Abstract
In this paper, two calibration methods for angular vibration pickups using a precision rotary encoder are proposed. The KRISS (Korea Research Institute of Standards and Science) primary angular vibration calibration system and the calibration procedures are briefly explained. The rotary encoder is shown to be calibrated in two methods: The one is to use the laser interferometer to calibrate the rotary encoder under test and the other is to exploit the certificate of the encoder supplied. Complex sensitivities measured from the first are shown to be less than 0.1 % difference in magnitude and $0.01^{\circ}$ difference in phase shift in reference to those of the primary calibration system. Their expanded uncertainties were observed to be less than 0.6 % in magnitude and $0.4^{\circ}$ in phase shift over the range of 0.4 to 200 Hz. Under the same calibration conditions, complex sensitivities evaluated by the second method are shown be 0.1 % difference in magnitude and $0.6^{\circ}$ difference in phase shift in reference to those of the primary calibration system. Their expanded uncertainties were seen to be less than 4.8 % in magnitude and $2.8^{\circ}$ in phase shift.
Keywords
Angular accelerometer calibration; Angular vibration; Angular vibration pickups; Rotary encoder; Measurement uncertainty;
Citations & Related Records
연도 인용수 순위
  • Reference
1 H. Martens, "Dynamic cailbration of rotational transducers by diffraction grating interferometry," PTB Mitt. 103, 19-26 (1992).
2 ISO 16063 part 21, Methods for the calibration of vibration and shock transducers: Vibration calibration by comparison to a reference transducer, 2001.
3 ISO 16063 part 15, Methods for the calibration of vibration and shock transducers: Primary angular vibration calibration by laser interferometry, 2006.
4 S. Nam, Y. Lee and W. Cheung "Method of calibrating charge amplifier using a dynamic spectrum analyzer," IMEKO 20th Congress, TC22, (2012).
5 W. Cheung and T. Licht, "Progress in development of calibration systems for angular vibration pickups," IMEKO 19th Congress, TC22, (2009).
6 W. Cheung and C. Chung, "Angle prism-based laser interferometer for high precision measurement of angular vibration," IMEKO 18th Congress, TC22, (2006).
7 W. Cheung "Improved features of angular vibration exciters," IMEKO 20th Congress, TC22, (2012).
8 W. Cheung, "Primary calibration procedure of angular vibration transducers," KRISS., Tech. Rep., 2011.
9 A. Ellin, and G. Dolsak. "The design and application of rotary encoders." Sens Rev. 28, 150-158 (2008).   DOI   ScienceOn
10 J. Bendat and A. Piersol, Random Data (WILEY Publication, INC., 2010), pp. 102-105.