Browse > Article
http://dx.doi.org/10.7776/ASK.2010.29.8.476

Influences of the Defective Piezo-Elements of a Medical Ultrasonic Probe on Transient Acoustic Fields and B-Mode Images  

Choi, Kwang-Yoon (부경대학교 대학원 음향진동공학과(협))
Ha, Kang-Lyeol (부경대학교 물리학과)
Kim, Moo-Joon (부경대학교 물리학과)
Kim, Jung-Soon (동명대학교 멀티미디어공학과)
Yang, Jeong-Hwa (제주한라대학 방사선과)
Kang, Gwan-Suk (제주대학교 의학과 의공학협동과정)
Choi, Min-Joo (제주대학교 의학과 의공학협동과정)
Abstract
The ultrasonic transient fields and B-mode images of a point target which were simulated for a medical ultrasonic probe with a few defective piezo-elements were compared with those for a normal probe. The present study considered a 3.5 MHz linear array probe whose acoustic beam was formed by the 64 active elements of total 192 elements. The results showed that the maximum amplitude and -3 dB width of the acoustic fields by main-lobes decreased linearly as the defective element number increased from one to four. However, the depths of foci remained almost unchanged, and the pressure differences between main-lobes and side-lobes tended to decrease due to rise in pressures in side-lobes. Such changes in ultrasonic fields affected the B-mode images of point targets. So the artifacts were formed in the right and left side of the target, and the lateral spatial resolutions were decreased while the axial resolution was almost the same.
Keywords
Ultrasonic Imaging; Medical Ultrasound Probe; Defective Element; Transient Acoustic Field; Spatial Resolution;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 L. J. Ziomek, Fundamentals of acoustic field theory and space-time signal processing, CRC press Inc., Boca Raton, 1995.
2 J. W. Goodman, Introduction to Fourier optics, McGraw-Hill Inc., New York, 2ed, 1996.
3 B. Weigang, G. W. Moore, J. Gessert, W. H. Phillips, and M. Schafer, "The methods and effects of transducer degradation on image quality and the clinical efficacy of diagnostic sonography," Jour. Diag. Med. Sonography, vol. 19, no. 1, pp. 3-13, 2003.   DOI
4 양정화, 이경성, 강관석, 팽동국, 최민주, "초음파프로브의 결함이 B-모드 영상의 공간 분해능에 미치는 영향," 한국음향학회 2009년도 춘계학술발표대회 논문집, 28권, 1(s)호, 176-179쪽, 2009.
5 M. Martensson, M. Olsson, B. Segall, A. G. Fraser, R. Winter, and L. Brodin, "High incidence of defective ultrasound transducers in use in routine clinical practice," Euro. Jour. Echocardiography, vol. 10, pp. 389-394, 2008.
6 G. S. Kino, Acoustic waves: devices, imaging, and analog signal processing, Prentice-Hall, Inc., Englewood Cliffs, pp. 154-390, 1987.
7 R. N. Thurston and A. D. Pierce, Ultrasonic instruments and devices I : Reference for modern instrumentation techniques, and technology, Physical Acoustics, vol. XXIII, Academic Press, San Diego, pp. 46-147, 1999.
8 T. A. Shoup and J. Hart, "Ultrasonic imaging system," Proc. IEEE Ultrason. Symp., pp. 863-871, Oct. 1988.
9 J.A. Jensen and N. B. Svendsen, "Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers," IEEE Trans. Ultrason. Ferroelec. Freq. Contr., vol. 39, pp. 262-267. 1992.   DOI   ScienceOn
10 M. Fatemi and A.C. Kak, "Ultrasonic b-scan imaging: Theory of image formation and a thecnique for restoration," Ultrason. Imaging, vol. 2, pp. 1-47, 1980.   DOI   ScienceOn
11 유병철, 최흥호, 노시철, 민해기, 권장우, "채널별 음장분포 분석을 통한 진단용 초음파 어레이 프로브의 평가방법에 관한 연구," 의공학회지, 27권, 229-236쪽, 2006   과학기술학회마을
12 P. Cormbie, A.Bascom, and S. Cobbold, "Calculating the pulsed response of linear arrays: Accuracy vs. computational efficiency," IEEE Trans. Ultrason. Ferroelect. Freq. Contr., vol. 44, pp. 997-1009, 1997.   DOI   ScienceOn
13 박은주, 송행용, 하강렬, 김무준, 김동현, 이수성, "과도음장 해석을 통한 초음파 진단 탐촉자의 성능 개선," 한국음향학회지, 21권 8호, 744-756쪽, 2002.   과학기술학회마을
14 J.A. Jensen, "Field: A program for simulating ultrasound systems." medical & biological engineering & computing, vol. 34, Supplement 1, Part 1, pp. 351-353, 1996.   DOI