Cancellation of MRI Artifact due to Rotational Motion  

김응규 (한밭대학교 정보통신ㆍ컴퓨터공학부)
Abstract
When the imaging object rotates in image plane during MRI scan, its rotation causes phase error and non-uniform sampling to MRI signal. The model of the problem including phase error non-uniform sampling of MRI signal showed that the MRI signals corrupted by rotations about an arbitrary center and the origin in image plane are different in their phases. Therefore the following methods are presented to improve the quality of the MR image which includes the artifact. The first, assuming that the angle of 2-D rotational motion is already known and the position of 2-D rotational center is unknown, an algorithm to correct the artifact which is based on the phase correction is presented. The second, in case of 2-D rotational motion with unknown rotational center and unknown rotational angle, an algorithm is presented to correct the MRI artifact. At this case, the energy of an ideal MR image is minimum outside the boundary of the imaging object to estimate unknown motion parameters and the measured energy increases when the imaging object has an rotation. By using this property, an evaluation function is defined to estimate unknown values of rotational angle at each phase encoding step. Finally, the effectiveness of this presented techniques is shown by using a phantom image with simulated motion and a real image with 2-D translational shift and rotation.
Keywords
MRI; Non-uniform sampling; Rotational motion artifact; Artifact correction;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 An efficient algorithm for MRI image reconstruction without low spatial frequencies /
[ H.Yan;J.C.Gore ] / IEEE Trans. Med. Imaging
2 김함겸, 박재성, 이혜경, 김대호, 영상진단학, 대학서림, pp.55-74, 1997
3 이성우, 은충기, 문치웅, 박수성, 초기 자기 공명 영상학, 여문각, pp.53-61, 1998   과학기술학회마을
4 김청월, '펄스방식의 핵자기 공명장치에 관한 새로운 구현방법', 대한전자공학회논문지, 제35권 C편, 제10호, pp.771-781, 1998   과학기술학회마을
5 H. Yan and JC. Gore, 'An efficient algorithm for MRI image reconstruction without low spatial frequencies,' IEEE Trans. Med. Imaging, Vol.9, No.2, pp.30-36, 1992   DOI   ScienceOn
6 Z Cho. JP Jone, M Singh, Foundations of Medical Imaging, New York, John Wiley, pp.237-267, 1993
7 J. P. Felmlee, R. L. Ehman, S. J. Riederer, and H. W. Korin, 'Adaptive motion compensation in MR imaging without use of navigator echoes,' Radiology, Vol.179, pp.139-142,1991   과학기술학회마을
8 M. Hedley and H. Yan, 'Suppression of Slice Selection Axis Motion Artifact in MRI,' IEEE Trans. Med, Imaging, Vol. 11 , No.2, pp.233-237, 1992   DOI   ScienceOn
9 김응규, 권영도, '촬상단면내의 MRI 체동 아티팩트의 제거', 1999년도 대한전자공학회 하계종합학술대회 논문집, 22권, 제1호, pp.631-634, 1999   과학기술학회마을
10 김응규, 'MRI에 있어서 체동 아티팩트의 제거', 대한전자공학회논문지, SP, 제37권, 제3호, pp.70-78, 2000   과학기술학회마을
11 M. Hedley, H. Yan and D, Rosenfeld, 'Motion artifact correction in MRI using generalized projection,' IEEE Trans. Med. Imaging, Vol.10, No.1, pp.40-46, 1991   DOI   ScienceOn
12 M. Hedley, H. Yan and D. Rosenfeld, 'An improved algorithm for 2-D translational motion artifact correction,' IEEE Trans. Med, Imaging, Vol.10, No.4, pp,548-553, 1991   DOI   ScienceOn
13 M. Hedley and H. Yan, 'Iterative restoration of MR images corrupted with translational motion,' Journal of visual Communication Image Representation, Vol.3, No.4, pp.325-337, 1992   DOI
14 김응규, '촬상단면내 MRI 아티팩트의 제거', 한국정보과학회논문지:소프트웨어 및 응용, 제27권, 제4호, pp.432-440, 2000   과학기술학회마을
15 ML. Wood, MJ. Shivzi and PL. Stanchev, 'Planar Motion Correction with use of k-space data acquired in Fourier MR imaging,' Journal of MRI, Vol.5, No.1, pp.57-64, 1995   DOI   ScienceOn
16 RW. Cox, 'AFNI: Software for Analysis and Visualization of Functional Magnetic Resonance Neuroimages,' Computers and Biomedical Research, Vol.29, pp.162-173, 1996   DOI   ScienceOn
17 김응규, '위상 엔코딩 방향의 체동에 의한 MRI 아티팩트의 제거', 1998년도 한국정보과학회 가을 학술발표논문집, 제25권, 제2호, pp.432-434, 1998   과학기술학회마을
18 Eung-Kyeu Kim, Nohpill Park et al, 'Cancellation of MRI Motion Artifact in Image Plane,' IEEE Instrumentation and Measurement Technology Conference, Anchorage, Alaska, U.S.A, May, 2002   DOI
19 H. W. Korin, F. Farzane h, R.C. Wright, and S. J. Riederer, 'Compensation for effects of linear motion in MR imaging,' Magn. Reson. Med. Vol.12, pp.99-113, 1989   DOI
20 C. E. Floyd, R. J. Jaazczak and R. E. Coleman, 'Image resampling on a cylindrical sector grid,' IEEE Trans. Med. Imaging, Vol.MI-5, No.3, pp.128-131, 1986   DOI   ScienceOn
21 ML. Wood, MJ. Shivzi and PL. Stanchev, 'Planar Motion Correction with use of k-space data acquired in Fourier MR imaging,' Journal of MRI, Vol.5, pp.57-64, 1995   DOI   ScienceOn
22 JK. Reik, AM. Tekalp, Smith et al, 'Out-of-plane motion compensation in multislice spin-echo MRI,' IEEE Trans. Med. Imag., Vol.14, pp.464-470, 1995   DOI   ScienceOn
23 E. M. Haacke and G. W. Lenz, 'Improving MR image quality in the presence of motion by using rephasing gradients,' AJR 148, PP.1251-1258, 1987   DOI
24 T. Mitsa, K. J. Paker, W. E. Smith and A. M. Tekalp, 'Correction of Periodic Motion Artifacts Along the Slice Section Axis in MRI,' IEEE Trans. Med. Imaging, Vol.9, No.2, pp.310-317, 1990   DOI   ScienceOn
25 GH. Glover, AT. Lee 'Motion Artifacts in fMRI: Comparison of 2 DFT with PR and spiral scan methods,' Magnetic Resonance in Medicine, Vol. 33, pp.624-635, 1995   DOI
26 R. A. Zoroofi, Y. Sato et al, 'Reduction of MRI artifact due to rotational motion in the imaging plane,' in Proc.JAMIT Frontier 95', Nagoya Univ., pp. 37-44, 1995
27 R. L. Ehman and J. P. Felmlee, 'Adaptive technique for high-definition MR imaging of moving structures,' Radiology Vol.173, pp.255-263, 1989
28 D. J. Nishimura, Introduction to Magnetic Resonance Imaging, textbook EE 369B, Stanford Univ., Press, Stanford, CA, pp.156-167, 1994
29 이문호, C 언어 영상통신의 신호처리, 대영사, pp.158-180, 1998