Browse > Article
http://dx.doi.org/10.11629/jpaar.2014.10.1.001

Development of a Real-time Monitoring Device for Measuring Particulate Matter  

Kim, Dae Seong (Global Frontier Center for Multiscale Energy Systems, Seoul National University)
Cho, Young Kuk (Global Frontier Center for Multiscale Energy Systems, Seoul National University)
Yoon, Young Hun (Department of Integrated Environmental systems, Pyeongtaek University)
Publication Information
Particle and aerosol research / v.10, no.1, 2014 , pp. 1-8 More about this Journal
Abstract
In this study, we have developed a real-time monitoring device for measuring $PM_{10/2.5/1}$ of ambient aerosol particles. The real-time PM (Particulate Matter) monitor was based on the light scattering method and had 16 channels in particle size. The laboratory and field tests were carried out to evaluate the performance of the PM monitor developed. Arizona Road Dust particles ranging from diameter of 0.1 to $20{\mu}m$ were generated as test particles in the laboratory test. The field test was carried out at the Seoul Meteorological Observatory. We can obtain the particle size and number concentration (particle size distribution) only from the real-time PM monitor developed. Therefore, the average density of aerosol particles was used to obtain the PM data from the particle size distribution. The $PM_{10/2.5/1}$ results of the PM monitor were compared with the data of the Grimm Dust Monitor (Model 1.108) and a beta ray gauge (Thermo Fisher Scientific). As a result, it was shown that the $PM_{10/2.5/1}$ results obtained by the real-time PM monitor agreed well with the data of the reference devices, and overall, the real-time PM monitor could be used as a PM monitoring device for real-time monitoring of the ambient particles.
Keywords
$PM_{10/2.5/1}$; real-time PM monitor; particle size distribution; aerosol measurement;
Citations & Related Records
연도 인용수 순위
  • Reference
1 U.S. EPA. (1987a). Reference Method for the Determination of Particulate Matter as PM10 in the Atmosphere, 40 CFR Part 50, U.S. Government Printing Office, Washington, DC.
2 Hinds, W. C. (1999). Aerosol technology, Second Edition, John Wiley & Sons, New York.
3 Dempsey, J. C., and Polinshuk, P. (1966). Radioisotopes for Aerospace, Part 2, Plenum Press, New York.
4 Wedding, J. B., and Weigand, M. A., (1993). An Automatic Particle Sampler with Beta Gauging, Journal of Air & Waste Management Association 43, 475- 479.
5 Park, S. S., Kim, Y. J., Lee, K. W., Chun, K. J., Lee, J. Y., Lim, Y. S., and Han, J. S. (2001). Development of an Automatic Beta Gauge Particulate Sampler with Filter Cassette Mechanism, Aerosol Science and Technology 35, 844-851.   DOI
6 Jaklevic, J. M., Gatti, R. C., Goulding, F. S., and Loo, B .W. (1981). A beta-Gauge Method Applied to Aerosol Samples, Environmental Science and Technology 15, 680-684.   DOI