Browse > Article
http://dx.doi.org/10.5369/JSST.2013.22.6.415

Performance Assessment on Temperature Calibration Capability of the Calibration Laboratories Using High-Precision Platinum Resistance Thermometers  

Gam, Kee Sool (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science)
Lee, Young Hee (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science)
Yang, Inseok (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science)
Publication Information
Journal of Sensor Science and Technology / v.22, no.6, 2013 , pp. 415-420 More about this Journal
Abstract
Calibration capabilities for thermometer calibration by comparison method were assessed using high-precision industrial platinum resistance thermometers (IPRT). It was found in the performance assessment that out of 31 laboratories who participated, 28 laboratories resulted magnitude of En number less than 1 at every calibration points they submitted results in the range from 50 to $500^{\circ}C$. The results of about 75% of the laboratories showed the difference from the assigned values less than 1/10 of the tolerance level of the class B IPRT. This indicates that the participating calibration laboratories performed with satisfactory level that was enough to calibrate IPRTs to significant precision. The sensors used in this work were manufactured and chosen by the criteria of long-term instability less than 4 mK and hysteresis less than 8 mK in the temperature range used in this work. Furthermore, the change in the resistance of the sensors in the calibration temperature range were less than the uncertainty of the calibration, 25 mK (k=2).
Keywords
Calibration by comparison; Industrial platinum resistance thermometer; Performance assessment; Temperature calibration; Uncertainty;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Korean Laboratory accreditation scheme, KOLASG-004 : 2012.
2 Korean Laboratory accreditation scheme, KOLASR-001 : 2012.
3 Korean Laboratory accreditation scheme, KOLASR-003 : 2012.
4 Korean Agency for Technology and Standards, KS Q ISO/IEC 17043 : 2011, "Conformity assessment - General requirements for proficiency testing", 2011.
5 K. S. Gam, S. H. Yoo, S. M. Kim, and I. S. Lee, "Experimental method and evaluation of the calibration capability for the national calibration centers using the platinum resistance temperature sensors", J. Sensor Sci. & Tech., Vol. 14, No. 4, pp. 231-236, 2005.   과학기술학회마을   DOI   ScienceOn
6 K. S. Gam, Y.-G. Kim, and S. G. Kim, "Measurement of traceability error for calibration services center using type S thermocouples", J. Sensor Sci. & Tech., Vol. 4, No. 1, pp. 43-50, 1995.   과학기술학회마을
7 K. S. Gam, I. Yang, and Y. G. Kim, "Thermal hysteresis in thin-film platinum resistance thermometers", Int. J. Thermophys, Vol. 32, No. 11, pp . 2388-2396, 2011.   DOI
8 International Electrotechnical Commission, IEC 60751 ed2.0, "Industrial platinum resistance thermometers and platinum temperature sensors", 2008.
9 I. Yang, Y.-G. Kim, K. S. Gam, and Y. H. Lee, "Improved interpolating equation for industrial platinum resistance thermometer", J. Sensor Sci. & Tech., Vol. 21, No. 2, pp. 109-113, 2012.   과학기술학회마을   DOI   ScienceOn