Evaluation of Robust Classifier Algorithm for Tissue Classification under Various Noise Levels  

Youn, Su Hyun (Department of Bio-Mechatronic Engineering, Sungkyunkwan University)
Shin, Ki Young (Department of Energy and Power Conversion Engineering, University of Science & Technology)
Choi, Ahnryul (Department of Bio-Mechatronic Engineering, Sungkyunkwan University)
Mun, Joung Hwan (Department of Bio-Mechatronic Engineering, Sungkyunkwan University)
Publication Information
ETRI Journal / v.39, no.1, 2017 , pp. 87-96 More about this Journal
Abstract
Ultrasonic surgical devices are routinely used for surgical procedures. The incision and coagulation of tissue generate a temperature of $40^{\circ}C-150^{\circ}C$ and depend on the controllable output power level of the surgical device. Recently, research on the classification of grasped tissues to automatically control the power level was published. However, this research did not consider the specific characteristics of the surgical device, tissue denaturalization, and so on. Therefore, this research proposes a robust algorithm that simulates noise to resemble real situations and classifies tissue using conventional classifier algorithms. In this research, the bioimpedance spectrum for six tissues (liver, large intestine, kidney, lung, muscle, and fat) is measured, and five classifier algorithms are used. A signal-to-noise ratio of additive white Gaussian noise diversifies the testing sets, and as a result, each classifier's performance exhibits a difference. The k-nearest neighbors algorithm shows the highest classification rate of 92.09% (p < 0.01) and a standard deviation of 1.92%, which confirms high reproducibility.
Keywords
Ultrasonic surgical device; Power level auto-control; Bioimpedance; Classifier algorithm; Additive white Gaussian noise;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 N.S. Amarin, "Harmonic Scalpel and Clipless Cholecystectomy," World J. Laparoscopic Surgery, vol. 1, no. 2, Aug. 2008, pp. 6-8.
2 C. Sutton, G.D. Adamson, and K.D. Jones, Modern Management of Endometriosis, Boca Raton, FL, USA: CPC press, 2005.
3 Ethicon Endo-Surgery, Generator 300 System Service Manual. http://www.frankshospitalworkshop.com/equipment/documents/electrosurgery/service_manuals/Ethicon_Harmonic_300_Generator_-_Service_manual.pdf
4 H. Kalvoy et al., "Impedance-Based Tissue Discrimination for Needle Guidance," Physiological Meas., vol. 30, no. 2, Jan. 2009, pp. 129-140.   DOI
5 D.A. Dean et al., "Electrical Impedance Spectroscopy Study of Biological Tissues," J. Electrostat., vol. 66, no. 3, Mar. 2008, pp. 165-177.   DOI
6 M. Simic, "Realization of Complex Impedance Measurement System Based on the Integrated Circuit AD5933," Telecommun. Forum, Belgrade, Serbia, Nov. 26-28, 2013, pp. 573-576.
7 R.O. Duda, P.E. Hart, and D.G. Stork, Pattern classification, Hoboken, NJ, USA: John Wiley & Sons, 2012.
8 V. Lepetit and P. Fua, "Keypoint Recognition Using Randomized Trees," IEEE Trans. Pattern Anal. Mach. Intell., vol. 28, no. 9, Sept. 2006, pp. 1465-1479.   DOI
9 N. Druzijanic et al., "Comparison of Lateral Thermal Damage of the Human Peritoneum using Monopolar Diathermy, Harmonic Scalpel and LigaSure," Canadian J. Surgery, vol. 55, no. 5, Oct. 2012, pp. 317-321.   DOI
10 T. Carus and K. Rackebrandt, "Collateral Tissue Damage by Several Types of Coagulation (monopolar, bipolar, cold plasma and ultrasonic) in a Minimally Invasive, Perfused Liver Model," ISRN Surgery, vol. 2011, May 2011, pp. 1-7.
