Abstract
A new type of diagnostic membranes based on methoxyethoxy and trifluoroethoxy co-substituted polyphosphazene has been prepared to measure blood glucose level of diabetics. Final absorbances at 680 nm through activated polyphosphazene membranes were measured at various concentration of glucose in plasma or blood. The end-point results of varing absorbance values as time (K/S) was found to have a linear relationship toward the blood glucose concentration. The effects of substitution rates with hydrophilic groups and hydrophobic groups on the measurements of glucose concentration were studied. Dose-response slope (DRS) values between glucose concentration and K/S values increased as the hydrophilic substitution rates increased. However, in more than 30% of the substitution rates, it was difficult to measure exact concentration level of glucose because DRS increased rapidly.
Keywords
polyphosphazene; membranes; diabetics; glucose;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 S. Kwon, I. Park, and D. Yoon, "The Preparation of Polyacrylonitrile Diagonostic Membranes for Blood Glucose Measurements(1): Effects of Temperature and Humidity on the Measurements of Glucose Concentration", Membrane Journal, 17, 4 (2007).
2 H. R. Allcock and S. Kwon, "Covalent Linkages of Proteins to Surface-Modified Poly(organophosphazenes): Immobilization of Glucose-6-Phosphate Dehydrogenase and Trypsin", Macromolecules, 19, 1502 (1986).   DOI
3 S. Kwon and C. Jun, "Studies on Polyphosphazenes- Bound Wittig Reactions", J. of Korean Ind. & Eng. Chem, 5, 5 (1994).
4 H. R. Allcock, M. Gebura, S. Kwon, and T. X. Neenan, "Amphiphilic Polyphosphazenes as Membrane Materials: Influence of Side Group on Radiation Crosslinking, Semipermeability, and Surface Morphology", Biomaterials, 9, 500 (1988).   DOI   ScienceOn
5 H. R. Allcock and S. Kwon, "Hydrophilic Polyphosphazenes as Hydrogels: Radiation Crosslinking and Hydrogel Characteristics of Poly[bis (methoxyethoxy)phosphazene]", Biomaterials, 7, 6 (1986).
6 S. Kwon and B. Lee, "A Study on the Preparation of Metal-Ion Separation Membrane with Hydrophilic Polyphosphazene", J. of Korean Ind. & Eng. Chem, 10, 3 (1999).
7 S. Kwon and B. Lee, "A Study on the Preparation of Ion-Exchange Membranes with Polyphosphazenes", J. of Korean Ind. & Eng. Chem, 9, 3 (1998).
8 S. Kwon, "Preparation of Polyurethane Diagnostic Membranes for Blood Glucose Measurements(6): Effects of Hematocrit on the Measurements of Glucose Concentration", Membrane Journal, 20, 3 (2010).
9 S. Kwon, "Studies on the Preparation of Diagnostic Membranes for Blood Glucose Measurements (3) : Effects of Hematocrit on the Measurements of Glucose Concentration", Hongik Industrial Technology, 14 (2004).
10 R. P. Back, "Biosensor Technology", Marcel Dekker, New York (1990).
11 S. Kwon, I. Park, and D. Yoon, "Studies on the Preparation of Polyurethane Diagnostic Membranes for Blood Glucose Measurements(5): Effects of Temperature and Humidity on the Measurements of Glucose Concentration", Membrane Journal, 17, 1 (2007).
12 S. Kwon, "Studies on the Preparation of Diagnostic Membranes for Blood Glucose Measurements (2): Effects of Temperature on the Measurements of Glucose Concentration", Hongik Industrial Technology, 13 (2003).
13 S. Kwon, "Basic Studies on the Preparation of Diagnostic Membranes by Using Multi-Layered Gelatin Films to Measure Blood Glucose Level of Diabetics", Membrane Journal, 8, 1 (1998).
14 S. Kwon and B. Lee, "Studies on the Multi-Layered Gelatin Diagnostic Membranes for Diabetes (1): Effects of Temperature and Humidity on the Diffusion-Controlled Rates of Glucose", Membrane Journal, 9, 2 (1992).
15 R. E. Kesting, "Synthetic Polymeric Membranes", Wiley-Interscience, New York (1985).
16 M. Gordon, "Polymer Membranes", Springer-Verlag, New York (1985).
17 S. Kwon, "Studies on the preparation of Diagnostic Membranes fir Blood Glucose Measurements (1) : Model Experiments by Using Microporous Polyurethane Membrane", Hongik Industrial Technology, 12 (2002).
18 S. Kwon, "Studies on the Polyurethane Diagnostic Membrane for Diabetes(2): Effects of Additives in Membrane Formulations for the Measurement of Urine Glucose", Polymer (Korea), 18, 6 (1994).
19 S. Kwon, "Studies on the Preparation of Diagnostic Membranes for Blood Glucose Measurements (4): Effects of Additives in Blood on the Measurements of Glucose Concentration", Hongik Industrial Technology, 15 (2005).
20 S. Kwon and J. Yu, "The Preparation of Polyacrylonitrile Diagnostic Membranes for Blood Glucose Measurements(3): Effects of Storage Environments on the Measurements of Glucose Concen- tration", Membrane Journal, 19, 3 (2009).
21 S. Kim, "Life Guide for Diabetes Control", Hyoseong, Seoul (1993).
22 T. Asano and T. Norioka, "Diabetes", Nihon Bungeisha, Tokyo (2000).
23 J. Kim, "Diabetes : Diagnostics and Control", Ohsung, Seoul (1997).
24 T. Asano, "Diabetic Food and Life Style", Shufu- To-Seikatsusha Co., Tokyo (2005).
25 Home Health Management Research Institute, "Diabetes", Kumyoung, Seoul (1992).
26 S. Kwon, "A Study on the Preparation of Polyurethane Diagnostic Membrane for Urine Glucose Test", J. Kor. Ind. & Eng. Chem., 5, 6 (1994).
27 S. Kwon and M. Choi, "The Preparation of Polyacrylonitrile Diagnostic Membranes for Blood Glucose Measurements(2) : Effects of Blood constituens on the Measurements of Glucose Concentration", Membrane Journal, 18, 4 (2008).
28 K. Huh, "Diabetic Health", D and C Media, Seoul (2006).
29 S. Kwon, "Studies on the Muti-Layered Gelatin Diagnostic Membranes for Diabetes(2): Effects of Interferents in Blood on the Diffusion-Controlled Rates of Glucose", Membrane Journal, 9, 4 (1999).