Development and Performance Evaluation of an Animal SPECT System Using Philips ARGUS Gamma Camera and Pinhole Collimator  

Kim, Joong-Hyun (Department of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Jae-Sung (Department of Nuclear Medicine, Seoul National University College of Medicine)
Kim, Jin-Su (Department of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Byeong-Il (Department of Nuclear Medicine, Seoul National University College of Medicine)
Kim, Soo-Mee (Department of Nuclear Medicine, Seoul National University College of Medicine)
Choung, In-Soon (Department of Nuclear Medicine, Seoul National University College of Medicine)
Kim, Yu-Kyeong (Department of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Won-Woo (Department of Nuclear Medicine, Seoul National University College of Medicine)
Kim, Sang-Eun (Department of Nuclear Medicine, Seoul National University College of Medicine)
Chung, June-Key (Department of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Myung-Chul (Department of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Dong-Soo (Department of Nuclear Medicine, Seoul National University College of Medicine)
Publication Information
The Korean Journal of Nuclear Medicine / v.39, no.6, 2005 , pp. 445-455 More about this Journal
Abstract
Purpose: We developed an animal SPECT system using clinical Philips ARGUS scintillation camera and pinhole collimator with specially manufactured small apertures. In this study, we evaluated the physical characteristics of this system and biological feasibility for animal experiments. Materials and Methods: Rotating station for small animals using a step motor and operating software were developed. Pinhole inserts with small apertures (diameter of 0.5, 1.0, and 2.0 mm) were manufactured and physical parameters including planar spatial resolution and sensitivity and reconstructed resolution were measured for some apertures. In order to measure the size of the usable field of view according to the distance from the focal point, manufactured multiple line sources separated with the same distance were scanned and numbers of lines within the field of view were counted. Using a Tc-99m line source with 0.5 mm diameter and 12 mm length placed in the exact center of field of view, planar spatial resolution according to the distance was measured. Calibration factor to obtain FWHM values in 'mm' unit was calculated from the planar image of two separated line sources. Te-99m point source with i mm diameter was used for the measurement of system sensitivity. In addition, SPECT data of micro phantom with cold and hot line inserts and rat brain after intravenous injection of [I-123]FP-CIT were acquired and reconstructed using filtered back protection reconstruction algorithm for pinhole collimator. Results: Size of usable field of view was proportional to the distance from the focal point and their relationship could be fitted into a linear equation (y=1.4x+0.5, x: distance). System sensitivity and planar spatial resolution at 3 cm measured using 1.0 mm aperture was 71 cps/MBq and 1.24 mm, respectively. In the SPECT image of rat brain with [I-123]FP-CIT acquired using 1.0 mm aperture, the distribution of dopamine transporter in the striatum was well identified in each hemisphere. Conclusion: We verified that this new animal SPECT system with the Phlilps ARGUS scanner and small apertures had sufficient performance for small animal imaging.
Keywords
Animal SPECT; Pinhole collimator; Center of rotation error; Performance evaluation;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Kim YH, Lee DS, Kang JH, Lee YJ, Chung JK, Roh JK, et al. Reversing the silencing of reporter sodium/iodide symporter transgene for stem cell tracking. J Nucl Med 2005;46:305-11
2 So MK, Kang JH, Chung JK, Lee YJ, Shin JH, Kim KI, et al. In vivo imaging of retinoic acid receptor activity using a sodium/iodide symporter and luciferase dual imaging reporter gene. Mol Imaging 2004;3:163-71   DOI   ScienceOn
3 Kim JH, Lee JS, Kim JS, Lee BI, Kim SM, Lee WW, et al. Performance evlution of an animal SPECT system using Phillips ARGUS scintillation camera and pinhole collimator. J Nucl Med 2005;46[suppl]:250 [abstract]
4 Bai C, Shao L, Ye J, DurbinM. The Effects of Center of Rotation Errors on Cardiac SPECT Imaging. IEEE Trans Nucl Sci 2003;50:1528-35   DOI   ScienceOn
5 Li J, Jaszczak RJ, Greer KL, Coleman RE. A filtered backprojection algorithm for pinhole SPECT with a displaced centre of rotation. Phys Med Biol 1994;39:165-176   DOI   ScienceOn
6 Kim KI, Chung JK, Kang JH, Lee YJ, Shin JH, Oh HJ, et al. Visualization of endogenous p53-mediated transcription in vivo using sodium iodide symporter. Clin Cancer Res 2005;11:123-8   PUBMED
7 Cherry SR, Shao Y, Silverman RW, Meadors K, Siegel S, Chatziioannou A, et al. MicroPET: a high resolution PET scanner for imaging small animals. IEEE Trans Nucl Sci 1997;44:1161-6   DOI   ScienceOn
8 MacDonald LR, Patt BE, Iwanczyk JS, Tsui BMW, Wang Y, Frey EC et al. Pinhole SPECT of mice using the LumaGEM gamma camera. IEEE Trans Nucl Sci 2001;48:830-6   DOI   ScienceOn
9 Smith MF, Jaszczak RJ. An analytic model of pinhole aperture penetration for 3D pinhole SPECT image reconstruction. Phys Med Biol 1998;43:761-775   DOI   ScienceOn
10 Missimer J, Maldi Z, Honer M, Keller C, Schubiger A, Ametamey S-M. Performance evaluation of the 16-module quad-HIDAC small animal PET camera. Phys Med Biol 2004;49:2069-81   DOI   ScienceOn
11 Shin JH, Chung JK, Kang JH, Lee YJ, Kim KI, So Y, et al. Noninvasive imaging for monitoring of viable cancer cells using a dual-imaging reporter gene. J Nucl Med 2004;45:2109-15
12 Beekman FJ, van der Have F, Vastenhouw B, van der Linden AJ, van Rijk PP, Burbach JP, et al. U-SPECT-I: a novel system for submillimeter-resolution tomography with radiolabeled molecules in mice. J Nucl Med 2005;46:1194-200   PUBMED
13 Tai C, Chatziioannou A, Siegel S, Young J, Newport D, Goble RN, et al. Performance evaluation of the microPET P4: a PET system dedicated to animal imaging. Phys Med Biol 2001;46:1845-1862   DOI   ScienceOn
14 Vanhove C, Defrise M, Franken PR, Everaert H, Deconinck F, Bossuyt A. Interest of the ordered subsets expectation maximization (OS-EM) algorithm in pinhole single-photon emission tomography reconstruction: a phantom study. Eur J Nucl Med 2000;27:140-6   DOI   ScienceOn
15 National Electrical Manufacturers Association: NEMA standards publication NU1-2001: performance measurements of scintillation cameras. Rosslyn, VA, National Electrical Manufacturers Association, 2001
16 Wu MC, Gao DW, Sievers RE, Lee RJ, Hasegawa BH, Dae MW. Pinhole single-photon emission computed tomography for myocardial perfusion imaging of mice. J Am Coll Cardiol 2003;42:576-82   DOI   ScienceOn
17 Wu MC, Hasegawa BH, Dae MW. Performance evaluation of a pinhole SPECT system for myocardial perfusion imaging of mice. Med Phys 2000;29:2830-9
18 Knoess C, Siegel S, Smith A, Newport D, Richerzhagen N, Winkeler A, et al. Performance evaluation of the microPET R4 PET scanner for rodents. Eur J Nucl Med Mol Imaging 2003 May;30:737-47   DOI   PUBMED   ScienceOn