Browse > Article
http://dx.doi.org/10.5369/JSST.2005.14.1.033

Development of PC-based and portable high speed impedance analyzer for biosensor  

Kim, Gi-Ryon (Dept. of Medical Engineering, College of Medicine, Dong-A University)
Kim, Gwang-Nyeon (Dept. of Medical Engineering, College of Medicine, Dong-A University)
Heo, Seung-Deok (Dept. of Medical Engineering, College of Medicine, Dong-A University)
Lee, Seung-Hoon (Dept. of Medical Engineering, College of Medicine, Dong-A University)
Choi, Byeong-Cheol (Dept. of Biomedical Engineering, Choonhae College)
Kim, Cheol-Han (Dept. of Biomedical Engineering, College of Medicine, Pusan National University)
Jeon, Gye-Rok (Dept. of Biomedical Engineering, College of Medicine, Pusan National University)
Jung, Dong-Keun (Dept. of Medical Engineering, College of Medicine, Dong-A University)
Publication Information
Abstract
For more convenient electrode-electrolyte interface impedance analysis in biosensor, a stand-alone impedance measurement system is required. In our study, we developed a PC-based portable system to analyze impedance of the electrochemical cell using microprocessor. The devised system consists of signal generator, programmable amplifiers, A/D converter, low pass filter, potentiostat, I/V converter, microprocessor, and PC interface. As a microprocessor, PIC16F877 which has the processing speed of 5 MIPS was used. For data acquisition, the sampling rate at 40 k samples/sec, resolution of 12 bit is used. RS-232 with 115.2 kbps speed is used for the PC communication. The square wave was used as stimuli signal for impedance analysis and voltage-controlled current measurement method of three-electrode-method were adopted. Acquired voltage and current data are calculated to multifrequency impedance signal after Fourier transform. To evaluate the implemented system, we set up the dummy cell as equivalent circuit of which was composed of resistor, parallel circuit of capacitor and resistor connected in parallel and measured the impedance of the dummy cell; the result showed that there exist accuracy within 5 % errors and reproduction within 1 % errors compared to output of Hioki LCR tester and HP impedance analyzer as a standard product. These results imply that it is possible to analyze electrode-electrolyte interface impedance quantitatively in biosensor and to implement the more portable high speed impedance analysis system compared to existing systems.
Keywords
electrochemical impedance spectroscopy; impedance analyzer; FFT;
Citations & Related Records
연도 인용수 순위
  • Reference
1 J. R. Macdonald, Impedance Spectroscopy, Willy- Interscience, New York, 1987
2 L. E. Baker, 'Some thoughts regarding electrical bioimpedance and its clinical use', J. Clin. Eng., vol. 18, pp. 221-3, 1993   DOI
3 A. J. Bard and Faulkner L. R., Electrochemical Methods, John Wiley & Sons, New York, 1880
4 J. O'M. Bockris and S. U. Khan, Surface Electrochemistry, Plenum Press, New York, 1993
5 황의진, 오상협, '푸리에 변환을 이용한 전기화학적 임피던스 측정 시스템 제작', 대한화학회지, 제35권, 제6호, pp. 713-719, 1991
6 백운기, 박수문, 계면과 전극과정의 과학 . 기술 전기화학, 청문각, 서울, 2001
7 A. Searle and L. Kirkup, 'Real time impedance plots with arbitrary frequency components', Physiol. Meas, vol. 20, pp. 103-114, 1999   DOI   ScienceOn
8 J. S. Yoo and S. M. Park, 'An Electrochemical Impedance Measurement Technique Employing Fourier Transform', Anal. Chem., vol. 72, pp. 2035-2041, 2000   DOI   ScienceOn
9 S. C. Creason and D. E. Smith, 'Fourier transform Faradic admittance measurements-on the use of high-precision data for characterization of very rapid electrode preocess kinetic parameter', Anal. Chem., vol. 45, pp. 2401-3, 1973   DOI
10 R. Gudivaka, D. Schoeller, and R. F. Kushner, 'Effect of skin temperature on multifrequency bio-electrical impedance analysis'. J. Appl. Phys., vol. 1, pp. 838-45, 1996
11 변수일, 재료전기화학개론, 시그마프레스, 서울, 2001
12 S. Grimnes, orjan Grottem Martinsen, Bioimped-ance and Bioelectricity Basic, Academic Press, New York, 2000