Dosimetric Effect on Selectable Optimization Parameters of Volumatric Modulated Arc Therapy  

Jung, Jae-Yong (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine)
Shin, Yong-Joo (Department of Radiation Oncology, Sanggye Paik Hospital, Inje University)
Sohn, Seung-Chang (Department of Radiation Oncology, Sanggye Paik Hospital, Inje University)
Kim, Yeon-Rae (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine)
Min, Jung-Wan (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine)
Suh, Tae-Suk (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine)
Publication Information
Progress in Medical Physics / v.23, no.1, 2012 , pp. 15-25 More about this Journal
Abstract
The aim of this study is to evaluate plan quality and dose accuracy for Volumetric Modulated Arc Therapy (VMAT) on the TG-119 and is to investigate the effects on variation of the selectable optimization parameters of VMAT. VMAT treatment planning was implemented on a Varian iX linear accelerator with ARIA record and verify system (Varian Mecical System Palo Alto, CA) and Oncentra MasterPlan treatment planning system (Nucletron BV, Veenendaal, Netherlands). Plan quality and dosimetric accuracy were evaluated by effect of varying a number of arc, gantry spacing and delivery time for the test geometries provided in TG-119. Plan quality for the target and OAR was evaluated by the mean value and the standard deviation of the Dose Volume Histograms (DVHs). The ionization chamber and $Delta^{4PT}$ bi-planar diode array were used for the dose evaluation. For treatment planning evaluation, all structure sets closed to the goals in the case of single arc, except for the C-shape (hard), and all structure sets achieved the goals in the case of dual arc, except for C-shape (hard). For the variation of a number of arc, the simple structure such as a prostate did not have the difference between single arc and dual arc, whereas the complex structure such as a head and neck showed a superior result in the case of dual arc. The dose distribution with gantry spacing of $4^{\circ}$ was shown better plan quality than the gantry spacing of $6^{\circ}$, but was similar results compared with gantry spacing of $2^{\circ}$. For the verification of dose accuracy with single arc and dual arc, the mean value of a relative error between measured and calculated value were within 3% and 4% for point dose and confidence limit values, respectively. For the verification on dose accuracy with the gantry intervals of $2^{\circ}$, $4^{\circ}$ and $6^{\circ}$, the mean values of relative error were within 3% and 5% for point dose and confidence limit values, respectively. In the verification of dose distribution with $Delta^{4PT}$ bi-planar diode array, gamma passing rate was $98.72{\pm}1.52%$ and $98.3{\pm}1.5%$ for single arc and dual arc, respectively. The confidence limit values were within 4%. The smaller the gantry spacing, the more accuracy results were shown. In this study, we performed the VMAT QA based on TG-119 procedure, and demonstrated that all structure sets were satisfied with acceptance criteria. And also, the results for the selective optimization variables informed the importance of selection for the suitable variables according to the clinical cases.
Keywords
Volumetric Modulated Arc Therapy (VMAT); Selectable optimization parameters; QA (Quality Assurance);
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Otto K: Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 35:310-317 (2008)   DOI   ScienceOn
2 Haga A, Nakagawa K, Shiraishi K, et al: Quality assurance of volumetric modulated arc therapy using Elekta Synergy. Acta Oncol 48:1193-1197 (2009)   DOI   ScienceOn
3 Ling CC, Zhang P, Archambault M, et al: Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys 72:575-581 (2008)   DOI   ScienceOn
4 Rao M, Yang W, Chen F, et al: Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: Plan quality, delivery efficiency and accuracy. Med Phy 37:1350-1359 (2010)   DOI   ScienceOn
5 Chen Q, Chen M, Lu W: Ultrafast convolution/superposition using tabulated and exponential kernels on GPU. Med Phys 38:1150-1161(2010)
6 Ezzell GA, Burmeister JW, Dogan N, et al: IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 36: 5359-5373 (2009)   DOI   ScienceOn
7 Nucletron: Oncentra MasterPlan Physics and Algorithms documentation, Nucletron, Netherlands (2011), pp. 66-72
8 Scandidos: $Delta^{4PT}$ Getting Started user guide. Scandidos AB, Sweden (2010), pp. 31-57
9 Low DA, Harms SB, Mutic S, Purdy JA: A technique for the quantitative evalution of dose distributions. Med Phys 25: 656-661 (1998)   DOI   ScienceOn
10 Feygelman V, Zhang G, Stevens C: Initial dosimetric evaluation of SmartArc - a novel VMAT treatment planning module implemented in a multi-vendor delivery chain. J Appl Clin Med Phys 11:99-116 (2010)   DOI
11 Saminathan S, Manickam R, Chandraraj V: Plan evaluation and dosimetric comparison of IMRT using AAPM TG119 test suites and recommendations. Australas Phys Eng Sci Med 34:55-61 (2011)   DOI   ScienceOn
12 안우상, 조병철: 세기조절방사선치료(IMRT)의 Commissioning 및 정도관리:AAPM TG 119 적용. 의학물리 22:99-105 (2011)
13 ESTRO Booklet no. 9: Guidelines for the verification of IMRT. ESTRO, Belgium (2008)
14 Nucletron: Oncentra MasterPlan Physics and Algorithms documentation. Nucletron, Netherlands (2011), pp. 86-95