Analysis of Acquisition Parameters That Caused Artifacts in Four-dimensional (4D) CT Images of Targets Undergoing Regular Motion  

Sheen, Heesoon (Boi-Medical Engineering, Medical School Sungkyunkwan University)
Han, Youngyih (Boi-Medical Engineering, Medical School Sungkyunkwan University)
Shin, Eunhyuk (Department of Radiation Oncology, Samsung Medical Center)
Publication Information
Progress in Medical Physics / v.24, no.4, 2013 , pp. 243-252 More about this Journal
Abstract
The aim of this study was to clarify the impacts of acquisition parameters on artifacts in four-dimensional computed tomography (4D CT) images, such as the partial volume effect (PVE), partial projection effect (PPE), and mis-matching of initial motion phases between adjacent beds (MMimph) in cine mode scanning. A thoracic phantom and two cylindrical phantoms (2 cm diameter and heights of 0.5 cm for No.1 and 10 cm for No.2) were scanned using 4D CT. For the thoracic phantom, acquisition was started automatically in the first scan with 5 sec and 8 sec of gantry rotation, thereby allowing a different phase at the initial projection of each bed. In the second scan, the initial projection at each bed was manually synchronized with the inhalation phase to minimize the MMimph. The third scan was intentionally un-synchronized with the inhalation phase. In the cylindrical phantom scan, one bed (2 cm) and three beds (6 cm) were used for 2 and 6 sec motion periods. Measured target volume to true volume ratios (MsTrueV) were computed. The relationships among MMimph, MsTrueV, and velocity were investigated. In the thoracic phantom, shorter gantry rotation provided more precise volume and was highly correlated with velocity when MMimph was minimal. MMimph reduced the correlation. For moving cylinder No. 1, MsTrueV was correlated with velocity, but the larger MMimph for 2 sec of motion removed the correlation. The volume of No. 2 was similar to the static volume due to the small PVE, PPE, and MMimph. Smaller target velocity and faster gantry rotation resulted in a more accurate volume description. The MMimph was the main parameter weakening the correlation between MsTrueV and velocity. Without reducing the MMimph, controlling target velocity and gantry rotation will not guarantee accurate image presentation given current 4D CT technology.
Keywords
4-dimensional computed tomography (4D CT); Partial volume effect (PVE); Partial projection effect (PPE); Mis-matching of initial motion phases between adjacent beds (MMimph);
Citations & Related Records
연도 인용수 순위
  • Reference
1 Wurstbauer K, Deutschmann H, Kopp P, Sedlmayer F: Radiotherapy planning for lung cancer: slow CTs allow the drawing of tighter margins. Radiother Oncol 75(2):165-170 (2005)   DOI   ScienceOn
2 Lagerwaard FJ, Van Sornsen de Koste JR, Nijssen-Visser MR, et al: Multiple "slow" CT scans for incorporating lung tumor mobility in radiotherapy planning. Int J Radiat Oncol Biol Phys 51(4):932-937 (2001)   DOI   ScienceOn
3 Vedam SS, Keall PJ, Kini VR, Mostafavi H, Shukla HP, Mohan R: Acquiring a four-dimensional computed tomography dataset using an external respiratory signal. Phys Med Biol 48(1):45-62 (2003)   DOI   ScienceOn
4 Keall PJ: 4-dimensional computed tomography imaging and treatment planning. Semin Radiat Oncol 14(1):81-90 (2004)   DOI   ScienceOn
5 Keall PJ, Starkschall G, Shukla H, et al: Acquiring 4D thoracic CT scans using a multislice helical method. Phys Med Biol 49(10):2053-2067 (2004)   DOI   ScienceOn
6 Mori S, Endo M, Asakura H: Improvement in banding artifacts in four-dimensional computed tomography for radiotherapy planning. Phys Med Biol 51(20): 5231-5244 (2006)   DOI   ScienceOn
7 Nakamura M, Narita Y, Sawada A, et al: Impact of motion velocity on four-dimensional target volumes: a phantom study. Med Phys 36(5):1610-1617 (2009)   DOI   ScienceOn
8 Pan T: Comparison of helical and cine acquisitions for 4D-CT imaging with multislice CT. Med Phys 32(2):627-634 (2005)   DOI   ScienceOn
9 Rietzel E, Pan T, Chen GT: Four-dimensional computed tomography: image formation and clinical protocol. Med Phys 32(4):874-889 (2005)   DOI   ScienceOn
10 Lecchi M, Fossati P, Elisei F, Orecchia R, Lucignani G: Current concepts on imaging in radiotherapy. Eur J Nucl Med Mol Imaging 35(4): 821-837 (2008)   DOI
11 Wong JW, Sharpe MB, Jaffray DA, et al: The use of active breathing control (ABC) to reduce margin for breathing motion. Int J Radiat Oncol Biol Physics 44:911-919 (1999)   DOI   ScienceOn
12 Hanley J, Debois NM, Mah D, et al: Deep inspiration Breath-hold technique for lung tumors: the potential value of target immobilization and reduced lung density in dose escalation: Int J Radiat Oncol Biol Physics 45:603-611 (1999)   DOI   ScienceOn
13 Shin EH, Park HC, Han YY, et al: Efficacy of a respiratory training system on the regularity of breathing. J. Korean Soc Ther Radiol 26(3):181-188 (2008)   DOI
14 Minohara S, Endo M, Kanai T, Kato H, Tsujii H: Estimating uncertainties of the geometrical range of particle radiotherapy during respiration. Int J Radiat Oncol Biol Phys 56(1):121-125 (2003)   DOI   ScienceOn