Browse > Article

A 16kw Rotating Anode type Monoblock X-ray Generator  

Oh, Jun-Yong ((주)에코트론)
Kim, Yuen-Chung ((주)에코트론)
Kim, Hack-Seong (동양공전 자동화시스템과)
Publication Information
The Transactions of the Korean Institute of Power Electronics / v.11, no.2, 2006 , pp. 97-103 More about this Journal
Abstract
This paper designs mono block type with rotating high power radiography x-ray generator and studies 16kW X-ray generator possible to adapt hospital mobile radiography and industrial X-ray equipment and design. This equipment uses rotating anode type x-ray tube at high voltage generator to generate x-ray and adds rotor operating circuit to operate rotor of x-ray tube. The size of high voltage transformer and high voltage generator is minimized by high voltage high frequency inverter has 100kHz switching frequency. Also this paper shows result of x-ray tube voltage and tube current correspond to variable load.
Keywords
Mono block; Radiography x-ray generator; High voltage generator; Rotating anode type x-ray tube; High voltage transformer;
Citations & Related Records
연도 인용수 순위
  • Reference
1 K.T. Compton: Magnetic Circuit and Transformers, John Wiley and sons, New-York, Chap. 18, 1943
2 Rs. A. McCurrie: Ferromagnetic Materials, Academic, London, Chap. 2, 1994
3 Hino, H., Hatakeyama, T. and NakaokaM.,'Resonant PWM Inverter Linked DC-DC Converter using parasitic Impedances of High Voltage Transformer and Its Applications to X-Ray Generator', PESC Record pp. 1212-1219, 1998
4 Izumi Ogura, and T. Aoyagi, 'Analysis of the Basic Motion of the Resonance Type Inverter X-ray Generator by the Model circuit'
5 김학성, 박영국 오준용, 성기봉, '진단 X-선용 40kW 고주파 고압 전원 시스템', 전력전자 학술대회 논문집, pp. 192-198, 2003. 4
6 Kelly J.P and Trout E.D, 'Physical Characteristics of the Radiation from 2-Pulse, 12-Pulse, and 1,000-Pulse X-Ray Equipment', Radiology100, pp. 653-661, 1971   DOI