Browse > Article
http://dx.doi.org/10.7776/ASK.2008.27.6.263

Optimal Design and Analysis of a Medical Imaging Ultrasonic Array Sensor  

Kim, Hoe-Yong (경북대학교 기계공학과)
Roh, Yong-Rae (경북대학교 기계공학과)
Abstract
The performance of an ultrasonic array sensor is determined by the properties of constituent materials and the effects of many structural parameters. In this study, with the finite element method, variation of the performances of an ultrasonic array sensor was analyzed in relation to its structural variables. Based on the analysis result, the structure of the ultrasonic array sensor was optimized to provide the highest sensitivity while satisfying such requirements as fractional bandwidth, center frequency and -20 dB pulse length. The optimization was carried out with the SQP-PD method for a target function composed of the ultrasonic array sensor performance. The optimized ultrasonic array sensor satisfied all the required specifications to be applicable to medical imaging diagnosis. The design technology in this paper can be utilized for other ultrasonic array sensors of a similar structure.
Keywords
Ultrasonic; Array sensor; Optimization; FEM;
Citations & Related Records
연도 인용수 순위
  • Reference
1 K. J. Kang and Y. R. Roh, "Optimization of structural variables of a flextensional transducer by the statistical multiple regression analysis method," J. Acoust. Soc. Am. 114(3), 1454 -1461, 2003   DOI   ScienceOn
2 D. A. Pierre, Optimization theory with applications, (Dover Publications, New York, 1986)
3 J. A. Jensen and N. B. Svendsen, "Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers." IEEE Trans. Ultrason., Ferroelec., Freq. Contr. 39, 262-267, 1992   DOI   ScienceOn
4 R. J. Freund and W. J. Wilson, Regression analysis: Statistical modeling of a response variable, (Academic Press, San Diego, 1998)
5 R. E. McKeighen, "Design Guidelines for Medical Ultrasonic Arrays," SPIE 3341, 2-18, 1998   DOI
6 J. Larson,"Non-ideal radiations in phased array transducers," IEEE Ultras. Symp. Proc., 673-684, 1984
7 K. R. Erikson, F. J. Fry and J. P. Jones, "Ultrasound in medicine-a review," IEEE Trans. on Sonics and Ultrasonics 21, 144-170, 1974   DOI
8 강국진, 노용래, "Class IV Flextensional 트랜스듀서의 주파수 특성 변화에 관한 연구," 한국음향학회지 18(7), 67-73, Nov. 1999
9 R. E. Miller, Optimization: foundations and applications, (John Wiley & Sons, New York, 2000)
10 N. Felix, D. Certon, E. Lacaze, M. Lethiecq and F. Patat, "Experimental investigation of cross-coupling and its influence on the elementary radiation pattern in 1D ultrasound arrays," IEEE ultras. Sym. Proc., 1053-1056, 1999
11 R. O. Kuehl, Design of experiments: Statistical principles of research design and analysis, (Duxbury Press, Pacific Grove, 2000)
12 S. C. Wooh and Y. Shi, "Optimum beam steering of linear phased arrays," Wave Motion 29, 245-265, 1999   DOI   ScienceOn
13 J. Ylitalo, "Synthetic aperture ultrasound imaging using convex array," IEEE Utras. Sym. Proc., 1337-1340, 1995
14 S. W. Smith, O. T. von Ramm, M. E. Haran and R. L. Thurstone, "Angular response of piezoelectric elements in linear phased array ultrasound scanners," IEEE Trans. Ultrason. Ferroelectr. Freq. Contr. 26, 185-191, 1979
15 D. J. Powell, G. L. Wojcik and B. K. Mukherjee, "Incremental 'Model-build-test'Validation Exercise for a 1-D Biomedical Ultrasonic Imaging Array," IEEE ultras. Sym. Proc., 1669-1674, 1997
16 A. D. Belegudu and T. R. Chandrupatla, Optimization concept and application in engineering, (Prentice Hall, New Jersey, Chap.5, 1999)