Abstract
In this study, we evaluated the performance of a commercial pixelated cadmium zinc telluride (CZT) detector for spectroscopy and identified its feasibility as a Compton camera for radiation monitoring in a nuclear power plant. The detection system consisted of a $20mm{\times}20mm{\times}5mm$ CZT crystal with $8{\times}8$ pixelated anodes and a common cathode, in addition to an application specific integrated circuit. The performance of the various radioisotopes $^{57}Co$, $^{133}Ba$, $^{22}Na$, and $^{137}Cs$ was evaluated. In general, the amplitude of the induced signal in a CZT crystal depends on the interaction position and material non-uniformity. To minimize this dependency, a drift time correction was applied. The depth of each interaction was calculated by the drift time and the positional dependency of the signal amplitude was corrected based on the depth information. After the correction, the Compton regions of each spectrum were reduced, and energy resolutions of 122 keV, 356 keV, 511 keV, and 662 keV peaks were improved from 13.59%, 9.56%, 6.08%, and 5%-4.61%, 2.94%, 2.08%, and 2.2%, respectively. For the Compton imaging, simulations and experiments using one $^{137}Cs$ source with various angular positions and two $^{137}Cs$ sources were performed. Individual and multiple sources of $^{133}Ba$, $^{22}Na$, and $^{137}Cs$ were also measured. The images were successfully reconstructed by weighted list-mode maximum likelihood expectation maximization method. The angular resolutions and intrinsic efficiency of the $^{137}Cs$ experiments were approximately $7^{\circ}-9^{\circ}$ and $5{\times}10^{-4}-7{\times}10^{-4}$, respectively. The distortions of the source distribution were proportional to the offset angle.
Keywords
Pixelated CZT detector; Spectroscopy; Drift time method; Compton camera;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 S.D. Sordo, L. Abbene, E. Caroli, A.M. Mancini, A. Zappettini, P. Ubertini, Progress in the development of CdTe and CdZnTe semiconductor radiation detectors for astrophysical and medical applications, Sensors 9 (2009) 3491-3526.   DOI
2 W. Jo, M. Jeong, H. Kim, S. Kim, J. Ha, Preliminary Research of CZT based PET system development in KAERI, J. Radiat. Prot. Res. 41 (2016) 81-86.   DOI
3 L. Cirignano, H. Kim, K. Shah, M. Klugerman, P. Wong, M. Squillante, L. Li, Evaluation of CZT detectors with capacitive Frisch grid structure, Proc. SPIE 5198 (2004) 1-8.
4 Z. He, W. Li, G.F. Knoll, D.K. Wehe, J. Berry, C.M. Stahle, 3-D position sensitive CdZnTe gamma-ray spectrometers, Nucl. Instrum. Methods A422 (1999) 173-178.
5 P.N. Luke, Unipolar charge sensing with coplanar electrodes - application to semiconductor detectors, IEEE Trans. Nucl. Sci. 42 (1995) 207-213.   DOI
6 D.S. McGregor, Z. He, H.A. Seifert, D.K. Wehe, R.A. Rojeski, Single charge carrier type sensing with a parallel strip pseudo-Frisch-grid CdZnTe semiconductor radiation detector, Appl. Phys. Lett. 72 (1998) 792-794.   DOI
7 D. Xu, Z. He, C.E. Lehner, F. Zhang, $4{\pi}$ Compton imaging with single 3D position sensitive CdZnTe detector, Proc. SPIE 5540 (2004) 114-115.
8 C.E. Lehner, Z. He, F. Zhang, $4{\pi}$ Compton imaging using a 3-D position-sensitive CdZnTe detector via weighted list-mode maximum likelihood, IEEE Trans. Nucl. Sci. 51 (2004) 1618-1624.   DOI
9 C. Wahl, W.R. Kaye, W. Wang, F. Zhang, J.M. Jaworski, A. King, A.Y. Boucher, Z. He, The Polaris-H imaging spectrometer, Nucl. Instrum. Methods Phys. Res. A Accel. Spectrom. Detect. Assoc. Equip. 784 (2015) 377-381.   DOI
10 A.E. Bolotnikov, K. Ackley, G.S. Camarda, C. Cherches, Y. Cui, G. De Geronimo, J. Fried, D. Hodges, A. Hossain, W. Lee, G. Mahler, M. Maritato, M. Petryk, U. Roy, C. Salwen, E. Vernon, G. Yang, R.B. James, An array of virtual Frisch-grid CdZnTe detectors and a front-end application-specific integrated circuit for large-area position-sensitive gamma-ray cameras, Rev. Sci. Instrum. 86 (2015) 1-10.
