Browse > Article

A Study on the Location of Supporting Members in Monoleaflet Polymer Valve to Minimize Stress and Deformation  

Lee Seong Wook (동아대학교 대학원 기계공학과)
Shim Jae Joon (동아대학교 대학원 기계공학과)
Han Dong Seop (동아대학교 대학원 기계공학과)
Han Geun Jo (동아대학교 기계공학과)
Kim Tae Hyung (경남정보대학 기계자동차산업계열)
Publication Information
Abstract
A monoleaflet polymer artificial heart valve showed the remarkable improvement in pressure drop compared with other types of artificial valve. So, in this study we designed a monoleaflet polymer artificial valve with two supporting members to minimize the deformation and bending stress of the valve with respect to the variation of the gap between two supporting members using nonlinear contact analysis. The marginal valve thickness was also predicted in accordance with the relationship between the thickness and horizontal displacement in order to prevent the dislocation of the valve tip from the frame wall.
Keywords
Monoleaflet; Artificial heart valve; Supporting member; Vertical displacement; Horizontal displacement; Contact analysis;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Chandran, K. B., Lee, C. S., Shipkowitz, T., Chen, L. D., Yu, L. S. and Wurzel, D., 'In Vitro Hemodynamic Analysis of Flexible Artificial Ventricles,' Artificial Organs, Vol. 15, pp. 420-426, 1991
2 Kim, S. H., Chandran, K. B. and Chen, C. J., 'Numerical Simulation of Steady Flow in a Two-Dimensional Total Artificial Heart Model,' J. Biomech. Eng. ASME, Vol. 114, pp. 497-503, 1992   DOI
3 Kim, S. H., Kim, W. K., Jang, B. C. and Cho, B. K., 'A Simulative Circulation Test of Monoleaflet Polymer Valve,' Proc. of KOSOMBE, Vol. 15, pp. 51-53, May 1993
4 Han, G. J., Ahn, S. C., Shim, J. J. and Kim, S. Y., 'Nonlinear Analysis of the Monoleaflet Polymer Valve according to Shape of Supporting Members,' J. KSPE, Vol. 20, No.3, pp. 120-124, 2003   과학기술학회마을
5 ANSYS User's Manual, ANSYS Inc., 1992
6 Jung, S. J., 'Heat Transfer Analysis,' TSNE Inc., pp. 181-199, 2000
7 Akutsu, T., Dreyer, B. and Kolff, W. C., 'Polyurethane Artificial Heart Valves in Animals,' J. Appl. Physiol., Vol. 14, pp. 1045-1048, 1959
8 Chandran, K. B., Cabell, G. N., Khalighi, B. and Chen, C. J., ' Laser Anemometry Measurements of Pulsatile Flow Past Aortic Valve Prostheses,' J. Biomech., Vol. 16, pp. 865-873. 1983   DOI
9 Hilbert, S. L. and Jones, M., 'Evaluation of Explanted Polyurethane Trileatlet Cardiac Valve Prosthesis,' J. Thorac. Cardiovasc. Surg., Vol. 94, pp. 419-429, 1987
10 Harold, M., Lo, H. B., Reul, H., Muchter, H., Tauchi, K., Giersiepen, M., Birkle, G., Holiweg, G., Rau, G. and Messmer, B. J., 'The Helmholtz Institute Trileaflet Polyurethane Heart Valve Prosthesis : Design, Manufacturing, and First In Vitro and In Vivo Results. Polyurethanes in Biomedical Engineering II,' Elsevier, pp. 321-356, 1987
11 Yu, L. S., Yuan, B., Bishop, D., Topaz, S., Griensven, J. V., Hofma, S., Swier, P., Klinkmann, J., Kolff, J. and Kolff, W. J., 'New Polyurethane Valves in New Soft Artificial Hearts,' ASAIO Trans. Vol. 35, pp. 301-304, 1989   DOI   ScienceOn
12 Jansen, J., Willeke, S., Reiners, B., Harbott, H., Reul, H., Lo, H. B., Dabritz, S., Rosenbaum, C., Bitter, A., Ziehe, K., Rau, G. and Messmer, B. J., 'Advanced in Design Principle and Fluid Dynamics of a Flexible Polymeric Heart Valve,' ASAIO Trans., Vol. 57, pp. 451-453, 1991