Browse > Article
http://dx.doi.org/10.7471/ikeee.2016.20.3.310

Design of Tester Apparatus for 48 Channel GM Tube Sensor  

Lee, Hee-Yeol (Dept. Electronic Engineering, Hanbat National University)
Lee, Joo-Hyun (Dept. Electronic Engineering, Hanbat National University)
Lee, Seung-Ho (Dept. Electronics&Control Engineering, Hanbat National University)
Publication Information
Journal of IKEEE / v.20, no.3, 2016 , pp. 310-313 More about this Journal
Abstract
In this paper, we propose the tester apparatus for 48 channel GM Tube sensor. The proposed apparatus can test up to 48 channel GM tube simultaneously to detect the defect and analyze the sensor characteristic. 300-1000V variable high voltage generation circuit is utilized for the apparatus suitable for the sensor characteristic. Thus, the proposed system is useful for various GM Tube sensor characteristic analysis. Multiple sensor testing environment is established for the early detection of the defect and the analysis to reduce the costs for manufacturing and rework. Developed 48 channel GM Tube sensor test is evaluated with certified testing equipment and shows excellent performance with respect to the uncertainty of the sensor test results.
Keywords
Radiation Detection; GM Tube Sensor; Variable High Voltage Generator; Plateau; Measurement Uncertainty Rate;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 Jang, Kyeong-Uk, Joo-Hyun Lee, and Seung-Ho Lee, "A Study On Hardware Design for High Speed High Precision Neutron Measurement," Journal of IKEEE, pp. 61-67, 2016.
2 D. Barclay, "Improved Response of Geiger Muller Detectors.," IEEE Transactions on Nuclear Science, vo.33, pp. 613-616, Feb. 1986.   DOI
3 Gillan K. Godfrey, "Tracing Field Voles (Microtus Agrestis) with a Geiger-Muller Counter," ECOLOGY, vo. 35, no. 1, pp. 5-10, Jan. 1954.   DOI
4 Glenn F Knoll, Radiation detection and measurement, Wiley, 2010.
5 R. Andres Cortez, "Smart radiation sensor management," IEEE Robotics & Automation Magazine, vo. 15, no. 3, pp. 85-93, Sep. 2008.   DOI
6 H. S. Son, et al., "Design of Hybrid Supply Modulator for Reconfigurable Power Amplifiers," The Korean Institute of Electromagnetic Engineering and Science, vo. 23, no. 4, pp. 475-483, Apr. 2012.   DOI
7 Lee, Joo-Hyun, and Seung-Ho Lee, "A Study On Radiation Detection Using CMOS Image Sensor," Journal of IKEEE, pp. 193-200, 2015.