Realization of the national standard of candela traceable to the absolute cryogenic radiometer at KRISS  

Park, Seung-Nam (Division of Optical Metrology, KRISS)
Kim, Yong-Wan (Division of Optical Metrology, KRISS)
Lee, Dong-Hoon (Division of Optical Metrology, KRISS)
Publication Information
Korean Journal of Optics and Photonics / v.15, no.5, 2004 , pp. 443-448 More about this Journal
Abstract
We realized the national standard of the candela, one of the SI units, by using two photometers with the spectral responsivity measured in reference to the absolute cryogenic radiometer. The external apertures of the photometers were fabricated using a diamond turning machine, and measured in terms of area with uncertainty of 0.05 %(k = 1). The candela is realized using a 1 kW FEL lamp and the characterized photometers on an optical bench. The uncertainty is budgeted to be 0.25 %(k = 1) considering the uncertainty of the spectral responsivity and the response uniformity of the detectors, the area of the external apertures, the color temperature of the lamp, and the positioning reproducibility of the photometers and the lamp. We verified the realized scale by comparing with the scale of National Institute of Standards and Technology, USA. They coincided with each other within 0.1%.
Keywords
Radiometry; Photometry; Candela;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 E. Ikonen, P. Karha, ALassila, F. Manoonchehri, H. Fagerlund and L. Liedquist, 'Radiometric realization of the candela with a trap detector,' Metrologia, vol. 32, pp. 689-692, 1996   DOI   ScienceOn
2 Albert Parr, 'The candela and photometric and radiometric measurements,' Research Journal of the National Institute of Standards and Technology, vol. 106, pp. 151-186, 2000
3 Yoshi Ohno, Handbook of Applied Photometry, (edited by Casimer DeCusatis, AIP, New York, 1997) pp. 55-98
4 Principles of Light Measurement, Commission Internationale de I'Eclairage(CIE) Pub. no. 18(E-12), 1970
5 BIPM, 'Calibration and Measurement Capabilities of National Metrology Institute,' (http://kcdb.bipm.org/AppendixC)
6 J. E. Martin, N. P. Fox, and P. J. Key, 'A cryogenic radiometer for absolute radiometric measurements,' Metrologia, vol. 21, pp. 147-155, 1985   DOI   ScienceOn
7 A. Lassila, P. Toivanen, and E. Ikonen, 'An optical method for direct determination of the radiometric aperture area at high accuracy,' Measure. Sci. Technology, vol. 8, pp. 973-977, 1997   DOI   ScienceOn
8 강창호, 김석원, 박승남, '레이저 스폿의 칼날주사 방법에 의한 복사계 개구 면적의 정밀측정,' 한국광학회지, 제15권 제3호, pp. 258-262, 2004   과학기술학회마을   DOI