Ceramic-Ceramic Wear of Zirconia/Alumina Composites for the Application of Total Hip Replacement  

Lee Kwon-Yong (Bioengineering Research Center, Sejong University)
Kim Hwan (Dept. of Mechanical Engineering, Sejong University)
Kim Dae-Jun (Bioengineering Research Center, Sejong University)
Lee Myong-Hyon (Advanced Material Analysis and Evaluation Team, KICET)
Seo Won-Seon (Advanced Material Analysis and Evaluation Team, KICET)
Publication Information
Tribology and Lubricants / v.21, no.5, 2005 , pp. 216-220 More about this Journal
Abstract
The sliding wear behaviors of three different compositions of novel low temperature degradation-free zirconia/alumina (LTD-free Z/A) composites were examined in a ceramic-ceramic contact pair. The wear tests were performed by using a pin-on-disk type wear tester in a linear reciprocal sliding motion with a line contact in both dry and bovine serum lubricated conditions at room temperature. From the results of dry sliding wear tests, Z/A#1((5.3Y, 4.6Nb)-TZP/80 $vol\%Al_2O_3$) showed the best wear resistance among three kinds of LTD-free Z/A composites. For the bovine serum lubricated sliding wear tests, wear was too little to be measured for all kinds of Z/A composites. These novel LTD-free Z/A composites having excellent wear resistance demonstrated a potential as the alternative materials for the ceramic-ceramic contact pairs of femoral head and acetabular liner in total hip replacement.
Keywords
zirconia/alumina composite; wear; low temperature degradation free; ceramic-ceramic contact pair.;
Citations & Related Records
연도 인용수 순위
  • Reference
1 김대준, '생체불활성 세라믹스,' 세라미스트 3(3), pp. 32-38, 2000
2 J. Briggs., 'Engineering Ceramics in Europe and The USA,' Enceram, UK. 1999
3 I.C. Clarke, B. Ben-Nissan, P.A. Williams and A. Gustafson., 'Abstract Book of 104th Ann. Meeting of the Am. Ceram. Soc,' The American Ceramic Society, USA, pp. 73, 2002
4 김대준, '생체세라믹스,' 한국세라믹스연감 2000, 월간세라믹스, pp. 169-175, 2000
5 D.-J. Kim, H.-J. Jung and D.-H. Cho, 'Phase transformations of Y,O, and Nb,O, doped tetragonal zirconia during low temperature aging in air,' Solid State Ionics, Vol. 80, pp. 67, 1995   DOI   ScienceOn
6 FDA: 'Recall of zirconia ceramic femoral heads for hip implants,' Sep. 13, 2001
7 FDA: 'Steam Re-sterilization causes deterioration of zirconia ceramic heads of total hip prosthesis,' Jun. 5, 1997
8 S. Lawson, 'Environmental Degradation of zirconia Ceramics,' Journal of European Ceramic Society, Vol. 15, pp. 485. 1995   DOI   ScienceOn