Fabrication and Characteristics of Chromel-Constantan Multijunction Thermal Converter with Evanohm R Alloy Heater  

Lee, Young-Hwa (Department of Material Engineering, Chungnam National University)
Kwon, Sung-Won (Division of Electromagnetic Metrology, Korea Research Institute of Standards and Science)
Kim, Kook-Jin (Division of Electromagnetic Metrology, Korea Research Institute of Standards and Science)
Park, Se-Il (Division of Electromagnetic Metrology, Korea Research Institute of Standards and Science)
Ihm, Young-Eon (Department of Material Engineering, Chungnam National University)
Publication Information
Abstract
A thin-film multijunction thermal converter was fabricated through the process using 6 inch silicon wafer semiconductor process and bulk micromachining. Evanohm R alloy and chromel-constantan were used as a heater and thermocouple materials, respectively. The temperature coefficient of resistance of Evanohm R heater was about 75.12 ppm/$^{\circ}C$ and the voltage sensitivity of the thermal converter indicated about 5.75 mV/mW in air. The transfer differences, measured by FRDC-DC method in the frequency range from 20 Hz to 10 kHz, showed the value under about 1.36 ppm, 0.83 ppm for the film thickness of 500, 200 nm, respectively. And in case of a 200 nm-thick thermal converter, the AC-DC transfer differences seems to be stabilized below the value of 1 ppm in the frequency range from 1 kHz to 500 kHz.
Keywords
thermal converter; voltage sensitivity; FRDC-DC; AD-DC;
Citations & Related Records
연도 인용수 순위
  • Reference
1 T.F. Wunsch et al., 'A new fabrication process for planar thin-film multijunction thermal converters,' IEEE Trans. Instrum. Meas., vol. 50, no. 2, pp. 330-332, Apr. 2001   DOI   ScienceOn
2 http://www.hpmetals.com/technical.asp
3 T.A. Kovacs, 'Micromachined Transducers Sourcebook,' McGraw-Hill, New York, NY, pp. 77-83, 1998
4 http://www.goodfellow.com/csp/active/
5 정인식 등, '박막형 크로멜-알루멜 다중접합 열전변환기,' 전자공학회논문지, 제36권 D편, 제9호, pp. 37-45, 1999
6 M. Klonz et al., 'Measuring thermoelectric effects in thermal converters with a fast reversed DC,' IEEE Trans. Instrum. Meas., vol. 44, no. 2, pp. 379-382, Apr. 1995   DOI   ScienceOn
7 권성원 등, '교류-직류 변환오차 자동 측정시스템,' 센서학회지, 제7권, 제6호, pp. 25-32, 1998
8 H. Laiz et al., 'A simulation tool for the AC-DC transfer difference of thermal converters at low frequencies,' IEEE Trans. Instrum. Meas., vol. 48, no. 6, pp. 1155-1160, Dec. 1999   DOI   ScienceOn
9 조현덕 등, '저출력저항의 박막 크로멜-알루멜 다중접합 열전변환기,' 센서학회지, 제9권, 제4호, pp. 288-296, 2000
10 B.D. Inglis, 'Standards for AC-DC transfer,' Metrologia, vol. 29, pp. 191-199, 1992   DOI   ScienceOn
11 H. Dintner et al., 'AC-DC-mV-transfer with highly sensitive thin-film multijunction thermal Converters,' IEEE Trans. Instrum. Meas., vol. 42, no. 2, pp. 612-614, Apr. 1993   DOI   ScienceOn
12 M. Klonz and T. Weimann, 'Accurate thin film multijunction thermal converter on a silicon chip,' IEEE Trans. Instrum. Meas., vol. 38, no. 2, pp. 335-337, Apr. 1989   DOI   ScienceOn
13 J.R. Kinard et al., 'Development of thin-film multijunction thermal converters at NIST,' IEEE Trans. Instrum. Meas., vol. 46, no. 2, pp. 347-351, Apr. 1997   DOI   ScienceOn
14 김진섭 외, 'Bi-Sb 다중접합 열전변환기의 교류직류 변환 특성,' 전자공학회논문지, 제35권 D편, 제11호, pp. 46-54, 1998
15 K. Takahashi et al., 'AC-DC voltage transfer difference due to Seebeck effect in thermal Converters,' IEEE Trans. Instrum. Meas., vol. 48, no. 2, pp. 391-394, Apr. 1999   DOI   ScienceOn