The Comparative Study of on Pump CABG during Pulsatile $(T-PLS^{TM})$ and Nonpulsatile $(Bio-pump^{TM})$ Perfusion  

Park Young-Woo (Department of Thoracic and Cardiovascular Surgery, College of Medicine, Soonchunhyang University)
Her Keun (Department of Thoracic and Cardiovascular Surgery, College of Medicine, Soonchunhyang University)
Lim Jae-Ung (Department of Thoracic and Cardiovascular Surgery, College of Medicine, Soonchunhyang University)
Shin Hwa-Kyun (Department of Thoracic and Cardiovascular Surgery, College of Medicine, Soonchunhyang University)
Won Yong-Soon (Department of Thoracic and Cardiovascular Surgery, College of Medicine, Soonchunhyang University)
Publication Information
Journal of Chest Surgery / v.39, no.5, 2006 , pp. 354-358 More about this Journal
Abstract
Background: Pulsatile pumps for extracorporeal circulation have been known to be better for tissue perfusion than non-pulsatile pumps but be detrimental to blood corpuscles. This study is intended to examine the risks and benefits of $T-PLS^{TM}$ through the comparison of clinical effects of $T-PLS^{TM}$ (pulsatile pump) and $Bio-pump^{TM}$ (non-pulsatile pump) used for coronary bypass surgery. Material and Method: The comparison was made on 40 patients who had coronary bypass using $T-PLS^{TM}\;and\;Bio-pump^{TM}$ (20 patients for each) from April 2003 to June 2005. All of the surgeries were operated on pump beating coronary artery bypass graft using cardiopulmonary extra-corporeal circulation. Risk factors before surgery and the condition during surgery and the results were compared. Result: There was no significant difference in age, gender ratio, and risk factors before surgery such as history of diabetes, hypertension, smoking, obstructive pulmonary disease, coronary infarction, and renal failure between the two groups. Surgery duration, hours of heart-lung machine operation, used shunt and grafted coronary branch were little different between the two groups. The two groups had a similar level of systolic arterial pressure, diastolic arterial pressure and mean arterial pressure, but pulse pressure was measured higher in the group with $T-PLS^{TM}\;(46{\pm}15\;mmHg\;in\;T-PLS^{TM}\;vs\;35{\pm}13\;mmHg\;in\;Bio-pump^{TM},\;p<0.05)$. The $T-PLS^{TM}$-operated patients tended to produce more urine volume during surgery, but the difference was not statistically significant $(9.7{\pm}3.9\;cc/min\;in\;T-PLS^{TM}\;vs\;8.9{\pm}3.6\;cc/min\;in\;Bio-pump^{TM},\;p=0.20)$. There was no significant difference in mean duration of respirator usage and 24-hour blood loss after surgery between the two groups. Plasma free Hb was measured lower in the group with $T-PLS^{TM}\;(24.5{\pm}21.7\;mg/dL\;in\;T-PLS^{TM}\;versus\;46.8{\pm}23.0mg/dL\;in\;Bio-pump^{TM},\;p<0.05)$. There was no significant difference in coronary infarction, arrhythmia, renal failure and morbidity rate of cerebrovascular disease. There was a case of death after surgery (death rate of 5%) in the group tested with $T-PLS^{TM}$, but the death rate was not statistically significant. Conclusion: Coronary bypass was operated with $T-PLS^{TM}$ (Pulsatile flow pump) using a heart-lung machine. There was no unexpected event caused by mechanical error during surgery, and the clinical process of the surgery was the same as the surgery for which $Bio-pump^{TM}$ was used. In addition, $T-PLS^{TM}$ used surgery was found to be less detrimental to blood corpuscles than the pulsatile flow has been known to be. Authors of this study could confirm the safety of $T-PLS^{TM}$.
Keywords
Coronary artery bypass; Cardiopulmonary bypass; Pulsatile flow; Perfusion;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 Wrigh G. Mechanical stimulation of cardiac function by means of pulsatile blood pumps. J Cardiothorac Vasc Anes 1997;299-309
2 ÜUndar A, Masai T, Fraser OH, Fraser CD Jr. Pulsatile and nonpulsatile flows can be quantified in terms of energy equivalent pressure during cardiopulmonary bypass for direct comparisons. ASAIO J 1999;45:610-4   DOI
3 ÜUndar A. Myths and truths of pulsatile and nonpulsatile perfusion during acute and chronic cardiac support. Artif Organs 2004;28:439-43   DOI   ScienceOn
4 Lee JJ, Lim CH, Son HS, et al. In vitro evaluation of the performance of korean pulsatile ECLS (T-PLS) using precise quantification of pressure-flow waveforms. ASAIO J 2005; 51:604-8   DOI   ScienceOn
5 Son HS, Sun K, Fang YH, et al. The effects of pulsatile versus non-pulsatile extracorporeal circulation on the pattern of coronary artery blood flow during cardiac arrest. Int J Artif Organs 2005;28:609-16   DOI
6 Wright G, Sum Ping JTS, Campbell CS, et al. Assessment of cardiovascular function in patients undergoing coronary artery bypass grafting. J Thorac Cardiovasc Surg 1988;96: 400-7
7 Rho YR, Choi H, Lee JC, et al. Applications of the pulsatile flow versatile ECLS: in vivo studies. Int J Artif Organs 2003;26:428-35   DOI
8 Gibbon JH. Application of a mechanical heart and lung apparatus to cardiac surgery. Minn Med 1954;37:171-85
9 Dodrill FD, Hill E, Gerisch R. Some physiologic aspects of the artificial heart problem. J Thorac Surg 1952;24:134-50
10 Shepard RB, Simpson DC, Sharp JF. Energy equivalent pressure. Arch Surg 1966;93:730-40   DOI   ScienceOn
11 ÜUndar A, Rosenberg G, Myers JL. Major factors in the controversy of pulsatile versus nonpulsatile flow during acute and chronic cardiac support. ASAIO J 2005;51:173-5   DOI   ScienceOn
12 Lim CH, Sun K, Son HS, et al. A study of optimal model for the circuit configuration of korean pulsatile extracorporeal life support system (T-PLS). Korean J Thorac Cardiovasc Surg 2005;38:13-22   과학기술학회마을
13 Lim CH, Son HS, Lee JJ, et al. Optimization of the circuit confuguration of a pulsatile ECLS: an in vivo experimental Study. ASAIO J 2005;51:609-13   DOI   ScienceOn
14 Hooker DR. A study of the isolated kidney: the influence of pulse pressure upon renal function. Am J Physiol 1910;27: 24-44
15 Zuckerman BD, Yin FC. Aortic impedance and compliance in hypertensive rats. Am J Physiol 1989;257:553-62
16 Undar A, Koenig KM, Frazier OH, Fraser Jr CD. Impact of membrane oxygenators on pulsatile versus nonpulsatile perfusion in a neonatal model. Perfusion 2000;15:111-20   DOI   ScienceOn
17 Wrigh G. Hemodynamic analysis could resolve the pulsatile blood flow controversy. Ann Thorac Surg 1994;58:1199-204   DOI   ScienceOn
18 Zumbro GL, Shera G, Fishback ME, Galloway RF. A prospective evaluation of the pulsatile assist device. Ann Thorac Surg 1979;25:269-72
19 Kim HK, Son HS, Fang YH, et al. The effects of pulsatile flow on renal tissue perfusion during cardiopulmonary bypass: a comparative study of pulsatile and nonpulsatile flow. ASAIO J 2005;51:30-6   DOI   ScienceOn