Browse > Article
http://dx.doi.org/10.5668/JEHS.2018.44.5.491

Comparison of Human Blood Cadmium Concentrations using Graphite Furnace Atomic Absorption Spectrometry (GF-AAS) and Inductively Coupled Plasma-mass Spectrometry (ICP-MS)  

Kwon, Jung-Yeon (Heavy metal exposure Environmental Health Center, Dong-A University)
Kim, Byoung-Gwon (Heavy metal exposure Environmental Health Center, Dong-A University)
Lim, Hyoun-Ju (Heavy metal exposure Environmental Health Center, Dong-A University)
Seo, Jeong-Wook (Heavy metal exposure Environmental Health Center, Dong-A University)
Kang, Min-Kyung (Heavy metal exposure Environmental Health Center, Dong-A University)
Kim, Yu-Mi (Heavy metal exposure Environmental Health Center, Dong-A University)
Hong, Young-Seoub (Heavy metal exposure Environmental Health Center, Dong-A University)
Publication Information
Journal of Environmental Health Sciences / v.44, no.5, 2018 , pp. 491-501 More about this Journal
Abstract
Objectives: The aims of this study were to compare concentrations and the correspondence of human blood cadmium by using graphite furnace atomic absorption spectrometry (GF-AAS) and inductively coupled plasma-mass spectrometry (ICP-MS), which are representative methods of heavy metal analysis. Methods: We randomly selected 79 people who agreed to participate in the research project. After confirming the linearity of the calibration curves for GF-AAS and ICP-MS, the concentrations of cadmium in a quality control standard material and blood samples were measured, and the correlation and the degree of agreement were compared. Results: The detection limit of ICP-MS (IDL: $0.000{\mu}g/L$, MDL: $0.06{\mu}g/L$) was lower than that of GF-AAS (IDL: $0.085{\mu}g/L$, MDL: $0.327{\mu}g/L$). The coefficient of variation of the quality control standard material showed stable values for both ICP-MS (clinchek-1: 5.35%, clinchek-2: 6.22%) and GF-AAS (clinchek-1: 7.92%, clinchek-2: 5.22%). Recovery was relatively high for both ICP-MS (clinchek-1: 95.1%, clinchek-2: 92.8%) and GF-AAS (clinchek-1: 91.4%, clinchek-2: 98.8%), with more than 90%. The geometric mean, median, and percentile of blood samples were all similar. The agreement of the two instruments compared with the bias of the analytical values found that about 81% of the analytical values were within ${\pm}30%$ of the deviation from the ideal reference line (y=0). As a result of the agreement limit, the value included in the confidence interval was about 94%, which shows high agreement. Conclusion: In this study, we confirmed there was no significant difference in concentrations of a quality control standard material and blood samples. Since ICP-MS showed lower concentrations than GF-AAS at concentrations below the method detection limit of GF-AAS, it is expected that more precise results will be obtained by analyzing blood cadmium with ICP-MS.
Keywords
Cadmium; ICP-MS; GF-AAS; degree of agreement;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 Fukui Y, Ohashi F, Sakuragi S, Moriguchi J, Ikeda M. Comparative Evaluation of GFAAS and ICP-MS for Analyses of Cadmium in Blood. Ind Health. 2011; 49(3): 338-343.   DOI
2 Zhang ZW, Moon CS, Watanabe T, Shimbo S, He FS, Wu YQ, et al. Background exposure of urban populations to lead and cadmium: comparison between china and Japan. Int. Arch. Occup. Environ. Health. 1997; 69: 273-281.   DOI
3 Kim HJ, Hong YS, Lee KE, Kim DS, Lee MJ, Yeah BJ, et al. The Levels of Blood Lead and Cadmium in Urban and Rural Population in Korea. Journal of Life Science. 2009; 19(4): 472-478.   DOI
4 Lee JS, Kim YH, Park JS, Lee JH, Im TH, Kang HG, et al. Establishment of analytical method for blood and urine using ICP/MS(I). National Institute of Environmental Research. 2008.
5 Kim SH. Study of the exposure to heavy metals in the Korean population. Samsung Seoul Hospital. 2002.
6 Lee SY, Oh HJ, Choi YH, Kim JW, Kim SH. Trace Metal Analysis Using Inductively Coupled Plasma-Mass Spectrometry (ICP-MS). Korean J Lab Med. 2004; 24: 362-370.
