Implementation of Impedance Method to Estimate Blood Flow Variation with Cuff Pressure Change  

Jeong, Do-Un (Dept. of Interdisciplinary Program in Biomedical Engineering, Pusan National University)
Bae, Jin-Woo (Dept. of Interdisciplinary Program in Biomedical Engineering, Pusan National University)
Shon, Jung-Man (Dept. of Interdisciplinary Program in Biomedical Engineering, Pusan National University)
Yae, Su-Yung (Dept. of Interdisciplinary Program in Biomedical Engineering, Pusan National University)
Choi, Byeong-Cheol (Dept. of Biomedical Engineering, Choonhae College)
Nam, Ki-Gon (Dept. of Electronic Engineering, College of Engineering, Pusan National University)
Kim, Cheol-Han (Dept. of Electronic Engineering, College of Engineering, Pusan National University)
Jeon, Gye-Rok (Dept. of Biomedical Engineering, College of Medicine, Pusan National University)
Publication Information
Journal of Sensor Science and Technology / v.13, no.6, 2004 , pp. 462-472 More about this Journal
Abstract
In this study, we measured the blood flow on arm by non-invasive method and implemented a system to measure variation of the blood flow by estimating bio-electrical impedance and arterial pressure according to cuff pressure. The implemented system measured impedance variation according to pressure variation applied by artificial cuff pressure on the measuring position. The system consisted of pressure measuring part and impedance measuring part using 4-electrode method. Pressure measuring part was composed of semiconductor pressure sensor and electronic circuit for signal processing of sensor output signal. In addition, impedance measuring part was composed of constant current source circuit and lock-in amplifier for detecting impedance signal. We conducted experiments of impedance measuring part using standard resistance for performance evaluation of the implemented system. In addition we experimented to estimate variation of the blood flow by measuring impedances of the experimental group. We estimated ratio of the blood flow resistance using mean arterial pressure and variation of the blood flow. As a result the ratio of the blood flow resistance and variation of blood flow were in an inverse relationship with each other and the correlation coefficient was -0.96776.
Keywords
blood flow; blood pressure; impedance; constant current;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Nyboer, J., 'Electrical Impedance Plethysmography', Second Edition. Charles C. Thomas, Springfield, IL, 1970
2 Nyboer, J., Bango, S., Barnett, A., and Halsey, R. H., 'Tadiocardiograms-the Electrical Impedance Changes of the Heart in Relation to Electrocardiograms and Heart Sounds', J. Clin. Invest., vol. 19, no. 963, 1940
3 F. Risacher, J. Jossinet, and E. T. McAdams, et al., 'Impedance Plethysmography for The Evaluation of Pulse-Wave Velocity in Limbs', Med. & Biol. Eng. & Comput., vol. 31, pp. 318-322, 1993   DOI   ScienceOn
4 John G. Webster, 'Medical Instrumentation : Design and Application', Yeo Moon Kack, pp. 492-508, 1998
5 Yanqun Wang, David R. Haynor, and Yongmin Kim, 'A Finite-Element Study of the Effects of Electrode Position on the Measured Impedance Change in Impedance Cardiography', IEEE Trans. Biomed. Eng., vol. 48, no. 12, pp. 1390-1401, 2001   DOI   ScienceOn
6 정도운, 전계록, 배진우, 김길중, 심윤보, '휴대형 혈액가스분석 시스템의 구현 및 성능평가', 센서학회지 vol. 12, no.1, 2003
7 최병철, 정동근, 정도운, 노정훈, 전계록, '광전용적 맥파 센서를 이용한 맥파전달시간의 측정', 센서학회지, vol. 13, no. 5, 2003
8 Thomasett, 'A Bioelectrical Properties of Tissue Impedance', Lyon Med, vol. 107, no. 118, 1962
9 Kubicek, W. G., 'Development and Evaluation of an Impedance Cardiographic System to Measure Cardiac Output and Other Cardiac Parameters', National Aeronautics and Space Administratioll (NASA). July 1, 1968 to June 30, Contract no. NAS 9-4500, 1969
10 강두희, 생리학, 신광출판사, Ch. 8, 2000
11 Jae Kyu Cheun, 'Cardio-Pulmonary Physiology for the Clinicians', Koon fa Publishing Inc., pp. 57-68, 1996
12 Joseph J. Carr and John M. Brown, 'Introduction to Biomedical Equipment Technology', Kyung Moon, pp. 13-32, 2000
13 Nyboer, J., Bango, S., Barnett, A., and Halsey, R.H., 'Tadiocardiograms-the Electrical Impedance Changes of the Heart in Relation to Electrocardiograms and Heart Sounds', J. Clin. Invest., vol. 19, no. 963, 1940