Testing Evaluation for Stand-Alone Cervical Cage  

Baek, Myong-Hyun (Medical Device Development Center, Osong Medical Innovation Foundation)
Lee, Mun-Hwan (Medical Device Development Center, Osong Medical Innovation Foundation)
Publication Information
Abstract
Stand-alone cervical cage consists of a PEEK body, Ti plate, and screw, which are configured as a single-piece. Through a single operation, this implantable medical device is capable of completely fixing the cervical vertebral body. For example, instead of a plate, which is normally used, the intervertebral disc is removed and replaced with a cervical cage. It should be noted that in Korea, KFDA guidelines for a stand-alone cervical cage have not yet been suggested. Therefore, the aim of this study is to present the systematic study of the static compression test, static torsion test, dynamic compression test, and expulsion test. Further, the test method is designed to refer to the ASTM standard and relative literature.
Keywords
Stand-Alone cervical cage; Medical device; Vertebral body; Intervertebral disc;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Health Insurance Review & Assessment Service. "In 2014, One Quarter of the Population of South Korea got Treatment for Spinal Disease," 2014.
2 Jeon, T.-S., Chang, H., and Choi, B.-W., "Current Concept on the Operative Treatment for Degenerative Cervical Disc Disease," Journal of the Korean Medical Association, Vol. 54, No. 9, 2011.
3 ASTM F2077-11, "Test Methods for Intervertebral Body Fusion Devices," 2011.
4 Goel, V. K., Dick, D., Jackson, M., Kuroki, H., and Ebraheim, N., "Expulsion Testing of an Intervertebral Fusion Cage: Effect of Tooth Design on Pull out Resistance," Proc. of the Summer Bioengineering Conference, pp. 111-112, 2003.