Browse > Article
http://dx.doi.org/10.6109/jkiice.2007.11.12.2343

Development of Medical System using Time-varying Magnetic field  

Kim, In-Soo S. (한국전기연구원 융합기술연구단)
Hong, Jung-Hwang ((주)엠말 기술연구소)
Abstract
In this paper, we describe the design and implementation of time-varying magnetic field stimulator. Novel design for power supply part to generate high repetitive magnetic field for very short time(less than $300{\mu}s$) was achieved. Using the novel designed power supply part and the circular type coil probe, we've achieved high magnetic field up to 1.2 Tesla in 20Hz repetition rate.
Keywords
Magnetic field; Power supply system; Induced magnetic coil probe;
Citations & Related Records
연도 인용수 순위
  • Reference
1 M.S. George, S.H. Lisanby, H.A. Sackeim, 'Transcranial Magnetic Stimulation,' Psychiatry, 1999
2 J.V. Escudero, J. Sancho, D. Bautista, M Escudero, J. Lpez-Trigo,'Prognostic Value of Motor Evoked Potential Obtained by Transcranial Magnetic Brain Stimulation in Motor Function Recovery in Patients With Acute Ischemic Stroke,' Stroke (J. of the American Heart Association), 1998
3 K. Davey, 'Magnetic Stimulation Coil and Circuit Design,' IEEE Trans. on Biom. Eng., Nov. 2000
4 Z, J. Daskalakis, B.K. Christensen, P.B. Fitzgerald, R. Chen, 'Transcranial Magnetic Stimulation: A New Investigational and Treatment Tool in Psychiatry,' J. Neuropsychiatry Clin. Neurosci. 14:4, 2002
5 D. McRobbie, 'Design and instrumentation of a magnetic nerve stimulatior,' J.Phys. E, 1985
6 M.F. del Olmo, Olalla Bello, Javier Cudeiro, 'Transcranial magnetic stimulation over dorsolateral prefrontal cortex in Parkinson's disease,' Clinical Neurophysiology, 2007