Browse > Article

Evaluation of methods for estimating the pulse reflection site with cardiovascular simulator  

Lee, Ju-Yeon (Dept. of East-West Medical Engineering, Graduate school, Sangji University)
Shin, Sang-Hoon (Depart. of Oriental Biomedical Engineering, Sangji University)
Publication Information
The Journal of the Society of Korean Medicine Diagnostics / v.19, no.1, 2015 , pp. 47-54 More about this Journal
Abstract
Objective Wave reflection is an important factor that determines the shape of the pulse wave. The purpose of this study is to compare the conventional method used for estimating the reflection site of pulse with a cardiovascular simulator. Methods: In this study, cardiovascular simulator with one elastic tube was used. The pressure and flow was measured simultaneously at three different points. The measured data were used to the conventional methods to estimate the pulse wave reflection site. The results were compared with the known length which were the distances from the measured points to the end of tube. Results & Conclusions: There is a significant error with the time domain method. While, the reflection site with the frequency domain method was similar to the actual reflection site.
Keywords
Pulse wave; Pulse reflection site; Cardiovascular simulator;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Murgo JP, Westerhoff N, Giolma JP, Altobelli SA. Aortic input impedance in normal men: relationship to pressure wave forms. Circulation. 1980; 62: 105-116.   DOI
2 Nichols WW, Nicolini FA, Pepine CJ. Determinants of isolated systolic hypertesion in the elderly. Journal of Hypertension. 1992; 10: S73-S77.
3 Newman DL, Masters NJ, McNulty JF. Pulse-wave reflection in elastic tubes. Medical and biological engineering. 1975; 13: 720-726.   DOI
4 Segers P, Verdonck P. Role of tapering in aortic wave reflection: hydraulic and mathematical model study. Journal of biomechanics. 2000; 33: 299-306.   DOI
5 Khir AW, Parker KH. Measurements of wave speed and reflected waves in elastic tubes and bifurcations. Journal of biomechanics. 2002; 35: 775-783.   DOI
6 Lee JY, Jang M, Shin SH. Mock circulatory system with a silicon tube for the study of pulse waves in an arterial system. Journal of the Korean Physical Society, 2014; 65(7): 1134-1141.   DOI
7 Legendre D, Fonseca J, Andrade A, Biscegli JF, Manrique R, Guerrino D, Lucchi JC. Mock circulatory system for the evaluation of left ventricular assist devices, endoluminal prostheses, and vascular diseases. Artificial organs. 2008; 32: 461-467.   DOI
8 Westerhof N, Stergiopulos N, Noble MI. Snapshots of hemodynamics: an aid for clinical research and graduate education. Springer Science & Business Media, New York, USA, 2010, 109-110.
9 W. W. Nichols, C. R. Conti, W. E. Walker and W. R. Milnor, "Input impedance of the systemic circulation in man," Circulation Research, vol. 40, no. 5, pp. 451-458, 1977.   DOI
10 S. Laxminarayan, "The calculation of forward and backward waves in the arterial system," Medical and Biological Engineering and Computing, vol. 17, no. 1, pp. 130-130, 1979.   DOI
11 O'Rourke MF, Cartmill TB: Influence of aortic coarctation on pulsatile hemodynamics in the proximal aorta. Circulation 44: 281, 1971.
12 Mills CJ, Gabe IT, Gault JH, Mason DT, Ross J Jr, Braunwald E, Shillingford JP. Pressure flow relationships and vascular impedance in man. Cardiovasc Res. 1970; 4: 405.   DOI
13 Zannoli R, Corazza I, Branzi A. Mechanical simulator of the cardiovascular system. Physica Medica. 2009; 25: 94-100.   DOI
14 권선민, 강희정, 임윤경, 이용흠, 김윤경, 이용홈. 5 단계 가압 맥파측정에 의한 연령별 혈관 경화도 분석. Korean Journal of Acupuncture. 2010; 27: 107-120.