Implementation of Wireless ECG Measurement System Attaching in Chair for Ubiquitous Health Care Environment  

Ye, Soo-Young (부산대학교 의학전문대학원 BK21 고급의료인력양성사업단)
Baik, Seong-Wan (부산대학교 의학전문대학원 마취통증의학교실)
Kim, Jee-Chul (부산대학교 의학전문대학원 의공학교실)
Jeon, Gye-Rok (부산대학교 의학전문대학원 의공학교실)
Publication Information
Journal of the Korean Institute of Electrical and Electronic Material Engineers / v.21, no.8, 2008 , pp. 776-781 More about this Journal
Abstract
In this study, ubiquitous health care system attaching in chair to monitor ECG for health care was developed at the unconsciousness state. The system conveniently and simple measured ECG at non-consciousness. We measured the contact impedance to skin-electrode of metal mesh electrodes of the system. Contact impedance enable the electrode to use for ECG measurement. The results are that the impedance of the metal mesh electrodes according to sizes is low when the size is 4$cm^2$. As the result, when the size of the metal mesh electrode is 4$cm^2$, the electrode is fit for ECG measurement. We can acquired by positing the arm on the metal mesh electrode. The ECG signal was detected using a high-input-impedance bio-amplifier, and then passed filter circuitry. The measured signal transmitted to a PC through the bluetooth wireless communication and monitored. Data of the non-constrained ECG system attaching in chair is noise-data when comparing metal mesh electrode with the Ag/Agcl electrode but the data is significant to monitor ECG for check the body state.
Keywords
Ubiquitous health care; Electrocardiogram (ECG); Metal mesh electrode; High-input-impedance bio-amplifier; Non-restrained biotelemetry;
Citations & Related Records
연도 인용수 순위
  • Reference
1 G. Troster, "The Agenda of Wearable Healthcare", IMIA Yearbook of Medical Informatics, 2005
2 M. J. Burke and D. T. Gleeson, "A micropower dry-electrode ECG preamplifier", IEEE Trans. BioMed. Eng., Vol. 47, No. 2, p. 155, 2000   DOI   ScienceOn
3 E. P. Scilingo, "Performance evaluation of sensing fabrics for monitoring physiological and biomechanical variables", IEEE Trans, BioMed, Vol. 9, No. 3, 2005
4 Y. K. Lim, K. K. Kim, and K. S. Park, "The ECG Measurement through Clothes with Capacitive Electrodes", in 6th Asian-Pacific Conference on Medical and Biological Engineering, 2005
5 I. Korhonen, J. Parkka, and M. V. Gils, "Health Monitoring in the of the Future", IEEE Engineering in Medicine and Biology Magazine, p. 66, 2003
6 G. E. Bergey, R. D. Squires, and W. C. Sipple, "Electrocardiogram recording with pastless electrodes", IEEE Trans. Biomed. Eng., Vol. BME-18, p. 206, 1971   DOI
7 W. H. Ko, M. R. Neuman, R. N. Wolfson, and E. T. Yon, "Insulated active electrodes", IEEE Trans. Ind. Elect. Contr. Instrum., Vol. IECI-17, p. 195, 1970   DOI   ScienceOn
8 C. Gondron, E. Siebert, P. Fabry, E. Novakov, and P. Y. Gumery, "Non-polarisable dry electrode based on NASICON ceramic", Med. Biol. Eng. Comput., Vol. 33, p. 452, 1995   DOI
9 지영준, "무구속 생체신호 측정 기술의 현황과 전망", 전자공학회지, 32권, 12호, p. 56, 2005
10 이충희, "생체신호계측 기술 정책", 기술동향분석보고, 한국과학기술정보연구원, 2003