Browse > Article

Effect of Stem Design on Contact Pressure and Stress Distribution of End-of-stem in Revision TKR  

Kim, Yoon-Hyuk (경희대학교 테크노공학대학, 산학협력기술원)
Kwon, Oh-Soo (카톨릭의대 의정부 병원)
Park, Si-Mon (경희대학교 대학원 기계공학과)
Publication Information
Abstract
In this study, the effect of stem-end design on contact pressure and stress distribution in revision TKR was investigated using finite element method. The finite element model of tibia, including the cortical bone, the cancellous bone and canal, was developed based on CT images. The implant models with various stem lengths, diameters, friction coefficients, and press-fit effects were considered. The results showed that the longer stem length, the stronger press-fit, the bigger stem diameter, and the higher friction coefficient increased both peak contact pressure and Von-Mises stress distributions. The results supported the clinical hypothesis that peak contact pressure and stress are related to the stem end pain. The results of this study will be useful to design the stem and reduce the end-of-stem pain in revision TKR.
Keywords
Revision TKR; Finite element analysis; Stem pain; Stem design;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Brooks, P. J., Walker, P. S. and Scott, R. D., 'Tibial com- ponent fixation in deficient tibial bone stock,' Clinical orthopaedics and related research, Vol. 184, pp. 302-308, 1984
2 Jason, P. H., Anthony, J. P. and Paul, J. R., 'Explicit finite element modeling of total Knee replacement mechanics,' Journal of Biomechanics, Vol. 38, pp. 323-331, 2005   DOI   ScienceOn
3 Wade, T. G., Harry, T. R., Allen, B., James, J. P., Robert, L. B. and Cecil, H. R., 'Revision total knee arthroplasty fixation with modular stems,' Clinical Orthopaedisc and Related Research, Vol. 404, pp. 158-168, 2002   DOI
4 Bae, D. K., Ahn, J. S. and Nha, K. N., 'Preliminary results of press-fit condylar knee,' Korean Knee Society, Vol. 1, No. 2, pp. 132-138, 1989
5 Hilding, M., Asplund. S. and Backbro, B., 'Alignment accuracy by two different instrumentation systems in total knee arthroplasty,' American Journal of Knee surgery, Vol. 6, p. 158, 1993
6 Anthony, G. A., Adrian, B. L., James, V. R. and Amirfazli, A., 'A parametric analysis of fixation post shape in tibial knee prostheses,' Medical Engineering & Physics, Vol. 27, pp. 123-134, 2005   DOI   ScienceOn
7 Taylor, M., Tanner, K. E. and Freeman, M. A. R., 'Finite element analysis of the implanted proximal tibia: a relationship between the initial cancellous bone stresses and implant migration,' Journal of Biomechanics, Vol. 31, pp. 303-310, 1998   DOI   ScienceOn
8 Barrack, R. L., Rorabeck, C., Burt, M. and Sawhney, J., 'Pain at the end of the stem after revision total knee arthroplasty,' Clinical Orthopaedics and related research, Vol. 367, pp. 216-25, 1999   DOI
9 Ramaniraka, N. A., Rakotomanana, L. R. and Leyvraz, P. F., 'The fixation of the cemented femoral component : Effects of stem stiffness, cemented thickness and roughness of the cement-bone surface,' The journal of bone and joint surgery(Br), Vol. 82-B, pp. 297-303, 2000
10 Simon, U., Augat, P., Ignatius, A. and Claes, L., 'Influence of the stiffness of bone defect implants on the mechanical conditions at the interface: a finite element analysis with contact,' Journal of Biomechanics, Vol. 36, pp. 1079-1086, 2003   DOI   ScienceOn
11 Kuster, M. S. and Stachowiak, G. W., 'Factors affecting polyethylene wear in total knee arthroplasty,' Orthopedics, Vol. 25, pp. 235-242, 2002
12 Fehring, T. K., Odum, S., Olekson, C., Griffin, W. L., Mason, J. B. and McCoy, T. H., 'Stem fixation in revision total knee arthroplasty,' Clinical orthopaedics and related research, Vol. 416, pp. 217-224, 2003   DOI
13 Robert, L. B., Tom, S., Mark, B. and Shelby, H., 'The effect of stem design on end-of-stem pain in revision total knee arthroplasty,' The Journal of Arthroplasty, Vol. 19, pp. 119-124, 2004   DOI   ScienceOn
14 James, A. R., Robert, T. T., Duane, M. I. and Scott, W. H., 'Factors affecting the durability of primary total knee prostheses,' The Journal of Bone and Joint Surgery(Am), Vol. 85, pp. 259-265, 2003   DOI
15 Taylor, M. and Tanner, K. E., 'Fatigue failure of cancellous bone: a possible cause of implant migration and loosening,' Journal of Bone and Joint Surgery, Vol. 79B, pp. 181-182, 1997   DOI
16 Lombardi, A. V. Jr., Berend, K. R. and Mallory, T, H., 'Pop-top tibial component elimination of back-side wear with potential modularity,' Surgical Technology International, Vol. 13, pp. 253-260, 2004