Because the rain gauges of tipping bucket type can easily use the digital signal, the rain gauges are widely used for the meteorological observation. In general, the resolution of rain gauges of tipping bucket type can be categorized by the 0.1mm, 0.5mm, and 1.0mm classes. But, the error of the tipping bucket rain gauges is made by the intensity of rainfalls and is expected to make the standard calibration method for error measurement. Thus, we developed the hardware of standard calibration facility for rain gauges by weighting measurement method and proposed the standard procedure by rainfall intensity in this study Also, we calculated the uncertainty for the rainfall intensity and obtained useful result through the proposed calibration method.
Keywords
rain gauges; standard calibration; rainfall; uncertainty;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 WMO, 'Guide to meteorological instruments and methods of observation,' WMO-no. 8. 1996
2 우덕모, '전도형 우량계의 강우강도별 특성,' 한국기상학회지, vol. 38, no. 5, pp. 479-491, 2002
3 김대원, 이부용, '우량계 개발과 측정 오차,' 한국환경과학회지, vol. 11, no. 10, pp. 1055-1060, 2002   과학기술학회마을   DOI
4 건설교통부, '수분관측매뉴얼,' 2004
5 한국교정시험기관인정기구, '측정불확도 추정사례집,' 2004
6 한국표준과학연구원, '불확도 평가 및 표현 사례집,' 2005