11 W. Lee, A.E. Bolotnikov, T. Lee, G.S. Camarda, T. Cui, R. Gul, A. Hossain, Y. Roy, G. Yang, R.B. James, Mini Compton camera based on an array of Frisch-grid CdZnTe detecttors, IEEE Trans. Nucl. Sci. 63 (1) (2016) 259-265.   DOI
12 F. Zhang, Z. He, D. Xu, G.F. Knoll, D.K. Wehe, J.E. Berry, Improved resolution for 3-D position sensitive CdZnTe spectrometers, IEEE Trans. Nucl. Sci. 51 (2004) 2427-2431.   DOI
13 F. Zhang, Z. He, C. Seifert, A prototype three-dimensional position sensitive CdZnTe detector array, IEEE Trans. Nucl. Sci. 54 (2007) 843-848.   DOI
14 R. Cesareo, A.L. Hanson, G.E. Gigante, L.J. Pedraza, S.Q.G. Mahtaboally, Interaction of keV photons with matter and new applications, Phys. Rep. 213 (1992) 117-178.   DOI
15 H.H. Barrett, T. White, L.C. Parra, List-mode likelihood, J. Opt. Soc. Am. A 14 (1997) 2914-2923.   DOI
16 S.J. Wilderman, N.H. Clinthorne, J.A. Fessler, W.L. Rogers, List-mode likelihood reconstruction of Compton scatter camera images in nuclear medicine, Proc. IEEE Nucl. Sci. Symp. Conf. Rec. (2000) 1716-1720.
17 J.B. Martin, A Compton Scatter Camera for Spectral Imaging of 0.5 to 3.0 MeV Gamma Rays, Ph. D. dissertation, department of nuclear engineering and radiological sciences, University of Michigan, Ann Arbor, MI, 1988.
18 C. Ordonez, A. Bolozdyna, W. Chang, Doppler broadening of energy spectra in Compton cameras, Proc. 7997 IEEE Nucl. Sci. Symp. 2 (1997) 1361-1365, 1997.
19 F. Biggs, L.B. Mendelsohn, J.B. Mann, Hartree-Fork Compton profiles for the elements, Atomic Data Nucl. Data Tables 16 (1975) 201-309.   DOI
20 C. Kim, S.E. Anderson, W. Kaye, F. Zhang, Y. Zhu, S.J. Kaye, Z. He, Charge sharing in common-grid pixelated CdZnTe detectors, Nucl. Instrum. Methods A 654 (1) (2011) 233-243.   DOI
21 H3D, H100 gamma-ray imaging spectrometer [Online] Available: https://h3dgamma.com/H100Specs.pdf,, 2019.
22 Jiyang Chu, Advanced Imaging Algorithms with Position-Sensitive Gamma-Ray Detectors, Ph. D. dissertation, department of nuclear engineering and radiological sciences, University of Michigan, Ann Arbor, MI, 2018.
23 D. Xu, Gamma-ray Imaging and Polarization Measurement Using 3-D Position-Sensitive CdZnTe Detectors, department of nuclear engineering and radiological sciences, University of Michigan, Ann Arbor, MI, 2006. Ph. D. dissertation.
24 A. Zoglauer, M. Galloway, M. Amman, S.E. Boggs, J.S. Lee, P.N. Luke, L. Mihailescu, K. Vetter, C.B. Wunderer, First results of the high efficiency multi-mode imager (HEMI), in: Nuclear Science Symposium Conference Record, IEEE, 2009.
25 C. Szeles, D. Bale, J. Grosholz, G.L. Smith, M. Blostein, J. Eger, Fabrication of high performance CdZnTe Quasi-hemispherical gamma-ray $CAPture^{TM}$ plus detectors, Hard X-Ray and Gamma-Ray Detector Physics VIII (2006) 6319.
26 A. Dempster, N. Laird, D. Rubin, Maximum likelihood from incomplete data via the EM algorithm, J. R. Stat. Soc. 39 (1977) 1-38.   DOI
27 L.A. Shepp, Y. Vardi, Maximum likelihood reconstruction for emission tomography, IEEE Trans. Med. Imaging MI-2 (1982) 113-122.
28 K. Lange, R. Carson, EM reconstruction algorithms for emission and transmission tomography, J. Comput. Assist. Tomogr. 8 (1984) 306-316.
29 R. Redus, Application Note ANCZT-2 Rev.3 Charge Trapping in XR-100T - CdTe and - CZT Detectors, 2007 [Online] Available:https://amptek.com/pdf/anczt2.pdf.
30 Y. Kim, T. Lee, W. Lee, Double-layered CZT Compton imager, IEEE Trans. Nucl. Sci. 64 (7) (2017) 1769-1773.   DOI
31 G. Montemont, M. Arques, L. Verger, J. Rustique, A capacitive Frisch grid structure for CdZnTe detectors, IEEE Trans. Nucl. Sci. 48 (2001) 278-281.   DOI
32 F. Zhang, C. Herman, Z. He, G. De Geronimo, E. Vernon, J. Fried, Characterization of the H3D ASIC readout system and 6.0 cm3 3-D position sensitive CdZnTe Detectors, IEEE Trans. Nucl. Sci. 59 (2012) 236-242.   DOI