Browse > Article
http://dx.doi.org/10.13104/imri.2018.22.4.218

T1-Based MR Temperature Monitoring with RF Field Change Correction at 7.0T  

Kim, Jong-Min (Department of Electronics and Information Engineering and ICT Convergence Technology for Health & Safety, Korea University)
Lee, Chulhyun (Bioimaging Research Team, Korea Basic Science Institute)
Hong, Seong-Dae (Department of Electronics and Information Engineering and ICT Convergence Technology for Health & Safety, Korea University)
Kim, Jeong-Hee (Korea Artificial Organ Center)
Sun, Kyung (Korea Artificial Organ Center)
Oh, Chang-Hyun (Department of Electronics and Information Engineering and ICT Convergence Technology for Health & Safety, Korea University)
Publication Information
Investigative Magnetic Resonance Imaging / v.22, no.4, 2018 , pp. 218-228 More about this Journal
Abstract
Purpose: The objective of this study is to determine the effect of physical changes on MR temperature imaging at 7.0T and to examine proton-resonance-frequency related changes of MR phase images and T1 related changes of MR magnitude images, which are obtained for MR thermometry at various magnetic field strengths. Materials and Methods: An MR-compatible capacitive-coupled radio-frequency hyperthermia system was implemented for heating a phantom and swine muscle tissue, which can be used for both 7.0T and 3.0T MRI. To determine the effect of flip angle correction on T1-based MR thermometry, proton resonance frequency, apparent T1, actual flip angle, and T1 images were obtained. For this purpose, three types of imaging sequences are used, namely, T1-weighted fast field echo with variable flip angle method, dual repetition time method, and variable flip angle method with radio-frequency field nonuniformity correction. Results: Signal-to-noise ratio of the proton resonance frequency shift-based temperature images obtained at 7.0T was five-fold higher than that at 3.0T. The T1 value increases with increasing temperature at both 3.0T and 7.0T. However, temperature measurement using apparent T1-based MR thermometry results in bias and error because B1 varies with temperature. After correcting for the effect of B1 changes, our experimental results confirmed that the calculated T1 increases with increasing temperature both at 3.0T and 7.0T. Conclusion: This study suggests that the temperature-induced flip angle variations need to be considered for accurate temperature measurements in T1-based MR thermometry.
Keywords
MR thermometry; 7T; RF hyperthermia; Proton resonance frequency shift; T1; RF field;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Lewa CJ, Majewska Z. Temperature relationships of proton spin-lattice relaxation time T1 in biological tissues. Bull Cancer 1980;67:525-530
2 Graham SJ, Bronskill MJ, Henkelman RM. Time and temperature dependence of MR parameters during thermal coagulation of ex vivo rabbit muscle. Magn Reson Med 1998;39:198-203   DOI
3 Ishihara Y, Calderon A, Watanabe H, et al. A precise and fast temperature mapping using water proton chemical shift. Magn Reson Med 1995;34:814-823   DOI
4 Kim JM, Lee C, Jo YS, et al. Monitoring and guidance on high-intensity focused ultrasound treatment by multiple fast field echo at 3.0 T MRI: ex-vivo studies with multiparametric mapping. In Proceedings of the 26th Annual Meeting ISMRM, Paris, France, 2018:1485
5 Moseley ME, Cohen Y, Mintorovitch J, et al. Early detection of regional cerebral ischemia in cats: comparison of diffusion- and T2-weighted MRI and spectroscopy. Magn Reson Med 1990;14:330-346   DOI
6 Young IR, Hand JW, Oatridge A, Prior MV. Modeling and observation of temperature changes in vivo using MRI. Magn Reson Med 1994;32:358-369   DOI
7 Langereis S, Keupp J, van Velthoven JL, et al. A temperature-sensitive liposomal 1H CEST and 19F contrast agent for MR image-guided drug delivery. J Am Chem Soc 2009;131:1380-1381   DOI
8 Schwan HP, Kay CF. The conductivity of living tissues. Ann N Y Acad Sci 1957;65:1007-1013   DOI
9 Yeo SY, Katscher U, Kim YS, Gruell H. Conductivity imaging for assessing the treatment outcome of MR-HIFU ablation of uterine fibroids. In Proceedings of the 25th Annual Meeting ISMRM, Honolulu, Hawaii, USA, 2017:2102
10 Kwon DH, Kim JM, Oh SH, et al. Seven-Tesla magnetic resonance images of the substantia nigra in Parkinson disease. Ann Neurol 2012;71:267-277   DOI
11 Goelman G, Liu S, Fleysher R, Fleysher L, Grossman RI, Gonen O. Chemical-shift artifact reduction in Hadamardencoded MR spectroscopic imaging at high (3T and 7T) magnetic fields. Magn Reson Med 2007;58:167-173   DOI
12 Rieke V, Butts Pauly K. MR thermometry. J Magn Reson Imaging 2008;27:376-390   DOI
13 Wang P. Evaluation of MR thermometry with proton resonance frequency method at 7T. Quant Imaging Med Surg 2017;7:259-266   DOI
14 Lee Y, Callaghan MF, Nagy Z. Analysis of the precision of variable flip angle T1 mapping with emphasis on the noise propagated from RF transmit field maps. Front Neurosci 2017;11:106   DOI
15 Fite BZ, Liu Y, Kruse DE, et al. Magnetic resonance thermometry at 7T for real-time monitoring and correction of ultrasound induced mild hyperthermia. PLoS One 2012;7:e35509   DOI
16 Henderson E, McKinnon G, Lee TY, Rutt BK. A fast 3D looklocker method for volumetric T1 mapping. Magn Reson Imaging 1999;17:1163-1171   DOI
17 Zhu DC, Penn RD. Full-brain T1 mapping through inversion recovery fast spin echo imaging with time-efficient slice ordering. Magn Reson Med 2005;54:725-731   DOI
18 Cheng HL, Wright GA. Rapid high-resolution T(1) mapping by variable flip angles: accurate and precise measurements in the presence of radiofrequency field inhomogeneity. Magn Reson Med 2006;55:566-574   DOI
19 Liberman G, Louzoun Y, Ben Bashat D. T(1) mapping using variable flip angle SPGR data with flip angle correction. J Magn Reson Imaging 2014;40:171-180   DOI
20 Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med 2007;57:192-200   DOI
21 Helms G, Dathe H, Dechent P. Quantitative FLASH MRI at 3T using a rational approximation of the Ernst equation. Magn Reson Med 2008;59:667-672   DOI
22 Kim HJ, Kim JM, Jo YS, et al. Experiment study of MR compatible RF hyperthermia system. In Proceedings of the 24th Annual Meeting ISMRM, Singapore, 2016:1644
23 Kim JM, Kumar S, Jo YS, et al. Temperature-rangedependent optimization of noninvasive MR thermometry methods. J Bio Eng Res 2015;36:241-250   DOI
24 Kim JM, Lee C, Hong SD, et al. Comparison of temperature mapping methods using proton resonance frequency shift and T1 in 3-T and 7-T MRI. In Proceedings of the 25th Annual Meeting ISMRM, Honolulu, Hawaii, USA, 2017:2588