Browse > Article
http://dx.doi.org/10.1016/j.net.2019.02.001

Minimum detectable activity of plastic scintillator for in-situ beta measurement system in ground water  

Choi, Woo Nyun (Nuclear Engineering, Ulsan National Institute of Science and Technology)
Lee, UkJae (Nuclear Engineering, Ulsan National Institute of Science and Technology)
Bae, Jun Woo (Nuclear Engineering, Ulsan National Institute of Science and Technology)
Kim, Hee Reyoung (Nuclear Engineering, Ulsan National Institute of Science and Technology)
Publication Information
Nuclear Engineering and Technology / v.51, no.4, 2019 , pp. 1169-1175 More about this Journal
Abstract
The minimum detectable activity (MDA) value was derived according to the flow rate of the sample and degree of amplification of the device by sending the sample directly from the collection site to the detection part through a pump. This method can lead to reduction in time and cost compared to the existing measurement method that uses a pre-treatment process. In this study, experiments were conducted on $^3H$ and $^{90}Sr$, which are the major pure beta-emitting radionuclides, by setting the sample flow rate and the amplification gain as factors. The MDA values were derived according to the flow rates, considering that the flow rate can affect the MDA values. There were no change in the MDA under different flow rates of 0, 600, 800, and 1000 mL/min. Therefore, it was confirmed that the flow rate may not be considered when collecting samples for monitoring in actual field. As the degree of amplification of the amplifier increased, the time required to reach the target MDA decreased. When the amplification was quadrupled, the detection efficiency increased by approximately 23.4 times, and the time to reach the MDA decreased to approximately 1/550 times. This method offers the advantage of real-time on-site monitoring.
Keywords
Groundwater monitoring; Gross beta measurements; Tritium; On-site measurement; Minimum detectable activity;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 A. Soreefan, Development of an Original Laboratory Prototype for a Field Tritium Detector Containing a PEM Electrolyzer Mounted in Series with a Gas Proportional Counter, 2009.
2 P. Alaei, Introduction to health physics, Med. Phys. 35 (2008), 5959-5959.   DOI
3 http://www.nssc.go.kr/nssc/information/law4_4.jsp, The code of nuclear safety and security commission JAE2012-59Ho: Regulation for the clearance of radioactive waste (2015).
4 S.W. Duce, H. Amir, M.L. Mohagheighi, L.M. Mark, R.R. Robert, R.M. David, In situ radiation detection demonstration, in: WM'00 Conference, Tucson, AZ, February 27 -, March 2, 2000.
5 S.J. Ko, et al., Radiation Detection & Measurement, 2001 (in Korean language).
6 S. Croft, et al., The estimation of the minimum detectable activity from measured passive neutron coincidence counter data, in: Proc. 46th Int. Symp. on Nuclear Material Management (INMM), 2005.
7 T.T. Alton, S.D. Monk, D. Cheneler, Beta particle energy spectra shift due to self-attenuation effects in environmental sources, Nucl. Eng. Technol. 49 (2017) 1483-1488.   DOI
8 Ortec Inc. AN34 Experiment 9 Time Coincidence Techniques and Absolute Activity Measurements. https://www3.nd.edu/-wzech/Application-Note-AN34-Experiments-Nuclear-Science-Experiment-9.pdf.
9 Ortec Inc. Model 556H High Voltage Power Supply Operating and Service Manual. https://www.ortec-online.com/-/media/ametekortec/manuals/556hmnl.pdf.
10 HAMAMATSU, Photomultiplier tube R878. www.hamamatsu.com.
11 Ortec Inc. Model 855 Dual Spectroscopy Amplifier Operating and Service Manual. https://www.ortec-online.com/-/media/ametekortec/manuals/855-mnl.pdf.
12 PASTERNACK, RF Adapters Technical Data Sheet (PE9363).
13 Ortec Inc. Model 551 Timing Single-Channel Analyzer Operating and Service Manual. https://www.ortec-online.com/-/media/ametekortec/manuals/551-mnl.pdf.
14 Ortec Inc, EASY-MCA-8KTM EASY-MCA-2KTM digital gamma-ray spectrometer user's manual. https://www.ortec-online.com/-/media/ametekortec/manuals/easy-mca-mnl.pdf.
15 PASTERNACK, RF Adapters Technical Data Sheet (PE9336).
16 http://rarecomponents.com/store/1017,Search:19-March-2018.
17 Saint-Gobain Crystals, Organic Scintillation Materials. https://www.crystals.saint-gobain.com/products/organic-scintillation-materials, 2014.