Abstract
The shielding materials designed for replacement of lead equivalent materials for lighter apron than that of lead in diagnostic photon beams. The absorption characteristics of elements were applied to investigate the lead free material for design the shielding materials through the 50 kVp to 110 kVp x-ray energy in interval of 20 kVp respectively. The idea focused to the effect of K-edge absorption of variable elements excluding the lead material for weight reduction. The designed shielding materials composited of Tin 34.1%, Antimon 33.8% and Iodine 26.8% and Polyisoprene 5.3% gram weight account for 84 percent of weight of lead equivalent of 0.5 mm thickness. The size of lead-free shielder was $200{\times}200{\times}1.5\;mm^3$ and $3.2\;g/cm^3$ of density which is equivalent to 0.42 mm of Pb. The lead equivalent of 0.5 mm thickness generally used for shielding apron of diagnostic X rays which is transmitted 0.1% for 50 kVp, 0.9% for 70 kVp and 3.2% for 90 kVp and 4.8% for 110 kVp in experimental measurements. The experiment of transmittance for lead-free shielder has showed 0.3% for 50 kVp, 0.6% for 70 kVp, 2.0% for 90 kVp and 4.2% for 110 kVp within ${\pm}0.1%$. respectively. Using the attenuation coefficient of experiments for 0.5 mm Pb equivalent of lead-free materials showed 0.1%. 0.3%, 1.0% and 2.4%, respectively. Furthermore, the transmittance of lead-free shielder for scatter rays has showed the 2.4% in operation energy of 50 kVp and 5.9% in energy of 110 kVp against 2.4% and 5.1% for standard lead thickness within ${\pm}0.2%$ discrepancy, respectively. In this experiment shows the designed lead-free shielder is very effective for reduction the apron weight in diagnostic radiation fields.
Keywords
Radiation protection; Lead-free shielding material; Spectrum of x-ray beam; Apron;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Bushong SC: Radiologic science for technologists. 2nd ed. The C.V. Mosby Company, ST. Louis, Toronto, London (1980), pp. 398-445
2 Redus R: XRF Spectra and Spectra Analysis Software. Amptek Application Note XRF-1, Amptek Inc (2008)
3 NSRDS-NBS29: Photon Cross Sections, Attenuation Coefficients, and Energy Absorption Coefficients From 10 keV to 100 GeV
4 Bureau of Radiological Health: Section III, Radiation Protection Data. Radiological Health HandBook (1970), pp. 129-169
5 최태진, 김옥배, 서수지: 투과선량의 수치해석에 의한 전산화단층영상장치 X선의 에너지분포결정. 의학물리 8:35-43 (1997)
6 최태진, 김진희, 김옥배: 연필터의 투과선량을 이용한 15 MV X선의 에너지스펙트럼 결정과 조직선량 비교. 의학물리 19:80-88 (2008)   과학기술학회마을
7 Johns HE, Cunningham JR: The Physics of Radiology. Diagnostic Radiology. 4th ed. Charles Thomas Publisher, Spring-field. Illinois, U.S.A (1983), pp. 557