Browse > Article

Optimal design of a flexure hinge-based XY AFM scanner for minimizing Abbe errors and the evaluation of pitch measuring uncertainty of a nano-accuracy AFM system  

Kim Dong-Min (한국과학기술원 기계공학과)
Lee Dong-Yeon (한국과학기술원 기계공학과 NOM Lab.)
Gweon Dae-Gab (한국과학기술원 기계공학과 NOM Lab.)
Publication Information
Abstract
To establish of standard technique of nano-length measurement in 2D plane, new AFM system has been designed. In the long range (about several tens of ${\mu}m$), measurement uncertainty is dominantly affected by the Abbe error of XY scanning stage. No linear stage is perfectly straight; in other words, every scanning stage is subject to tilting, pitch and yaw motion. In this paper, an AFM system with minimum offset of XY sensing is designed. And XY scanning stage is designed to minimize rotation angle because Abbe errors occur through the multiply of offset and rotation angle. To minimize the rotation angle optimal design has performed by maximizing the stiffness ratio of motion direction to the parasitic motion direction of each stage. This paper describes the design scheme of full AFM system, especially about XY stage. Full range of fabricated XY scanner is $100{\mu}m\times100{\mu}m$. And tilting, pitch and yaw motion are measured by autocollimator to evaluate the performance of XY stage. As a result, XY scanner can have good performance. Using this AFM system, 3um pitch specimen was measured. The uncertainty of total system has been evaluated. X and Y direction performance is different. X-direction measuring performance is better. So to evaluate only ID pitch length, X-direction scanning is preferable. Its expanded uncertainty(k=2) is $\sqrt{(3.96)^2+(4.10\times10^{-5}{\times}p)^2}$ measured length in nm.
Keywords
AFM; nano-metrology; Flexure hinge stage; Abbe error minimization; Uncertainty evaluation; pitch length;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Zygo, 'ZMI 2000 System manual,' http://www.zygo.com
2 Gamaes, J., Kofod, N., Kuhle, A., Nielsen, C., Dirscherl, K. and Blunt, L., 'Calibration of step heights and roughness measurements with atomic force microscopes,' Precision Engineering, Vol. 27, pp. 91-98, 2003   DOI   ScienceOn
3 Dixon, R. G, Koning, R. and Tsai, V. W., 'Metrology, Inspection and Process Control for Microlithography ?XIII,' Proc. SPIE, Bhanwar Singh, Ed., Vol. 3677, pp. 20-34, 1999   DOI
4 Meli, F. and Thalmann, R., 'Long-range AFM profiler used for accurate pitch measurements,' Meas. Sci. Technol., Vol. 9, pp. 1087-1092, 1998   DOI   ScienceOn
5 Gonda, S., Doi, T., Kurosawa, T. and Tanimura, Y., 'Real-time interferometrically measuring atomic force microscope for direct calibration of standards,' Review of Sci. Instr., Vol. 70, pp. 3362-3368, 1999   DOI
6 Hasche, K., Herrmann, K., Mirande, W., Seemann, R., Vitushkin, L., Xu, M. and Yu, G, 'Calibrated Scanning Force Microscope with Capabilities in the Subnanometre Range,' Surf. Interface Analysis, Vol. 33, pp. 71-74,2002   DOI   ScienceOn
7 www.npl.co.uk/euromet/length/annuaI_reports/an_rep_2004_2005.pdf
8 Meli, F., Thalman, R. and Blattner, P., 'High precision pitch calibration of grating using laser diffractometry,' 1st international conference and general meeting of the euspen society for precision engineering and nanometrology, pp. 252-255, 1999
9 Bienias, M., Gao, S., Rasche, K., Seeman, R. and Thiele, K., '3D Calibration of a Scanning Force Microscope with Internal Laser Interferometers,' Surf. Interface Anal., Vol. 25, pp. 606-610,1997   DOI   ScienceOn