Signal Change and Compensation of Pulse Pressure Sensor Array Due to Wrist Surface Temperature  

Jun, Min-Ho (KM Fundamental Research Division, Korea Institute of Oriental Medicine (KIOM))
Jeon, Young Ju (KM Fundamental Research Division, Korea Institute of Oriental Medicine (KIOM))
Kim, Young-Min (KM Fundamental Research Division, Korea Institute of Oriental Medicine (KIOM))
Publication Information
Journal of Sensor Science and Technology / v.26, no.2, 2017 , pp. 141-147 More about this Journal
Abstract
A pressure sensor in pulse measurement system is a core component for precisely measuring the pulse waveform of radial artery. A pulse sensor signal that measures the pulse wave in contact with the skin is affected by the temperature difference between the ambient temperature and skin surface. In this study, we found experimentally that the signal changes of the pressure sensors and a temperature sensor were caused by the temperature of the wrist surface while the pressure sensor was contacted on the skin surface for measuring pulse wave. To observe the signal change of the pulse sensor caused by temperature increase on sensor surface, Peltier device that can be kept at a set temperature was used. As the temperature of Peltier device was kept at $35^{\circ}C$ (the maximum wrist temperature), the device was put on the pulse sensor surface. The temperature and pressure signals were obtained simultaneously from a temperature sensor and six pressure sensors embedded in the pulse sensor. As a result of signal analysis, the sensor pressure was decreased during temperature increase of pulse sensor surface. In addition, the signal difference ratio of pressure and temperature sensors with respect to thickness of cover layer in pulse sensor was increased exponentially. Therefore, the signal of pressure sensor was modified by the compensation equation derived by the temperature sensor signal. We suggested that the thickness of cover layer in pulse sensor should be designed considering the skin surface temperature.
Keywords
pulse sensor array; pulse signal distortion; wrist temperature; pulse signal compensation; Peltier device;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Jun, M. H., Kim, Y. M., Bae, J. H., Jung, C. J., Cho, J. H., Jeon, Y. J., "Development of a Tonometric Sensor with a Decoupled Circular Array for Precisely Measuring Radial Artery Pulse", Sensors, Vol. 16, No. 6, 2016.
2 Wang, D. M., Zhang, D., Lu, G. M., "A Novel Multichannel Wrist Pulse System With Different Sensor Arrays", IEEE Trans. Instrum. Meas., Vol. 64, No. 7, pp. 2020-2034, 2015.   DOI
3 Joshi, A. B., Kalange, A. E., Bodas, D., Gangal, S. A., "Simulations of piezoelectric pressure sensor for radial artery pulse measurement", Mater. Sci. Eng. B Solid State Mater. Adv. Technol., Vol. 168, No. 1-3, pp. 250-253, 2010.   DOI
4 C. S. Hu, Y. F. Chung, C. C. Yeh, and C. H. Luo, "Temporal and spatial properties of arterial pulsation measurement using pressure sensor array", Evid. Based Complement Alternat. Med., Vol. 2012, Article ID 745127, 9 pages, 2012.
5 Lee, Sang-Suk, Myoung-Chone An, and Sung-Hoon Ahn, "A new measurement method of a radial pulse wave using multiple Hall array devices", Journal of Magnetics, Vol. 14, No. 3, pp. 132-136, 2009.   DOI
6 McLaughlin, J., Mcneil, M., Braun, B., and McCormack, P.D., "Piezoelectric sensor determination of arterial pulse wave velocity", Physiol. Meas., Vol. 24, No. 3, pp. 693-702, 2003.   DOI
7 Pressman, G. L., and P. M. Newgard, "A transducer for the continuous external measurement of arterial blood pressure", IEEE Trans. Biomed. Eng., Vol. 10, No. 2, pp. 73-81, 1963.
8 Yu-Wen Chu, Ching-Hsing Luo, Yu-Feng Chung, Chung-Shing Hu, Cheng-Chang Yeh, "Using an array sensor to determine differences in pulse diagnosis-Three positions and nine indicators.", Eur. J. Integr. Med., Vol. 6, No. 5, pp. 516-523, 2014.   DOI
9 전민호, 전영주, 김영민. "맥 센서 어레이 (array) 의 실리콘 (silicone) 코팅 두께에 따른 센서 간 간섭효과", J. Sensor Sci. & Tech., Vol. 25, No. 1, pp. 35-40, 2016.   DOI
10 Yoo, S. K., Shin, K. Y., Lee, T. B., Jin, S. O., and Kim, J. U., "Development of a radial pulse tonometric (RPT) sensor with a temperature compensation mechanism", Sensors, Vol. 13, No. 1, pp. 611-625, 2013.   DOI
11 https://en.tdk.eu/inf/57/ds/c33_barometric.pdf (retrieved on April 3, 2017)
12 Martinez-Nicolas A, Ortiz-Tudela E, Rol MA, Madrid JA, "Uncovering different masking factors on wrist skin temperature rhythm in free-living subjects", PloS One, Vol. 8, No. 4, e61142, 2013.   DOI