11 B.J. O'Daly et al., "High-Power Low-Frequency Ultrasound: A Review of Tissue Dissection and Ablation in Medicine and Surgery," J. Mater. Process. Technol., vol. 200, no. 1, May 2008, pp. 38-58.   DOI
12 S.J. Lee and S.J. Park, "Ultrasonic Energy in Endoscopic Surgery," Yonsei Med. J., vol. 40, no. 6, Dec. 1999, pp. 545-549.   DOI
13 D. Sun et al., "Development and Application of Ultrasonic Surgical Instruments," IEEE Trans. Biomed. Eng., vol. 44, no. 6, June 1997, pp. 462-467.   DOI
14 H.R. Kim et al., "Classification of Organs Using Impedance of Ultrasonic Surgical Knife to Improve Surgical Efficiency," J. Biomed. Eng. Res., vol. 34, no. 3, Sept. 2013, pp. 141-147.   DOI
15 C.W. Hsu and C.J. Lin, "A Comparison of Methods for Multiclass Support Vector Machines," IEEE Trans. Neural Netw., vol. 13, no. 2, Mar. 2002, pp. 415-425.   DOI
16 P.F. Vetere and C. Apostolis, "Ureteral Injury Due to a Harmonic Scalpel during Laparoscopic Salpingo-oophorectomy," J. Soc. Laparoendoscopic Surgeons, vol. 14, no. 1, Mar. 2010, pp. 115-119.   DOI
17 C. Ying, Z. Zhaoying, and Z. Ganghua, "Effects of Different Tissue Loads on High Power Ultrasonic Surgery Scalpel," Ultrasound Med. Biol., vol. 32, no. 3, Mar. 2006, pp. 415-420.   DOI
18 M. Fernandez-Corazza, N. Von Ellenrieder, and C.H. Muravchik, "Spatial Filtering in Electrical Impedance Tomography," J. Physics. Conf., Joinville, Brazil, Nov. 6-9, 2012, pp. 1-6.
19 S.H. Youn et al., "Multi-class Biological Tissue Classification Based on a Multi-classifier: Preliminary Study of an Automatic Output Power Control for Ultrasonic Surgical Units," Comput. Boil. Med., vol. 61, no. 1, June 2015, pp. 92-100.   DOI
20 J.S. Dadd, T.P. Ryan, and R. Platt, "Tissue Impedance as a Function of Temperature and Time," Biomed. Sci. Instrum., vol. 32, Feb. 1996, pp. 205-214.
21 A. Kiourti, "Biomedical Telemetry: Communication between Implanted Devices and the External World," Opticon 1826, no. 8, Apr. 2010, pp. 1-7.
22 K.W. Jung et al., "Prediction of Cancer Incidence and Mortality in Korea, 2014," Cancer Res. Treatment, vol. 46, no. 2, Apr. 2014, pp. 124-130.   DOI
23 S. Schmidbauer et al., "Experience with Ultrasound Scissors and Blades (UltraCision) in Open and Laparoscopic Liver Resection," Anna. Surgery, vol. 235, no. 1, Jan. 2002, pp. 27-30.   DOI
24 A. Choi et al., "Electrical Properties of Four Kinds of Organs of the Susscrofadomesticus from Impedance Measurement," Proc. Korean Soc. Precis. Eng. Conf., Jeju, Rep. of Korea, May 13-15, 2015, pp. 954-955.
25 C.C. Chung et al., "Double-Blind, Randomized Trial Comparing Harmonic Scalpel (TM) Hemorrhoidectomy, Bipolar Scissors Hemorrhoidectomy, and Scissors Excision: Ligation Technique," Diseases Colon Rectum, vol. 45, no. 6, June 2002, pp. 789-794.   DOI
26 C.K. Phillips et al., "Tissue Response to Surgical Energy Devices," Urology, vol. 71, no. 4, Apr. 2008, pp. 744-748.   DOI
27 C. Chen et al., "The Effects of Ultrasonic and Electrosurgery Devices on Nerve Physiology," Britsh J. Neurosurgery, vol. 26, no. 6, June 2012, pp. 856-863.   DOI