7 Mortada WI, Sobh MA, el-Defrawy MM, Farahat SE. Reference intervals of cadmium, lead, and mercury in blood, urine, hair, and nails among residents in Mansoura city, Nile delta, Egypt. Environ Res. 2002; 90(2): 104-110.   DOI
8 Kristiansen J, Christensen JM, Iversen BS, Sabbioni E. Toxic trace element reference levels in blood and urine: influence of gender and lifestyle factors. Sci Total Environ. 1997; 204(2): 147-160.   DOI
9 Rahil-Khazen R, Bolann BJ, Ulvik RJ. Trace element reference values in serum determined by inductively coupled plasma atomic emission spectrometry. Clin Chem Lab Med 2000; 38: 765-772.
10 Khassouani CE, Soulaymani R, Mauras Y, Allain P. Blood cadmium concentration in the population of the Rabat area, Morocco. Clin Chim Acta. 2000; 302: 155-160.   DOI
11 Kim JM, Ahn JM, Kim WS, Kim JI, Shin HR, Jung KY, et al. Blood Lead, MAnganese, Aluminium and Silicon Concentrations in Korean Adults. Korean J Prev Med. 2000; 33(2): 157-164.
12 Lee CR, Lee JH, Kim SR, Jeong JY, Ryu CI. Urinary levels of arsenic, cadmium, and zinc of children in ulsan industrial area. Korean J Prev Med. 1999; 32(1): 1-8.
13 Jessilyn T, Rob DW, Fanny KE. Toxicological profile for cadmium. Agency for toxic substances and disease registry. 1999.
14 Zhang ZW, Shimbo S, Ochi N, Eguchi M, Watanabe T, Moon C-S, et al. Determination of lead and cadmium in food and blood by inductively coupled plasma mass spectrometry: a comparison with graphite furnace atomic absorption spectrometry. Sci Total Environ. 1997; 205: 179-187.   DOI
15 Jones DR, Jarrett JM, Tevis DS, Franklin M, Mullinix NJ, Wallon KL, et al. Analysis of whole human blood for Pb, Cd, Hg, Se, and Mn by ICPDRC- MS for biomonitoring and acute exposures. Talanta. 2017; 162: 114-122.   DOI
16 Trzcinka-Ochocka M, Brodzka R, Janasik B. Useful and Fast Method for Blood Lead and Cadmium Determination Using ICP-MS and GF-AAS; Validation Parameters. J. Clin. Lab. Anal. 2016; 30(2): 130-139.   DOI
17 Yedomon B, Menudier A, Etangs FLD, Anani L, Fayomi B, Druet-Cabanac M, et al. Biomonitoring of 29 trace elements in whole blood from inhabitants of Cotonou (Benin) by ICP-MS. J Trace Elem Med Biol. 2017; 43: 38-45.   DOI
18 Xiao M, Huang Z, Cai J, Jia J, Zhang Y, Dong W, et al. Comparison of different sample preparation methods for platinum determination in cultured cells by graphite furnace atomic absorption spectrometry. PeerJ. 2017.
19 Zhong WS, Ren T, Zhao LJ. Determination of Pb (Lead), Cd (Cadmium), Cr(Chromium), Cu (Copper), and Ni (Nickel) in Chinese tea with high-resolution continuum source graphite furnace atomic absorption spectrometry. Journal of food and drug analysis. 2016; 24(1): 46-55.   DOI
20 Soares AR, Nascentes CC. Development of a simple method for the determination of lead in lipstick using alkaline solubilization and graphite furnace atomic absorption spectrometry. Talanta. 2013; 105: 272-277.   DOI
21 Palmer CD, Lewis Jr ME, Geraghty CM, Barbosa Jr F, Parsons PJ. Determination of lead, cadmium and mercury in blood for assessment of environmental exposure: A comparison between inductively coupled plasma-mass spectrometry and atomic absorption spectrometry. Spectrochimica Acta Part B. 2006; 61: 980-990.   DOI
22 Al-Rmalli SW, Belazi AM, Mustafa HA, Abdulla AA. Blood cadmium concentrations in general population of Tripoli region, Libya. IOSR-JAC. 2015; 8(7): 76-80.
23 White MA. A comparison of inductively coupled plasma mass spectrometry with electrothermal atomic absorption spectrometry for determination of trace elements in blood and urine from non occupationally exposed populations. J Trace Elements Med Biol. 1999; 13: 93-101.   DOI
24 Heitland P, Koster HD. Biomonitoring of 37 trace elements in blood samples from inhabitants of northern Germany by ICP-MS. J Trace Elem Med Biol. 2006; 20(4): 253-262.   DOI
25 Bocca B, Forte G, Petrucci F, Senofonte O, Violante N, Alimonti A. Development of methods for the quantification of essential and toxic elements in human biomonitoring. Ann Ist Super Sanita. 2005; 41(2): 165-70.
