Quantitative Differences between X-Ray CT-Based and $^{137}Cs$-Based Attenuation Correction in Philips Gemini PET/CT  

Kim, Jin-Su (Departments of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Jae-Sung (Departments of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Dong-Soo (Departments of Nuclear Medicine, Seoul National University College of Medicine)
Park, Eun-Kyung (Departments of Nuclear Medicine, Seoul National University College of Medicine)
Kim, Jong-Hyo (Departments of Radiology, Seoul National University College of Medicine)
Kim, Jae-Il (Departments of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Hong-Jae (Departments of Nuclear Medicine, Seoul National University College of Medicine)
Chung, June-Key (Departments of Nuclear Medicine, Seoul National University College of Medicine)
Lee, Myung-Chul (Departments of Nuclear Medicine, Seoul National University College of Medicine)
Publication Information
The Korean Journal of Nuclear Medicine / v.39, no.3, 2005 , pp. 182-190 More about this Journal
Abstract
Purpose: There are differences between Standard Uptake Value (SUV) of CT attenuation corrected PET and that of $^{137}Cs$. Since various causes lead to difference of SUV, it is important to know what is the cause of these difference. Since only the X-ray CT and $^{137}Cs$ transmission data are used for the attenuation correction, in Philips GEMINI PET/CT scanner, proper transformation of these data into usable attenuation coefficients for 511 keV photon has to be ascertained. The aim of this study was to evaluate the accuracy in the CT measurement and compare the CT and $^{137}Cs$-based attenuation correction in this scanner. Methods: For all the experiments, CT was set to 40 keV (120 kVp) and 50 mAs. To evaluate the accuracy of the CT measurement, CT performance phantom was scanned and Hounsfield units (HU) for those regions were compared to the true values. For the comparison of CT and $^{137}Cs$-based attenuation corrections, transmission scans of the elliptical lung-spine-body phantom and electron density CT phantom composed of various components, such as water, bone, brain and adipose, were performed using CT and $^{137}Cs$. Transformed attenuation coefficients from these data were compared to each other and true 511 keV attenuation coefficient acquired using $^{68}Ge$ and ECAT EXACT 47 scanner. In addition, CT and $^{137}Cs$-derived attenuation coefficients and SUV values for $^{18}F$-FDG measured from the regions with normal and pathological uptake in patients' data were also compared. Results: HU of all the regions in CT performance phantom measured using GEMINI PET/CT were equivalent to the known true values. CT based attenuation coefficients were lower than those of $^{68}Ge$ about 10% in bony region of NEMA ECT phantom. Attenuation coefficients derived from $^{137}Cs$ data was slightly higher than those from CT data also in the images of electron density CT phantom and patients' body with electron density. However, the SUV values in attenuation corrected images using $^{137}Cs$ were lower than images corrected using CT. Percent difference between SUV values was about 15%. Conclusion: Although the HU measured using this scanner was accurate, accuracy in the conversion from CT data into the 511 keV attenuation coefficients was limited in the bony region. Discrepancy in the transformed attenuation coefficients and SUV values between CT and $^{137}Cs$-based data shown in this study suggests that further optimization of various parameters in data acquisition and processing would be necessary for this scanner.
Keywords
attenuation coefficient; PET/CT; SUV;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Zaidi H, Hasegawa B. Determination of the attenuation map in emission tomography. J Nucl Med 2003;44:291-315
2 Kinahan PE, Hasegawa BH, Beyer T. X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 2003;33:166-79   DOI   ScienceOn
3 Watson CC, Schaefer A, Luk WK, Kirsch CM. Clinical evaluation of single-photon attenuation correction for 3D whole-body PET. IEEE Trans Nucl Sci 1999;46:1024-31   DOI   ScienceOn
4 Daube-Witherspoon ME, Matej S, Karp JS, Lewitt RM. Application of the row action maximum likelihood algorithm with spherical basis functions to clinical PET imaging. IEEE Trans Nucl Sci 2001;48:24-30   DOI   ScienceOn
5 Nakamoto Y, Osman M, Cohade C, Marshall LT, Links JM, Kohlmyer S et al,. PET/CT: comparison of quantitative tracer uptake between germanium and CT transmission attenuationcorrected images. J Nucl Med 2002;43:1137-43   PUBMED
6 National Electrical Manufacturers Association: NEMA Standards Publication NU2-2001: Performance Measurements of Positron Emission Tomographs. Rosslyn, VA, National Electrical Manufacturers Association, 2001
7 Burger C, Goerres G, Schoenes S, Buck A, Lonn AH, Von Schulthess GK. PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients. Eur J Nucl Med Mol Imaging 2002;29:922-7   DOI   PUBMED   ScienceOn
8 Valk PE, Bailey DL, Townsend DW. Maisey MN. Positron Emission Tomography: Basic Science and Clinical Practice.: Springer (U.S.A); 2002, p.138-41
9 Cherry SR, Sorenson JA, Phelps ME. Physics in nuclear medicine. 3rd ed.: Elsevier Science (U.S.A.); 2003. p.84
10 Kinahan PE, Townsend DW, Beyer T, Sashin D. Attenuation correction for a combined 3D PET/CT scanner. Med Phys 1998;25: 2046-53   DOI   ScienceOn
11 Nehmeh SA, Erdi YE, Pan T, Pevsner A, Rosenzweig KE, Yorke E et al,. Four-dimensional (4D) PET/CT imaging of the thorax. Med Phys 2004;31:3179-86   DOI   ScienceOn
12 Kim JS, Lee JS, Lee BI, Lee DS, Chung J-K, Lee MC. Performance Characteristics of 3D GSO PET/CT Scanner (Philips GEMINI PET/CT). Korean J Nucl Med 2004;38:318-32
13 Karp JS, Muehllehner G, Qu H, Yan XH. Singles transmission in volume-imaging PET with a $^{137}Cs$ source. Phys Med Biol 1995;40: 929-944.   DOI   ScienceOn