Browse > Article

High Frequency High Voltage 40kW Power System for Diagnosis X-ray  

김학성 (동양공업전문대학 자동화시스템과)
박영국 (㈜포스콤 연구소)
오준용 (㈜포스콤 연구소)
성기봉 (㈜포스콤 연구소)
Publication Information
The Transactions of the Korean Institute of Power Electronics / v.8, no.2, 2003 , pp. 192-198 More about this Journal
Abstract
Recently, the Inverter type X-ray generator is rapidly replacing the conventional single-phase or three-phase X-ray generator, it has several merits of space-saving: high accuracy and reproducibility. This paper presents a 40kw(12kV, 80mA) high tension generator system for diagnosis X-ray. The control circuit and design consideration of the proposed high tension power supply are given. Issues in the design of high voltage isolating transformer are discussed. Experimental results are presented to verify the performance of the designed power supply for varying load conditions. 1'he proposed apparatus has several advantages, e. g., the fast rising time of tube voltage, accuracy and reduced component size etc.
Keywords
X-ray generator; high tension power supply; high voltage isolating transformer;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 /
[ Rs. A. McCurrie ] / Ferromagnetic Materials
2 이온주입형 정밀고압 발행장치 연구 /
[ 유동욱(외6인) ] / 전력전자 학술대회 논문집
3 Analysis of the Basic Motion of the Resonance Type Inverter X-ray Generator by the Model Circuit /
[ Izumi Ogura;T. Aoyagi ] / 東京部立醫療技術短期 大學紀要
4 Physical Characteristics of the Radiation from 2-Pulse, 12-Pulse, and 1,000-Pulse X-Ray Equipment /
[ Kelly J.P;Trout E.D ] / Radiology   DOI
5 고압 대전류 펄스 자계를 이용한 저온 살균장치 시스템 연구 /
[ 백주원(외5인) ] / 전력전자학회 전력전자 학술대회 논문집
6 Resonant PWM Inverter Linked DC-DC Converter using parasitic Impedances of High Voltage Transformer and Its Applications to X-Ray Generator /
[ Hino, H.;Hatakeyama, T.;Nakaoka, M. ] / PESC Record
7 /
[ K.T. Compton(et al.) ] / Magnetic Circuit and Transformers
8 インバ-タ式 X線裝置の特性(共振形, 最高出力:50kW) /
[ 安部;小倉;靑柳;地 ] / 日放技學