Browse > Article

Design and Fabrication of FSK Transmitter for Miniaturized Wireless Endoscope  

장경만 (경북대학교 전기전자공학부)
문연관 (경북대학교 전기전자공학부)
류원열 (경북대학교 전기전자공학부)
윤영섭 (경북대학교 전기전자공학부)
조진호 (경북대학교 의과대학 의공학교실)
최현철 (경북대학교 전기전자공학부)
Publication Information
Abstract
The miniaturized wireless endoscope consists of CMOS Image sensor, FPGA, LED, Battery, DC to DC Converter, Antenna and Transmitter. FSK transmitter is designed and fabricated with 10 mm(diameter)${\times}$23 mm(thickness) dimension considering the maximum permission exposure(MPE), system size, power consumption, linearity and modulation method. Experimental results is - 3.67 dBm output power level, 20 MHz frequency deviation, and - 99 dBc/Hz(@100 kHz offset) phase noise at 1.2 GHz. From the in-vivo experiment, the designed FSK transmitter has a acceptable capability for wireless endoscope.
Keywords
Miniaturized Wireless Endoscope; FSK Transmitter;
Citations & Related Records
연도 인용수 순위
  • Reference
1 IEEE Standard For Satety Levels With Respect To Human Exposure To Radio Frequency Electromagnetic Fields, 3kHz To 300 GHz /
[ IEEE ] / IEEE Std. C95-1
2 /
[ 정보통신부 ] / 전자파인체보호기준, 정보통신부 고시 제2001-88호
3 /
[ Wayne Tomasi ] / Electronic Communication Systems
4 /
[] / Norika3
5 Visualization of Induced Current and SAR in Human's Head in Cellular Telecommunications /
[ M.Kacarska;G.L.Ololoska;S.Loskovska;L.Grcev ] / IEEE-APs Int'l Conference, Sessi-on40, Digest AP-40
6 Evaluating Compliance With FCC Guidelines For Human Exposure To Radio frequency Electromagnetic Fields /
[ FCC ] / OET Bulletin 65 Edition 97-01
7 /
[] / Given imaging
8 /
[] / 지능형 마이크로시스템 개발 사업단
9 Guidelines For Limiting Exposure To Time-Varying Electric, Magnetic, and Electromagnetic Fields(up to 300 GHz) /
[ ICNIRP ] / Health Physics
10 Safety Levels With Respect To Human Exposure To Radio Frequency Electromagnetic Fields, 300 kHz To 100 GHz /
[ ANSI ] / ANSI C95.1