26 Nixon DE, Moyer TP. Routine clinical determination of lead, arsenic, cadmium, and thallium in urine and whole blood by inductively coupled plasma mass spectrometry. Spectrochimica Acta Part B. 1996; 51: 13-25.   DOI
27 Heitland P, Koster HD. Fast, simple and reliable routine determination of 23 elements in urine by ICP-MS. J Anal At Spectrom, 2004; 19: 1552-1558.   DOI
28 Kira CS, Sakuma AM, Gouveia Nda C. Fast and simple multi-element determination of essential and toxic metals in whole blood with quadrupole ICPMS. J App Pharm Sci. 2014; 4(5): 39-45.
29 Torrence KM, McDaniel RL, Self DA, Chang MJ. Slurry sampling for the determination of arsenic, cadmium, and lead in mainstream cigarette smoke condensate by graphite furnace-atomic absorption spectrometry and inductively coupled plasma-mass spectrometry. Anal Bioanal Chem. 2002; 372: 723-731.   DOI
30 Madiha B, Khuram Shahzad A, Zahidqureshi, Nida M and Nimra. Determination of Heavy Metal Toxicity in Blood and Health Effect by AAS (Detection of Heavy Metals and its Toxicity in Human Blood). Arch Nano Op Acc J. 2018; 1(2): 22-28.
31 Chen X, Dai Y, Wang Z, Zhu G, Ding X, Jin T. The association between serum vitamin D levels and renal tubular dysfunction in a general population exposed to cadmium in China. PLOS ONE. 2018; 13(4): 1-11.
32 Lee JM, Seok KJ, Ryu JY, Jung WS, Park JB, Shin KH, et al. Association between Heavy Metal Exposure and Prevalence of Metabolic Syndrome in Adults of South Korea. Korean J Fam Pract. 2017; 7(2): 172-178.   DOI
33 Kim HJ, Garcia-Esquinas E, A Navas-Acien, Choi YH. Blood and urine cadmium concentrations and walking speed in middle-aged and older U.S. adults. Environmental Pollution. 2018; 232: 97-104.   DOI
34 Park HR, Kwun KS, Kim MH, Kim SK, Kim SY, Choi KS, et al. The Monitoring of Heavy Metals in Human Bloods of Middle School Students. J. Fd Hyg. Safety. 2005; 20(2): 83-88.
35 Lee MH. The Blood Cadmium and Lead Levels in Students. J. Exp. Biomed. Sci. 2006; 12: 371-374.
36 Oh JS, Lee SH. Pb, Hg and Cd Concentration of Blood and Exposure-Related Factors. Journal of the Korea Academia-Industrial cooperation Society. 2015; 16(3): 2089-2099.   DOI
37 Kim YA, Kim YN, Cho KD, Kim MY, Kim EJ, Baek OH, et al. Blood Heavy Metal Concentrations of Korean Adults by Seafood Consumption Frequency: Using the Fourth Korea National Health and Nutrition Examination Survey (KNHANES IV), 2008. Korean J Nutr. 2011; 44(6): 518-526.   DOI
38 Eom SY, Yim DH, Moon SI, Ochirpurev B, Choi YS, Park CH, et al. The Association of Blood Concentrations of Healvy Metals and Blood Pressure in Residents Living Near Janghang Copper Smelter in Korea. J Agric Med Community Health. 2017; 42(1): 13-23.   DOI
39 Seo JW, Kim BG, Kim YM, Choe BM, Seo SM, Hong YS. Associations of Blood Lead and Cadmium Levels with Hypertension using the Korea National Health and Nutrition Examination Survey III-VI. J Environ Health Sci. 2018; 44(4): 380-390.
40 Yoo IY. The Blood Levels of Lead, Mercury, and Cadmium and Metabolic Syndrome of Korean Adults. J. Korean Soc. Living Environ. Sys. 2014; 21(2): 251-259.   DOI
41 Lee KJ, Lim HB. Preparation and analysis of lyophilized whole blood as external quality control materials for Pb and Cd determination by graphite furnace atomic absorption spectrometry. Journal of the Korean Society of Analytical Sciences. 1995; 8(3): 134-140.
42 Im HS, Yang MH, Oh SL, Kim SY, Jung HJ, Cho YM, et al. Development and Validation of Reference Material for Proficiency Testing of Blood Heavy Metals in Environmental Biomonitoring. J Environ Health Sci. 2016; 42(6): 450-464.