Browse > Article

Biomechanical Analysis of the Implanted Constrained and Unconstrained ICR Types of Artificial Disc using FE Model  

Yun Sang-Seok (단국대학교 대학원)
Jung Sang-Ki (서울대학교 신경외과)
Kim Young-Eun (단국대학교 기계공학과)
Publication Information
Abstract
Although several artificial disc designs have been developed for the treatment of discogenic low back pain, biomechanical changes with its implantation were rarely studied. To evaluate the effect of artificial disc implantation on the biomechanics of functional spinal unit, a nonlinear three-dimensional finite element model of L4-L5 was developed with 1-mm CT scan data. Biomechanical analysis was performed for two different types of artificial disc having constrained and unconstrained instant center of rotation(ICR), ProDisc and SB Charite III model. The implanted model predictions were compared with that of intact model. Angular motion of vertebral body, forces on the spinal ligaments and facet joint, and stress distribution of vertebral endplate for flexion-extension, lateral bending, and axial rotation with a compressive preload of 400N were compared. The implanted model showed increased flexion-extension range of motion compared to that of intact model. Under 6Nm moment, the range of motion were 140%, 170% and 200% of intact in SB Charite III model and 133%, 137%, and 138% in ProDisc model. The increased stress distribution on vertebral endplate for implanted cases could be able to explain the heterotopic ossification around vertebral body in clinical observation. As a result of this study, it is obvious that implanted segment with artificial disc suffers from increased motion and stress that can result in accelerated degenerated change of surrounding structure. Unconstrained ICR model showed increased in motion but less stress in the implanted segment than constrained model.
Keywords
Artificial Disc; FE Model; Motion Segment; ICR; Stress;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Dooris, A.P., Goel, V.K., Grosland, N.M., Gilbertson, L.G. and Wilder, D.G., 'Load-sharing between Anterior and Posterior Elements in Lumbar Motion Segment Implanted with an Artificial Disc,' Spine, Vol. 26, pp. E122-E129, 2001   DOI   ScienceOn
2 Rohlmann, A., Zander T., and Bergmann G., 'Effect of Total Disc Replacement with ProDisc on Intersegmental Rotation of the Lumbar Spine,' Spine, Vol. 30, pp.738-743, 2005   DOI   ScienceOn
3 Kim, Y.E., Cho, S.Y. and Choi, H.Y., 'Analysis of Dural-sac OCclusion in a Lumbar Spinal Motion Segment FE Model,' J. of Musculoskeletal Research, Vol. 5, pp. 243-252, 2001   DOI   ScienceOn
4 McAfee, P.C., Cunningham, B.W., Devine, J., Williams, E. and Yu-Yahiro, J., 'Classification of Heterotopic Ossification in Artificial Disk Replacement,' J. of Spinal Disorder & Techniques, Vol.16, pp.384-389, 2003   DOI   ScienceOn
5 De Kleuver, M., Oner, F.C. and Jacobs, W.C.H., 'Total Disc Replacement for Chronic Low Back pain: background and a Systematic Review of the Literature,' Eur. Spine J., Vol. 12, pp.108-116, 2003   DOI   ScienceOn
6 Butter-Janz, K., Schellnack, K. and Zippel, H., 'Biomechanics of the SB Charite Lumbar Intervertebral Disc Endoprosthesis,' Int. Orthop., Vol. 13, pp.173-176, 1989   DOI
7 Lemaire, J.P., Skalli, W., Lavesate, F., Templier, A, Mendes, F., Diop, A., Sauty, V. and Laloux, E., 'Intervetevral Disc Prosthesis. Results and Prospects for the Year 2000,' Clin. Orthop., Vol. 337, pp. 64-76, 1997   DOI
8 Cunningham, B.W., Gordon, J.D., Dmitriev, AE., Hu, N. and McAfee, C., 'Biomechanical Evaluation of Total Disc Replacement Arthroplasty: An in Vitro human Cadaveric Model,' Spine, Vol. 28, pp. S110-S117, 2004   DOI   ScienceOn