medical Center)
Oh, Young Kee (Department of Radiation Oncology, Keimyung University Dongsan Medical Center)
Jeun, Kyung Soo (Department of Radiation Oncology, Keimyung University Dongsan Medical Center)
Lee, Yong Hee (Department of Radiation Oncology, Keimyung University Dongsan Medical Center)
Kim, Jin Hee (Department of Radiation Oncology, Keimyung University Dongsan Medical Center)
Kim, Ok Bae (Department of Radiation Oncology, Keimyung University Dongsan Medical Center)
Oh, Se An (Department of Radiation Oncology, Youngnam University Medical Center)
Kim, Sung Kyu (Department of Radiation Oncology, Youngnam University Medical Center)
Ye, Ji Woon (Department of Radiation Oncology, Youngnam University Medical Center)
Publication Information
Progress in Medical Physics / v.25, no.1, 2014 , pp. 15-22 More about this Journal
Abstract
The IMRT is proper implement to get high dose deliver to tumor as its shape and selective approach in radiation therapy. Since the IMRT is performed as modulated the radiation fluence by the MLC created the open shapes and its irradiation time, the dose of segment of radiation field effects on the cumulated portal dose. The accurate output factor of small and step shape of segment is important to improve the determination of deliver tumor dose as it is directly proportional to dose. This experiment performed with the 6 MV photon beam of Clinac Ex(Varian) from $3{\times}3cm^2$ to $0.5{\times}0.5cm^2$ small field size for collimator jaw in MLC free and/or for MLC open field in fixed collimator jaw $10{\times}10cm^2$ using the CC01 ion chamber, SFD diode, diamond detector and X-Omat film dosimetry. As results of normalized to the reference field of $10{\times}10cm^2$ of MLC, the output factor of $3{\times}3cm^2$ showed $0.899{\pm}0.0106$, $0.855{\pm}0.0106$ for $2{\times}2cm^2$, $0.764{\pm}0.0082$ for $1{\times}1cm^2$ and $0.602{\pm}0.0399$ for $0.5{\times}0.5cm^2$. The output factor of MLC open field has shown a maximum 3.8% higher than that of the collimator jaw open field.
Keywords
Small field; Output factor; Film dosimetry;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Duggan DM and Coffey II CW: Small Photon Field Dosimetry for Stereotactic Radiosurgery. Medical Dosimetry 23(3):153-159 (1998)   DOI   ScienceOn
2 Das IJ, Ding GX, Ahnesjo A: Small fields: Nonequilibrium radiation dosimetry. Medical Physics 35:206-215 (2008)   DOI   ScienceOn
3 Gavin CS, Steve W, Narinder PS, et al: Experimental small field 6 MV output ratio analysis for various diode detector and accelerator combinations Radiotherapy and Oncology 100: 429-435 (2011)   DOI   ScienceOn
4 Cadman P, Sidhul NPS, Ibbott G, Nelson A: Dosimetric considerations for validation of sequential IMRT process with a commercial treatment planning system. Phys Med Biol 47: 3001-3010 (2002)   DOI   ScienceOn
5 Thomas SJ, Eaton DJ, et al: Equivalent squares for small field dosimetry. The British Journal of Radiology. 81:897-901 (2008)
6 Choi TJ, Kim JH, Kim OB, et al: Development of a New Radiotherapy Technique using the Quasi-conformation Method. J Korean Soc Ther Radiol 9(2):343-350 (1991)
7 Kenichiro Hasumi, Yukimasa Aoki, Ryuko Wantanabe, Dean L Mann: Clinical respondse of advanced cancer patients to cellular immunotherapy and intensity-modulated radiation therapy. Oncoimmunology 2(10) (2013)
8 Alfonso R, Andreo P, Capote R, et al: A new formalism for reference dosimetry of small and nonstandard fields. Medical Physics 35:5179-5186 (2008)   DOI   ScienceOn
9 Francescon P, Cora S, Cavedon C: Total scatter factors of small beams: A multidetector and Monte Carlo study. Medical Physics 30:341-347 (2003)   DOI   ScienceOn
10 Oh SA, Kang MK, et al: Study of the Penumbra for High-energy Photon Beams with GafchromaticTM EBT2. Journal of the Korean Physical Society 60(11):1973-1976 (2012)   DOI   ScienceOn
11 Varian medical system. Eclipse Algorithms Reference Guide. p43-45 (2009)
12 Francisco SD, Günther HH, Javier P, et al: Uncertainty estimation in intensity-modulated radiotherapy absolute dosimetry verification. Int J Radiation Oncology Biol Phys 68(1): 301-310 (2007)   DOI
13 Alfonso R, Andreo P, Capote R, et al: A new formalism for reference dosimetry of small and nonstandard fields. Medical Physics 35:5179-5186 (2008)   DOI   ScienceOn
14 Das IJ, Ding GX, Ahnesjo A: Small fields: Nonequilibrium radiation dosimetry. Medical Physics 35:206-215 (2008)   DOI   ScienceOn
15 Klein EE, Hanley J, Bayouth J, et al: Task Group 142 report: Quality assurance of medical accelerators. Med Phys 36(9): 4197-4212 (2009)   DOI   ScienceOn
16 Barbosa NA, Rosa LAR, et al: Development of a phantom for dose distribution verification in Stereotactic Radiosurgery. Physica Medica 29:461-469 (2013)   DOI   ScienceOn
17 Followill DS, Kry SF, Qin L, et al: The Radiological Physics Center's standard dataset for small field size output factors. Journal of Applied Clinical Medical Physics 13(5): 282- 289 (2012)   DOI
18 Klein DM, Tailor RC. Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors. Med Phys 37(10):5541-5549 (2010)   DOI   ScienceOn
19 Kim JO, Siebers JV, Keall PJ, et al: A Monte Carlo study of radiation transport through multileaf collimators. Med Phys 28(12):2497-2506 (2001)   DOI   ScienceOn
20 Scott AJD, Nahum AE, Fenwick JD: Using a Monte Carlo model to predict dosimetric properties of small radiotherapy photon fields. Medical Physics 35:4671-4684 (2008)   DOI