Abstract
This paper describes the design of an elbow rehabilitation robot based on force measurement that enables a severe stroke patient confined to their bed to receive elbow rehabilitation exercises. The developed elbow rehabilitation robot was providewitha two-axis force/torque sensor which can detect force Fz and torque Tz, thereby allowing it to measure therotational force (Tz) exerted on the elbow and the signal force Fz which can be used as a safety device. The robot was designed and manufactured for severe stroke patients confined to bed, and the robot program was manufactured to perform flexibility elbow rehabilitation exercises. Asa result of the characteristics test of the developed rehabilitation robot, the device was safely operated while the elbow rehabilitation exercises were performed. Therefore, it is thought that the developed rehabilitation robot can be used for severe stroke patients.
Keywords
elbow rehabilitation robot; two-axis force sensor; rated output; rehabilitation robot; interference error; force control;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 P. R. Culmer, A. E. Jackson, S. Makower, R. Richardson, J. A. Cozens, M. C. Levesley, and B. B. Bhakta, "A control strategy for upper limb robotic rehabilitation with a dual robot system," IEEE/ASME Transactions on Mechatronics, vol. 15, no. 4, pp. 575-585, 2010.   DOI
2 L. Pan, A. Song, G. Xu, H. Li, H. Zeng, and B. Xu, "Safety supervisory strategy for an upper-limb rehabilitation robot based on impedance control," International Journal of Advanced Robotic Systems, vol. 10, pp. 1-12, 2013.
3 H. C. Kim, L. M. Miller, I. Fedulow, M. Simkins, G. M. Abrams, N. Byl, and J. Rosen, "Kinematic data analysis for post-stroke patients following bilateral versus unilateral rehabilitation with an upper limb wearable robotic system," IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol. 21, no. 2, 2013.
4 C. Li, Y. Inoue, T. Liu, K. Shibata, and K. Oka, "A self-controlled robot for upper limb rehabilitation," Proc. of the 2010 IEEE International Conference on Information and Automation, pp. 115-118, 2010.
5 M. H. Rahman, M. Saad, J. P. Kenne, and P. S. Archambault, "Modeling and development of an exoskeleton robot for rehabilitation of wrist movements," Advanced Intelligent Mechatronics (AIM), IEEE/ASME International Conference on, pp. 25-30, 2010.
6 B. C. Tsai, W. W. Wang, L. C. Hsu, L. C. Fu, and J. S. Lai, "An articulated rehabilitation robot for upper limb physiotherapy and training," Intelligent Robots and Systems (IROS), IEEE/RSJ International Conference on, pp. 1470-1475, 2010.
7 S. Liu, Y. Xie, S. Jiang, and J. Meng, "Interactive control for the arm rehabilitation robot," Intelligent Systems and Applications, International Workshop on, pp. 1-4, 2009.
8 W. Huo, J. Huang, Y. W. Jun, and L. C. Wu, "Control of upper-limb power-assist exoskeleton based on motion intention recognition," IEEE International Conference on Robotics and Automation, pp. 2243-2248, 2011.
9 A. Umemura, Y. Saito, and K. Fujisaki, "A study on powerassisted rehabilitation robot arms operated by patient with upper limb disabilities," Rehabilitation Robotics, ICORR, IEEE International Conference on, pp. 451-456, 2009.
10 Y. Ren, H. S. Park, and L. Q. Zhang, "Developing a whole-arm exoskeleton robot with hand opening and closing mechanism for upper limb stroke rehabilitation," Rehabilitation Robotics, ICORR, IEEE International Conference on, pp. 761-765, 2009.
11 B. Wu and P. Cai, "Decoupling analysis of a sliding structure six-axis force/torque sensor," Measurement Science Review, vol. 13, no. 4, 2013.
12 G. S. Kim, "Development of a hook-type finger force measuring system with force sensors," Journal of Institute of Control, Robotics and Systems (in Korean), vol. 20, no. 6, pp. 663-668, 2014.   DOI
13 K. Nagai, Y. Ito, M. Yazaki, K. Higuchi, and S. Abe, "Development of a small Six-component force/torque sensor based on the double-cross structure," Journal of the Robotics Society of Japan, vol. 22, no. 3, pp. 361-369, 2004.   DOI
14 K. J. Lee, and G. S. Kim, "Design of structure of four-axis force/torque sensor with parallel step plate beams," Journal of Institute of Control, Robotics and Systems (in Korean), vol. 20, no. 11, pp. 1147-1152, 2014.   DOI
15 G. S. Kim, "Development of a six-axis force/moment sensor with rectangular taper beams for an intelligent robot," International Journal of Control, Automation, and Systems, vol. 5, no. 4, pp. 419-428, 2007.
16 Y. K. Kim and G. S. Kim, "Design of six-axis force/moment sensor for ankle-rehabilitation robot," Journal of Institute of Control, Robotics and Systems (in Korean), vol. 19, no. 4, pp. 357-363, 2013.   DOI   ScienceOn
17 ATI Industrial Automation, "Multi-axis forcre/torque sensor," ATI Industrial Automation, pp. 4-45, 2005.
18 G. S. Kim and J. W. Yoon, "Development of calibration system for multi-axis force/moment sensor and its uncertainty evaluation," Korean Society Precision Engineering, vol. 24, no. 10, pp. 91-98, 2007.