Accurate Localization of Metal Electrodes Using Magnetic Resonance Imaging  

Joe, Eun-Hae (School of Electrical and Electronic Engineering, Yonsei University)
Ghim, Min-Oh (School of Electrical and Electronic Engineering, Yonsei University)
Ha, Yoon (Department of Neurosurgery, College of Medicine, Yonsei University)
Kim, Dong-Hyun (School of Electrical and Electronic Engineering, Yonsei University)
Publication Information
Investigative Magnetic Resonance Imaging / v.15, no.1, 2011 , pp. 11-21 More about this Journal
Abstract
Purpose : Localization using MRI is difficult due to susceptibility induced artifacts caused by metal electrodes. Here we took an advantage of the B0 pattern induced by the metal electrodes by using an oblique-view imaging method. Materials and Methods : Metal electrode models with various diameters and susceptibilities were simulated to understand the aspect of field distortion. We set localization criteria for a turbo spin-echo (TSE) sequence usingconventional ($90^{\circ}$ view) and $45^{\circ}$ oblique-view imaging method through simulation of images with various resolutions and validated the criteria usingphantom images acquired by a 3.0T clinical MRI system. For a gradient-refocused echo (GRE) sequence, which is relatively more sensitive to field inhomogeneity, we used phase images to find the center of electrode. Results : There was least field inhomogeneity along the $45^{\circ}$ line that penetrated the center of the electrode. Therefore, our criteria for the TSE sequence with $45^{\circ}$ oblique-view was coincided regardless of susceptibility. And with $45^{\circ}$ oblique-view angle images, pixel shifts were bidirectional so we can detect the location of electrodes even in low resolution. For the GRE sequence, the $45^{\circ}$ oblique-view anglemethod madethe lines where field polarity changes become coincident to the Cartesian grid so the localization of the center coordinates was more facilitated. Conclusion : We suggested the method for accurate localization of electrode using $45^{\circ}$ oblique-view angle imaging. It is expected to be a novelmethodto monitoring an electrophysiological brain study and brain neurosurgery.
Keywords
Microelectrode localization; Field inhomogeneity; Susceptibility artifact; Metallic artifact;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Truwit CL, Liu H. Prospective stereotaxy: a novel method of trajectory alignment using real-time image guidance. J Magn Reson Imaging 2001;13:452-457   DOI   ScienceOn
2 Martin AJ, Larson PS, Ostrem JL. Placement of deep brain stimulator electrodes using real-time high-field interventional magnetic resonance imaging. Magn Reson Med 2005;54:1107-1114   DOI   ScienceOn
3 Zrinzo L, Zrinzo LV, Tisch S, et al. Stereotactic localization of the human pedunculopontine nucleus: atlas-based coordinates and validation of a magnetic resonance imaging protocol for direct localization. Brain 2008;131:1588-1598   DOI   ScienceOn
4 Oya H, Kawasaki H, Dahdaleh NS, Wemmie JH, Howard MA. Stereotactic atlas-based depth electrode localization in the human amygdala. Stereotact Funet Neurosurg 2009;87:219-228   DOI   ScienceOn
5 Salomir R, De Senneville BD, Moonen CTW. A fast calculation method for magnetic field inhomogeneity due to an arbitrary distribution of bulk susceptibility. Concept Magn Reson B 2003; 19B:26-34   DOI
6 Koch KM, Papademetris X, Rothman DL, de Graaf RA. Rapid calculations of susceptibility-induced magnetostatic field perturbations for in vivo magnetic resonance. Phys Med Biol 2006;51:6381-6402   DOI   ScienceOn
7 Hopper TAJ, Vasilic B, Pope JM, et al. Experimental and computational analyses of the effects of slice distortion from a metallic sphere in an MRI phantom. Magn Reson Imaging 2006;24:1077-1085   DOI   ScienceOn
8 Schrader B, Hamel W, Weinert D, Mehdorn HM. Documentation of electrode localization. Mov Disord 2002;17:S167-S174   DOI   ScienceOn
9 Ludeke KM, Roschmann P, Tischler R. Susceptibility artifacts in NMR imaging. Magn Reson Imaging 1985;3:329-343   DOI   ScienceOn
10 Ladd ME, Erhart P, Debatin JF, Romanowski BJ, Boesiger P, McKinnon GC. Biopsy needle susceptibility artifacts. Magn Reson Med 1996;36:646-651   DOI   ScienceOn
11 Muller-Bierl B, Graf H, Lauer V, Steidle G, Schick F. Numerical modeling of needle tip artifacts in MR gradient echo imaging. Med Phys 2004;31:579-587   DOI   ScienceOn
12 Martinez-Santiesteban FM, Swanson SD, Non DC, Anderson DJ. Magnetic field perturbation of neural recording and stimulating microelectrodes. Phys Med Biol 2007;52:2073-2088   DOI   ScienceOn
13 Liu H, Hall WA, Martin AJ, Truwit CL. Biopsy needle tip artifact in MR-Guided Neurosurgery. J Magn Reson Imaging 2001;13:16-22   DOI   ScienceOn
14 Matsui T, Koyano KW, Koyama M, et al. MRI-based localization of electrophysiological recording sites within the cerebral cortex at single-voxel accuracy. Nat Methods 2007;4:161-168   DOI   ScienceOn