Browse > Article

Evaluation of Setup Uncertainty on the CTV Dose and Setup Margin Using Monte Carlo Simulation  

Cho, Il-Sung (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Kwark, Jung-Won (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Cho, Byung-Chul (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Kim, Jong-Hoon (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Ahn, Seung-Do (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Park, Sung-Ho (Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine)
Publication Information
Progress in Medical Physics / v.23, no.2, 2012 , pp. 81-90 More about this Journal
Abstract
The effect of setup uncertainties on CTV dose and the correlation between setup uncertainties and setup margin were evaluated by Monte Carlo based numerical simulation. Patient specific information of IMRT treatment plan for rectal cancer designed on the VARIAN Eclipse planning system was utilized for the Monte Carlo simulation program including the planned dose distribution and tumor volume information of a rectal cancer patient. The simulation program was developed for the purpose of the study on Linux environment using open source packages, GNU C++ and ROOT data analysis framework. All misalignments of patient setup were assumed to follow the central limit theorem. Thus systematic and random errors were generated according to the gaussian statistics with a given standard deviation as simulation input parameter. After the setup error simulations, the change of dose in CTV volume was analyzed with the simulation result. In order to verify the conventional margin recipe, the correlation between setup error and setup margin was compared with the margin formula developed on three dimensional conformal radiation therapy. The simulation was performed total 2,000 times for each simulation input of systematic and random errors independently. The size of standard deviation for generating patient setup errors was changed from 1 mm to 10 mm with 1 mm step. In case for the systematic error the minimum dose on CTV $D_{min}^{stat{\cdot}}$ was decreased from 100.4 to 72.50% and the mean dose $\bar{D}_{syst{\cdot}}$ was decreased from 100.45% to 97.88%. However the standard deviation of dose distribution in CTV volume was increased from 0.02% to 3.33%. The effect of random error gave the same result of a reduction of mean and minimum dose to CTV volume. It was found that the minimum dose on CTV volume $D_{min}^{rand{\cdot}}$ was reduced from 100.45% to 94.80% and the mean dose to CTV $\bar{D}_{rand{\cdot}}$ was decreased from 100.46% to 97.87%. Like systematic error, the standard deviation of CTV dose ${\Delta}D_{rand}$ was increased from 0.01% to 0.63%. After calculating a size of margin for each systematic and random error the "population ratio" was introduced and applied to verify margin recipe. It was found that the conventional margin formula satisfy margin object on IMRT treatment for rectal cancer. It is considered that the developed Monte-carlo based simulation program might be useful to study for patient setup error and dose coverage in CTV volume due to variations of margin size and setup error.
Keywords
IMRT; Setup uncertainty; Setup margin; Monte Carlo simulation; Treatment planning;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Bijhold J, Lebesque JV, Hart AA, et al: Maximizing setup accuracy using portal images as applied to a conformal boost technique for prostatic cancer. Radiother Oncol 24:261-271 (1992)   DOI   ScienceOn
2 Bel A, van Herk M, Bartelink H, et al: A verification procedure to improve patient set-up accuracy using portal images, Radiother Oncol 29(2):253-260 (1993)   DOI   ScienceOn
3 de Boer HC, Heijmen BJ: A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload. Int J Radiat Oncol Biol Phys 50(5):1350-1365 (2001)   DOI   ScienceOn
4 Van Herk M: Errors and Margins in Radiotherapy. Seminars in Radiation Oncol 14(1):52-64 (2004)
5 Mageras GS, Kutcher GJ, Leibel SA, et al: A method of incorporating organ motion uncertainties into three-dimensional conformal treatment plans. Int J Radiat Oncol Biol Phys 35(2): 333-342 (1996)   DOI   ScienceOn
6 Killoran JH, Kooy HM, Gladstone DJ, et al: A numerical simulation of organ motion and daily setup uncertainties: implications for radiation therapy. Int J Radiat Oncol Biol Phys 37(1):213-221 (1997)   DOI   ScienceOn
7 Mageras GS, Fuks Z, Leibel SA, et al: Computerized design of target margins for treatment uncertainties in conformal radiotherapy. Int J Radiat Oncol Biol Phys 43(2):437-445 (1999)   DOI   ScienceOn
8 Chen AM, Farwell DG, Luu Q, et al: Evaluation of the planning target volume in the treatment of head and neck cancer with intensity-modulated radiotherapy: what is the appropriate expansion margin in the setting of daily image guidance? Int J Radiat Oncol Biol Phys 81(4):943-949 (2011)   DOI   ScienceOn
9 Huang E, Dong L, Chandra A, et al: Intrafraction prostate motion during IMRT for prostate cancer. Int J Radiat Oncol Biol Phys 53:261-268 (2002)   DOI   ScienceOn
10 Britton KR, Takai Y, Mitsuya M, et al: Evaluation of internd intrafraction organ motion during intensity modulated radiation therapy (IMRT) for localized prostate cancer measured by a newly developed on-board image-guided system. Radiat Med 23(1):14-24 (2005)
11 Samuelian JM, Callister MD, Ashman JB, et al: Reduced acute bowel toxicity in patients treated with intensitymodulated radiotherapy for rectal cancer. Int J Radiat Oncol Biol Phys 82(5):1981-1987 (2012)   DOI   ScienceOn
12 Duthoy W, De Gersem W, Vergote K, et al: Clinical implementation of intensity-modulated arc therapy (IMAT) for rectal cancer. Int J Radiat Oncol Biol Phys 60(3):794-806 (2004)   DOI   ScienceOn
13 Guerrero Urbano MT, Henrys AJ, Adams EJ, et al: Intensity-modulated radiotherapy in patients with locally advanced rectal cancer reduces volume of bowel treated to high dose levels. Int J Radiat Oncol Biol Phys 65(3):907-916 (2006)   DOI   ScienceOn
14 RTOG 0529: A Phase II Evaluation of Dose-Painted IMRT in Combination with 5-Fluorouracil and Mitomycin-C for Reduction of Acute Morbidity in Carcinoma of the Anal Canal
15 ROOT: An Object Oriented Data Analysis Framework, http://root.cern.ch
16 Matsumoto M, Nishimura T: Mersenne Twistor: A 623-diminsionally equidistributed uniform pseudorandom number generator, ACM Trans. Model Comput Simul 8:3-30 (1998)   DOI   ScienceOn
17 Van Herk M, Remeijer P, Rasch C, et al: The Probability of correct target dosage: Dose-Population histograms for deriving treatment margins in radiotherapy. Int J Radiation Oncology Biol Phys 47:1121-1135 (2000)   DOI   ScienceOn