Browse > Article
http://dx.doi.org/10.5757/JKVS.2008.17.3.234

Development and its Characteristics of the 40kV x-ray transmission anode target tube  

Kim, Sung-Soo (Dept. of Techno-marketing, Mokwon Univ.)
Kim, Do-Yun (VSI CO. LTD.)
Publication Information
Journal of the Korean Vacuum Society / v.17, no.3, 2008 , pp. 234-239 More about this Journal
Abstract
Tungsten and rhodium target tube for a 40 kV x-ray transmission anode was developed to apply to the hand-held XRF(X-Ray Fluorescence) apparatus and its characteristics were evaluated. From the measurement of the energy distribution and dose of x-ray, it was confirmed that our results were good agreements with the known ones. The optimum thickness of metal film deposited on Be window to extract the maximum dose were $2.6{\mu}m$ and $2.7{\mu}m$ in case of W-target tube and Rh-target tube, respectively. When it was continuously worked during 30 min. at 40 kV in tube voltage and at $60{\mu}A$ in tube current, the temperature at target did not exceed $50^{\circ}C$. Our results reveals that the 40 kV x-ray transmission anode tube can be applied to the hand-held XRF apparatus.
Keywords
x-ray tube; transmission anode; x-ray fluorescence; Tungsten; Rhodium; XRF;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Analytical chemistry H. A. Liebhafsky 1972 WILEY INTERSCIENCE p16
2 X-rays in Theory and Experiment. Arthur H. Compton 1967 D.VAN NOSTRAND COMPANY, INC. p89
3 S. S. Kim, D. Y. Kim, J. Kor. Vac. Soc. 16 (4), 305 (2007)   과학기술학회마을   DOI   ScienceOn
4 United States Patent 4,034,251 July 5, 1977
5 Comprehensive Analytical Chemistry Volume XXIV G.Svehla 1989 ELSEVIER p70