Browse > Article
http://dx.doi.org/10.5012/bkcs.2005.26.1.115

Determination of Water Content in Ethanol by Miniaturized Near-Infrared (NIR) System  

Cho, Soo-Hwa (Department of Chemistry, College of Natural Sciences, Hanyang University)
Chung, Hoe-Il (Department of Chemistry, College of Natural Sciences, Hanyang University)
Woo, Young-Ah (College of Pharmacy, Dongduk Women's University)
Kim, Hyo-Jin (College of Pharmacy, Dongduk Women's University)
Publication Information
Abstract
The miniaturized NIR (Near-infrared) spectrometer has been utilized for the determination of water content (1-19% range) in ethanol that is the most popular organic solvent in pharmaceutical industries. It has many potential capabilities that can replace the conventional analyzers especially for the on-line measurement since it is compact, versatile and cost-effective. By using two dimensional (2D) correlation spectroscopy, it was preliminarily investigated to find any unforeseen spectral distortion among the spectra collected from the miniaturized spectrometer. The 2D study revealed that the spectral variation clearly followed the variation of water concentration without any spectral distortion or abnormality. PLS (Partial Least Squares) was employed to build the calibration model and the resulting prediction performance was acceptable and stable over several days. Even though the miniaturized NIR system was evaluated to fairly simple chemical matrix, the overall study demonstrates the sufficient feasibility for diverse practical and industrial applications.
Keywords
Near-infrared spectroscopy; Miniaturized spectrometer; Ethanol solution; 2D correlation spectroscopy;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
Times Cited By Web Of Science : 12  (Related Records In Web of Science)
Times Cited By SCOPUS : 13
연도 인용수 순위
1 Burns, D. A.; Ciurczak, E. W. Handbook of Near-Infrared Analysis; Marcel Dekker Inc., New York, 1992
2 Noda, I. Appl. Spectrosc. 1993, 47, 1329   DOI   ScienceOn
3 Noda, I.; Dowrey, A. E.; Marcott, C.; Story, G. M.; Ozaki, Y. Appl. Spectrosc. 2000, 54, 236A   DOI
4 Noda, I. Appl. Spectrosc. 2000, 54, 994   DOI   ScienceOn
5 Thomas, E. V.; Haaland, D. M. Anal. Chem. 1990, 62, 1091   DOI
6 Adachi, D.; Katsumoto, Y.; Sato, H.; Ozaki, Y. Appl. Spectrosc. 2002, 56, 357   DOI   ScienceOn
7 Chung, H.; Lee, J. S.; Ku, M. S. Appl. Spectrosc. 1998, 52, 885   DOI   ScienceOn
8 Broad, N. W.; Jee, R. D.; Moffat, A. C.; Eaves, M. J.; Mann, W. C.; Dziki, W. Analyst 2000, 125, 2054   DOI   ScienceOn
9 Wetzel, D. L. Anal. Chem. 1983, 55, 1165A   DOI
10 Kamat, M. S.; Lodder, R. A.; DeLuca, P. P. Pharm. Res. 1989, 6, 961   DOI   ScienceOn
11 Woo, Y. A.; Kim, H. J. Useful and Advanced Information in the field of Near Infrared Spectroscopy; Research Signpost: Kerela, India, 2003; chap. 14
12 Mark, H. Anal. Chem. 1986, 58, 2814   DOI
13 Haaland, D. M.; Thomas, E. V. Anal. Chem. 1988, 60, 1202   DOI   ScienceOn
14 Martens, H.; Naes, T. M. Multivariate Calibration; John Wiley and Sons: New York, 1989
15 Beebe, K. R.; Pell, R. J.; Seasholtz, M. B. Chemometrics: A Practical Guide; John Wiley and Sons: New York, 1998
16 Lee, S. H.; Nam, J. J.; Son, B. M. Bulll. Korean Chem. Soc. 2003, 24(2), 246   DOI   ScienceOn