Browse > Article
http://dx.doi.org/10.5369/JSST.2011.20.1.46

Realization of Swept Source-Optical Coherence Tomography using FDML Laser  

Eom, Jin-Seob (Department of Electronics Engineering, The School of Information Technology, Kangwon National University)
Publication Information
Abstract
In this paper, the swept source-optical coherence tomography system using frequency domain mode locked(FDML) laser has realized. The FDML swept source laser showed 55.03 kHz sweeping speed, 125 nm sweeping range, and 9 mW output optical power, which are the superiority of FDML laser compared to previous swept source lasers. Also, through the cross-sectional image captured at 5 frames per second for a mirror, a 1 mm-thickness glass plate, and a thumb bottom, the performance of the system has demonstrated.
Keywords
Swept source-Optical Coherence Tomography; Wavelength tunable filter; Dual balanced detector;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Wook Jae Yoo, Ji Yeon Heo, Dong Hyun Cho, Kyoung Won Jang, Jeong Ki Seo, Bongsoo Lee, Young-Ho Cho, Joo Hyun Moon, and Byung Gi Park, “Spectroscopic study on the development of fiber-optic pH sensor”, J. Kor. Sensors. Soc., vol. 18, no. 5, pp. 365-371, 2009.   과학기술학회마을   DOI   ScienceOn
2 David Huang and James G. Fujimoto, “Optical coherence tomography”, Science, vol. 254, pp. 1178-1181, 1991.   DOI
3 R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, “Performance of fourier domain vs. time domain optical coherence tomography”, Opt. Express, vol. 11, pp. 889-894, 2003.   DOI
4 M. A. Choma, M. V. Sarunic, C. Yang, and J. Izatt, “Sensitivity advantage of swept source and fourier domain optical coherence tomography”, Opt. Express, vol. 11, pp. 2183-2189, 2003.   DOI
5 R. Huber, M. Wojtkowski, K. Taira, J. G. Fujimoto, and K. Hsu, “Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging : design and scaling principles,” Opt. Express, vol. 13, pp. 3513-3528, 2005.   DOI
6 A. G. Podoleanu, “Unbalanced versus balanced operation in an optical coherence tomography system,” Appl. Opt. vol. 39, pp. 173-182, 2000.   DOI
7 Eugene Hecht, Optics, Addison-Wesley, 1987.
8 H. S. Kim, M. Y. Namgoong, J. R. Lee, and J. S. Eom, “Real of swept source optical coherence tomography system using loop mirror within reference arm”, Conference on Optoelectronics & Optical Communications(COOC) 2007, vol. 14, pp. 402-403, 2007.