Browse > Article

A Method to Calculate a Pass Rate of the ${\gamma}$-index Analysis in Tomotherapy Delivery Quality Assurance (DQA)  

Park, Dahl (Department of Radiation Oncology, Pusan National University Hospital)
Kim, Yong-Ho (Department of Radiation Oncology, Pusan National University Hospital)
Kim, Won-Taek (Department of Radiation Oncology, Pusan National University School of Medicine)
Kim, Dong-Won (Department of Radiation Oncology, Pusan National University School of Medicine)
Kim, Dong-Hyun (Department of Radiation Oncology, Pusan National University Hospital)
Jeon, Ho-Sang (Department of Radiation Oncology, Pusan National University Yangsan Hospital)
Nam, Ji-Ho (Department of Radiation Oncology, Pusan National University Yangsan Hospital)
Lim, Sang-Wook (Department of Radiation Oncology, Kosin University College of Medicine)
Publication Information
Progress in Medical Physics / v.21, no.4, 2010 , pp. 340-347 More about this Journal
Abstract
DQA, a patient specific quality assurance in tomotherapy, is usually performed using an ion chamber and a film. The result of DQA is analysed with the treatment planning system called Tomo Planning Station (TomoPS). The two-dimensional dose distribution of film measurement is compared with the dose distribution calculated by TomoPS using the ${\gamma}$-index analysis. In ${\gamma}$-index analysis, the criteria such as 3%/3 mm is used and we verify that whether the rate of number of points which pass the criteria (pass rate) is within tolerance. TomoPS does not provide any quantitative information regarding the pass rate. In this work, a method to get the pass rate of the ${\gamma}$-index analysis was suggested and a software PassRT which calculates the pass rate was developed. The results of patient specific QA of the intensity modulated radiation therapy measured with I'mRT MatriXX (IBA Dosimetry, Germany) and DQA of tomotherapy measured with film were used to verify the proposed method. The pass rate was calculated using PassRT and compared with the pass rate calculated by OmniPro I'mRT (IBA Dosimetry, Germany). The average difference between the two pass rates was 0.00% for the MatriXX measurement. The standard deviation and the maximum difference were 0.02% and 0.02%, respectively. For the film measurement, average difference, standard deviation and maximum difference were 0.00%, 0.02% and 0.02%, respectively. For regions of interest smaller than $24.3{\times}16.6cm^2$ the proposed method can be used to calculate the pass rate of the gamma index analysis to one decimal place and will be helpful for the more accurate DQA in tomotherapy.
Keywords
TomoTherapy$^{(R)}$; Patient specific quality assurance; ${\gamma}$-index;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Van VM, Field C, Raaijmakers CP: Comparing step-andshoot IMRT with dynamic helical tomotherapy IMRT plans for head-and-neck cancer. Int J Radiat Oncol Biol Phys 62:1535-1539 (2005)   DOI   ScienceOn
2 김동욱, 윤명근, 박성용 등: 두경부암 환자에 대한 선량체적 히스토그램에 따른 토모치료외 선형가속기기반 세기변조방사선 치료의 정량적 비교. 의학물리 19:89-94 (2008)
3 Chen YJ, Liu A, Han C, et al: Helical tomotherapy for radiotherapy in esophageal cancer: a preferred plan with better conformal target coverage and more homogeneous dose distribution. Med Dosim 32:166-171 (2007)   DOI   ScienceOn
4 Ramsey C, Seibert R, Scaperoth D, et al: Image-guided helical tomotherapy for localized prostate cancer: Technique and initial clinical observations. J Appl Clinc Med Phys 8:37-51 (2007)
5 Mavroidis P, Stathakis S, Gutierrez A, et al: Expected clinical impact of the differences between planned and delivered dose distributions in Helical Tomotherapy for treating head and neck cancer using Helical Megavoltage CT Images. J Appl Clinc Med Phys 10:125-139 (2009)   DOI
6 Low DA, Harms WB, Mutic S, Purdy JA: A technique for the quantitative evaluation of dose distributions. Med Phys 25: 656-661 (1998)   DOI   ScienceOn
7 TomoTherapy: Hi-Art Treatment System User Guide - Delivery Quality Assurance Version 3.x.xl
8 Amerio S, Boriano A, Bourhaleb F, et al: Dosimetric characterization of a large area pixel-segmented ionization chamber. Med Phys 31:414-420 (2004)   DOI   ScienceOn
9 Saminathan S, Manickan R, Chandraraj V, et al: Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification. J Appl Clinc Med Phys 11: 116-127 (2010)   DOI
10 Microsoft Help and Support: DIBs and Their Uses. (2005-02-11)