Abstract
For the purpose of treating health checkups and recovery of patients in a super-aged society, hospitals use devices designed with a reduction circuit of electromagnetic waves associated with the specific absorption rate of electromagnetic waves absorbed by the human body. In this paper, we proposed a filter improvement design method capable of reducing electromagnetic waves. As a result of confirming the validity of the proposed technique through simulation and experimental results, the following result values were obtained. Applying the common-mode (CM) inductor 4 mH to a calibration circuit, noise decreased in a multiband spectrum. Using the differential mode(DM) inductor 40 µH element in the primary calibration circuit, the noise decreased by 15 dB or more in the 3 MHz band spectrum. Also, applying the Admittance Capacitance (Y-Cap) 10 nF element in the secondary calibration circuit resulted in the decrease by more than 30 dB in the band spectrum before 2 MHz. After using a common-mode inductor 4 mH element in the tertiary calibration circuit, it decreased by more than 15 dB in the band spectrum after 2 MHz.
Keywords
Electro-magnetic compatibility; Artifact; Driving system; Specific absorption rate; Medical image equipment; calibration circuit;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 N. Narendran, N. Maliyagoda, L. Deng, and R. M. Pysar, Proc. SPIE 4445, Solid State Lighting and Displays (SPIE, San Diego, USA, 2001). [DOI: https://doi.org/10.1117/12.450037]   DOI
2 J. N. Seo, Ph. D. Thesis, Chonnam National University, Gwangju (2009).
3 S. Ye, W. Eberle, and Y. F. Liu, IEEE Trans. Power Electron., 19, 1668 (2004). [DOI: https://doi.org/10.1109/TPEL.2004.836629]   DOI
4 H. W. Ott, Electromagnetic Compatibility Engineering (John Wiley & Sons, New York, 2009).
5 T. C. Neugebauer and D. J. Perreault, IEEE Trans. Power Electron., 21, 282 (2006). [DOI: https://doi.org/10.1109/TPEL.2005.861103]   DOI
6 D. A. Paice, Power Electronics Converter Harmonics, Multipulse Methods for Clean Power (Wiley-IEEE Press, 1991).
7 J. S. Kim, J. Korean Inst. Electr. Electron. Mater. Eng., 33, 151 (2020). [DOI: https://doi.org/10.4313/JKEM.2020.33.2.151]   DOI